Canceling a dynamic impedance measurement
Once you have started the process, you can cancel the measurement in the progress
dialog displayed.
1. Select "Abort".
The R&S LCX displays a message informing you on the recorded log file and its
storage location.
2. Confirm with "OK".
3. To evaluate the report in the log file:
a) If necessary, connect a USB memory device at the front panel "USB" connec-
tor.
b) Use the file manager to copy the log file to the memory device, see Sec-
tion 9.1.3, "File manager settings", on page 122.
c) Upload the file on a PC.
For evaluation, you can use any application that supports *.csv file format.
8.8.5 Binning
Option: R&S LCX-K107
The binning function enables you to sort a set of electrical components according to
the measured characteristics. The R&S LCX measures a connected component and
categorizes it based on a predefined bin configuration. You can see the corresponding
result on the screen, and also get the category assigned at one of the digital outputs of
the interface I/O port.
Specific instrument functions
Instrument functions
R&S
®
LCX Series
111User Manual 1179.2260.02 ─ 08
9
4
6
10
21 3
5
78
Figure 8-25: Binning measurement display and controls
1 = Measurement function indicator
2 = Settings softkey
3 = CSV file
4 = Counter
5 = Sorting characteristics
6 = End of measurement bin
7 = Out bin
8 = Unconfigured bin
9 = User-specified bins
10 = Histogram
8.8.5.1 About binning
To perform binning, the R&S LCX requires the bin configuration you must define and
upload in advance. In addition, you must configure the test signal (i.e. level and fre-
quency) and bias, compensation and measurement range settings before. Changing
these settings is not possible during a binning session.
To specify the bin configuration, you can define up to 8 categories according to 2 mea-
surement characteristics in a *.csv file, e.g. using a suitable application. Optionally,
you can configure the last bin to denote the end of each measurement, independently
of the result. Another control signal denotes if a component exceeds the specified lim-
its. You can assign these results to a bin also.
To upload the file, use a USB flash drive. When the binning function is enabled, the
R&S LCX measures the connected component initiated by a trigger signal. It indicates
the result, i.e. the corresponding bin on the screen, and also assigns a signal at one of
the 8 digital bin outputs.
Specific instrument functions
Instrument functions
R&S
®
LCX Series
112User Manual 1179.2260.02 ─ 08
Contents of the *.csv file:
Up to 2 measurement parameters
The measured value range of each of the two parameters for the individual bin.
The value ranges of the measurement parameters defined for the corresponding
categories (up to 8 bins).
How the limits are specified:
Minimum and maximum values
Nominal value and percentage deviation
Nominal and deviation value
Additional signals you want to have assigned at the output, i.e. end of measure-
ment, out of bin, both).
With the binning function enabled, the R&S LCX switches to measurement mode "trig-
gered" automatically. The measurement starts with the first trigger signal. You can
either trigger the measurement manually, using the [Trig.] key at the front or by sending
the remote command *TRG. It is also possible to apply an external trigger signal to the
"Trigger Input" connector at the rear of the instrument.
During the binning measurements, the R&S LCX is permanently in the triggered mode
and cannot be disabled.
How to: see Section 8.8.5.5, "Configuring and using the binning function",
on page 116.
For details on the trigger modes, see the specifications document, section "Digital trig-
ger and control interfaces".
8.8.5.2 Assignment of the binning interface
The digital I/O port of the binning interface is a 15-pole D-Sub female socket, including
eight data lines. The remaining pins provide digital input and output of control signals.
Figure 8-26: Digital I/O connector, front view
For details on the interface specification, see section "Digital trigger and control interfa-
ces" in the specifications document.