LCX_UserManual_en_08_1.pdf - 第213页

Remote control commands R&S ® LCX Series 213 User Manual 1 179.2260.02 ─ 08 Example: See Example "Selecting the measurement function and setting the impedance range" on page 210. Manual operation: See &quot…

100%1 / 265
Remote control commands
R&S
®
LCX Series
212User Manual 1179.2260.02 ─ 08
Example: See Example "Selecting the measurement function and setting
the impedance range" on page 210.
Manual operation: See "Range: / [Range]" on page 75
FUNCtion:IMPedance:SOURce <Impedance>
Selects the output impedance for the measurement.
Parameters:
<Impedance> LOW | R10 | HIGH | R100
LOW | R10
Sets 10 Ω output impedance.
HIGH | R100
Sets 100 Ω output impedance.
*RST: HIGH
Example: See Example "Selecting the measurement function and setting
the impedance range" on page 210.
Manual operation: See "Source Impedance" on page 80
FUNCtion:IMPedance[:TYPE][:PRIMary] <Impedance Type>
Selects the primary impedance parameter for the measurement corresponding to the
measurement type, see FUNCtion:MEASurement:TYPE on page 214.
Parameters:
<Impedance Type> DEFault | CPD | CPQ | CPG | CPRP | CSD | CSQ | CSRS |
LPD | LPQ | LPG | LPRP | LSD | LSQ | LSRS | RX | RPB | RDC |
MTD | NTD | ZTD | ZTR | GB | YTD | YTR
DEFault
Sets the default setting ZTD (impedance phase angle degree).
CPD | CPQ | CPG | CPRP | CSD | CSQ | CSRS
Capacitive measurement type: Cp (parallel capacitance), Cs
(serial capacitance), D (dissipation factor), Q (quality factor), G
(conductance), Rp (parallel resistance), Rs (serial resistance)
LPD | LPQ | LPG | LPRP | LSD | LSQ | LSRS
Inductive measurement type: Lp (parallel inductance), Ls (serial
inductance), D (dissipation factor), Q (quality factor), G (conduc-
tance), Rp (parallel resistance), Rs (serial resistance), RDc
(direct current resistance)
RX | RPB | RDC | MTD | NTD | ZTD | ZTR | GB | YTD | YTR
Resistance measurement type: R (resistance), X impedance, Rp
(parallel resistance), RDC (direct current resistance), B (suscep-
tance), M (mutual inductance), N (transformer ratio), Z (impe-
dance), G (conductance), Y (admittance), TD (phase angle
degree), TR (phase angle rad)
FUNCtion subsystem
Remote control commands
R&S
®
LCX Series
213User Manual 1179.2260.02 ─ 08
Example: See Example "Selecting the measurement function and setting
the impedance range" on page 210.
Manual operation: See "<measurement pair>" on page 73
FUNCtion:IMPedance:TYPE:SECondary <Impedance Type>
Selects the secondary impedance parameter for the measurement corresponding to
the measurement type, see FUNCtion:MEASurement:TYPE on page 214.
Parameters:
<Impedance Type> DEFault | CPD | CPQ | CPG | CPRP | CSD | CSQ | CSRS |
LPD | LPQ | LPG | LPRP | LSD | LSQ | LSRS | RX | RPB | RDC |
MTD | NTD | ZTD | ZTR | GB | YTD | YTR
DEFault
Sets the default setting ZTD (impedance phase angle degree).
CPD | CPQ | CPG | CPRP | CSD | CSQ | CSRS
Capacitive measurement type: Cp (parallel capacitance), Cs
(serial capacitance), D (dissipation factor), Q (quality factor), G
(conductance), Rp (parallel resistance), Rs (serial resistance)
LPD | LPQ | LPG | LPRP | LSD | LSQ | LSRS
Inductive measurement type: Lp (parallel inductance), Ls (serial
inductance), D (dissipation factor), Q (quality factor), G (conduc-
tance), Rp (parallel resistance), Rs (serial resistance), RDc
(direct current resistance)
RX | RPB | RDC | MTD | NTD | ZTD | ZTR | GB | YTD | YTR
Resistance measurement type: R (resistance), X impedance, Rp
(parallel resistance), RDC (direct current resistance), B (suscep-
tance), M (mutual inductance), N (transformer ratio), Z (impe-
dance), G (conductance), Y (admittance), TD (phase angle
degree), TR (phase angle rad)
Example:
See Example "Selecting the measurement function and setting
the impedance range" on page 210.
Manual operation: See "<measurement pair>" on page 73
FUNCtion:MEASurement:AUToselect
Triggers the automatic selection of the measurement function and the corresponding
default measurement pairs.
Example:
See Example "Selecting the measurement function and setting
the impedance range" on page 210.
Usage: Event
Asynchronous command
Manual operation: See "[Auto]" on page 76
FUNCtion subsystem
Remote control commands
R&S
®
LCX Series
214User Manual 1179.2260.02 ─ 08
FUNCtion:MEASurement:TYPE <Measurement Type>
Selects the measurement function.
Parameters:
<Measurement Type> L | C | R | T
L
Impedance measurement.
C
Capacitance measurement.
R
Resistance measurement.
T
Transformer measurement.
*RST: C
Example: See Example "Selecting the measurement function and setting
the impedance range" on page 210.
Usage: Asynchronous command
Manual operation: See "Functions [hardkeys]" on page 73
FUNCtion:TRANsformer:RANGe[:TYPE] <Turn Ratio>
Selects the impedance range for transformer measurement.
Parameters:
<Turn Ratio> N50 | N500
Example: See Example "Selecting the measurement function and setting
the impedance range" on page 210.
11.12 HANDler subsystem
The HANDler subsystem contains the commands for configuring the binning function.
HANDler subsystem