Remote control commands R&S ® LCX Series 193 User Manual 1 179.2260.02 ─ 08 CORRection:LOAD:MODE? Queries the state of load correction. Return values: <Load Correction Mode> OFF | UNCorrected | SPOT | QSPot | F…

100%1 / 265
Remote control commands
R&S
®
LCX Series
192User Manual 1179.2260.02 ─ 08
// ***************************
// Query the current status of correction
CORRection:OPEN:STATe?
// Response: 1 // 1:active, 0: inactive
CORRection:SHORt:STATe?
CORRection:LOAD:STATe?
// ***************************
// Query the activated mode of a correction
CORRection:OPEN:MODE?
// Response: "FULL"
CORRection:SHORt:MODE?
// Response: "SPOT"
CORRection:LOAD:MODE?
// Response: "UNCorrected"
CORRection:LENGth......................................................................................................192
CORRection:LOAD:MODE?............................................................................................193
CORRection:LOAD:STATe.............................................................................................. 193
CORRection:OPEN:MODE?............................................................................................193
CORRection:OPEN:STATe..............................................................................................194
CORRection:OPEN[:EXECute]........................................................................................194
CORRection:SHORt:MODE?.......................................................................................... 194
CORRection:SHORt:STATe.............................................................................................195
CORRection:SHORt[:EXECute].......................................................................................195
CORRection:SPOT:LOAD:STANdard............................................................................... 195
CORRection:SPOT:LOAD[:EXECute]...............................................................................196
CORRection:SPOT:OPEN[:EXECute]...............................................................................196
CORRection:SPOT:SHORt[:EXECute]..............................................................................196
CORRection:STORage:RESTore.....................................................................................196
CORRection:STORage:SAVE..........................................................................................196
CORRection:LENGth <Cable Length>
Sets the length of the leads to the connected test fixture i.e. the DUT.
Parameters:
<Cable Length> Range: 0 m to 1 m
*RST: 0 m
Default unit: m
Example: See Example "Performing open/short/load correction"
on page 191.
Manual operation: See "Cable Length" on page 80
CORRection subsystem
Remote control commands
R&S
®
LCX Series
193User Manual 1179.2260.02 ─ 08
CORRection:LOAD:MODE?
Queries the state of load correction.
Return values:
<Load Correction
Mode>
OFF | UNCorrected | SPOT | QSPot | FULL
OFF
The instrument disabled correction data.
UNCorrected
Correction is enabled, but the correction data does not match
the current settings or is not available.
SPOT
Correction is enabled with the spot correction data of the current
measurement settings.
QSPot
Correction is enabled with spot correction data of measurement
settings other than the current settings.
FULL
Correction is enabled with full correction data.
Example:
See Example "Performing open/short/load correction"
on page 191.
Usage: Query only
CORRection:LOAD:STATe <Load Correction State>
Enables the load correction function.
Parameters:
<Load Correction
State>
*RST: OFF
Example: See Example "Performing open/short/load correction"
on page 191.
Manual operation: See "Open/Short/Load Correction Enabled" on page 83
CORRection:OPEN:MODE?
Queries the current state of open correction.
Return values:
<Open Correction
Mode>
OFF | UNCorrected | SPOT | QSPot | FULL
OFF
The instrument disabled correction data.
UNCorrected
Correction is enabled, but the correction data does not match
the current settings or is not available.
CORRection subsystem
Remote control commands
R&S
®
LCX Series
194User Manual 1179.2260.02 ─ 08
SPOT
Correction is enabled with the spot correction data of the current
measurement settings.
QSPot
Correction is enabled with spot correction data of measurement
settings other than the current settings.
Example: See Example "Performing open/short/load correction"
on page 191.
Usage: Query only
CORRection:OPEN:STATe
<Open Correction State>
Enables the open correction function.
Parameters:
<Open Correction
State>
1 | 0 | ON | OFF
*RST: OFF
Example: See Example "Performing open/short/load correction"
on page 191.
Manual operation: See "Open/Short/Load Correction Enabled" on page 83
CORRection:OPEN[:EXECute]
Executes an open correction on all frequencies.
Example:
See Example "Performing open/short/load correction"
on page 191.
Usage: Event
Asynchronous command
Manual operation: See "Start Short/Open Correction (Full)" on page 84
CORRection:SHORt:MODE?
Queries the state of short correction.
Return values:
<Short Correction
Mode>
OFF | UNCorrected | SPOT | QSPot | FULL
OFF
The instrument disabled correction data.
UNCorrected
Correction is enabled, but the correction data does not match
the current measurement settings or is not available.
SPOT
Correction is enabled with the spot correction data of the current
measurement settings.
CORRection subsystem