00197471-03_Service Manual Internal WPC5_6, EN_01-2019.pdf - 第143页

Service Manual In ternal WPC5 / WPC6  Page 4-143 4.5 Storing and Res toring Machine Da ta Storing machine da ta ➢ Log in as Machine Service . ➢ Select Sensors and Functions (1) => Location . ➢ Select the relevant loc…

100%1 / 172
Service Manual Internal WPC5 / WPC6
Page 4-142
4.4 Performing a WPC Endurance Run
After calibration, we recommend that you test the function by performing a WPC endurance run.
Select
Manual functions -> WPC endurance runs
.
Set a continuous run of 15 mins and start.
Service Manual Internal WPC5 / WPC6
Page 4-143
4.5 Storing and Restoring Machine Data
Storing machine data
Log in as
Machine Service
.
Select
Sensors and Functions
(1) =>
Location
.
Select the relevant location (2).
Select the option
WPC Machine Data
(3).
Enter the WPC serial number (4) and select
Accept
.
Select
Save machine data
(5).
A dialog box for the machine data storage location will open.
Enter the required storage location and then save the machine data..
Service Manual Internal WPC5 / WPC6
Page 4-144
Restoring the machine data
Log in as
Machine service
.
Select
Sensors and Functions
(1) =>
Location
.
Select the relevant location (2).
Select the option
WPC Machine Data
(3).
Enter the WPC serial number (4) and select
Accept
.
Select
Restore machine data
(5).
A dialog box for the machine data storage location will open.
Navigate to the required storage location and then restore the machine data.