00197471-03_Service Manual Internal WPC5_6, EN_01-2019.pdf - 第111页

Service Manual In ternal WPC5 / WPC6  Page 3-111 3.7.18 Sensor 16 N SM mo dul e (WPC6) – Safety Sensor for the Hand Guard Spare Part • Safety sensor hand guard [03056990- xx] Preparations Loosening t he belt tension ➢ L…

100%1 / 172
Service Manual Internal WPC5 / WPC6
Page 3-110
See also.
@3.5.2 Replacing the Load Axis Drive Motor [
3-35]
Loosen the counternut fastening
the sensor on the underside of the
base plate and unscrew the sensor
(1).
Screw the new sensor into the table
plate.
The sensor should protrude
between 0.2 mm and 0.5 mm
above the table plate. Check this
with a straight edge (2).
Use a counternut to fix the sensor
on the underside (1).
Reconnect connector X5 to the
circuit board.
Service Manual Internal WPC5 / WPC6
Page 3-111
3.7.18 Sensor 16 NSM module (WPC6) Safety Sensor for the Hand Guard
Spare Part
Safety sensor hand guard [03056990-xx]
Preparations
Loosening the belt tension
Loosen the tension of the load axis
drive belt (2) at the tensioning
screw behind the tower (1).
Removing the base plate
Loosen the screws fixing the motor
to the base plate at (3).
Lift off the base plate and turn it
over.
Service Manual Internal WPC5 / WPC6
Page 3-112
Removal / Installation
The sensor for the safety flap when open is located on the underside of the base plate.
See also.
@
3.5.1.4
Loosen the Tension of the Load Axis Drive Belt [ 3-34]
@
3.5.6
Setting the Load Axis Belt Tension [
3-44]
Disconnect connector X7 for the
safety flap when open from the
circuit board.
Remove the two screws fastening
the sensor (1).
Insert the new sensor and fix into
place with the two screws.
Reconnect connector X7 to the
circuit board.