LCX_UserManual_en_08_1.pdf - 第188页
Remote control commands R&S ® LCX Series 188 User Manual 1 179.2260.02 ─ 08 VOL T age[:LEV el] <V oltage Level> Sets the test signal voltage in RMS (root mean square). Parameters: <V oltage Level> <num…

Remote control commands
R&S
®
LCX Series
187User Manual 1179.2260.02 ─ 08
DEFault
Uses the default setting "SHORt".
LONG
Sets the measurement time ≤0.500 s.
MEDium
Sets the measurement time ≤0.100 s.
SHORt
Sets the measurement time ≤.015 s.
SSHort
Sets the measurement time ≤0.004 s.
*RST:
LONG
Example: See Example "Setting key parameters" on page 186.
Manual operation: See "Speed" on page 80
CURRent[:LEVel]
<Current Level>
Sets the test signal current in RMS (root mean square).
Parameters:
<Current Level> Range: 0.0001 A to 0.2 A (depends on measurement set-
tings)
Increment: 0.00001 A
Default unit: A
Example: See Example "Setting key parameters" on page 186.
Manual operation: See "Level: / [Level]" on page 74
FREQuency[:CW] <Test Signal Frequency>
Sets the frequency of the test signal.
Parameters:
<Frequency> float
Range:
4 to 300 * 10
3
(depends on instrument model, the
equipped options and measurement settings)
Increment: 0.1 Hz
*RST: 1E+3 Hz
Default unit: Hz
Example: See Example "Setting key parameters" on page 186.
Manual operation: See "Frequency: / [Freq.]" on page 74
Test signal commands

Remote control commands
R&S
®
LCX Series
188User Manual 1179.2260.02 ─ 08
VOLTage[:LEVel] <Voltage Level>
Sets the test signal voltage in RMS (root mean square).
Parameters:
<Voltage Level> <numeric>
Sets the value.
MIN | MINimum
Queries the lower limit of the signal level.
MAX | MAXimum
Queries the upper limit of the signal level.
DEF | DEFault
Queries the signal level the instrument sets by default.
Range: 0.01 V to 10 V (depends on measurement settings)
Increment: 0.001 V
*RST: 1 V
Default unit: V
Example: See Example "Setting key parameters" on page 186.
Manual operation: See "Level: / [Level]" on page 74
11.5 BIAS subsystem
The BIAS subsystem contains the commands for bias configuration.
Example: Bias settings
The example represents the remote control commands for applying either an external
or an internal DC bias level.
// Assign an internally generated voltage level and enable bias.
BIAS:VOLTage[:LEVel] 1
BIAS:CURRent[:LEVel] 0.1
BIAS:STATe 1
// Query an externally applied voltage level and enable bias.
BIAS:EXTernal:MEASure:VOLTage?
// Response: 0.2
BIAS:EXTernal[:VOLTage][:STATe]ON
// Response: 1
BIAS:CURRent[:LEVel]...................................................................................................188
BIAS:EXTernal:MEASure:VOLTage?................................................................................ 189
BIAS:EXTernal[:VOLTage][:STATe]...................................................................................189
BIAS:STATe...................................................................................................................189
BIAS:VOLTage[:LEVel]....................................................................................................189
BIAS:CURRent[:LEVel] <Bias Current Level>
Sets the internal bias current value.
BIAS subsystem

Remote control commands
R&S
®
LCX Series
189User Manual 1179.2260.02 ─ 08
To activate the bias, use command BIAS:STATe on page 189.
Parameters:
BiasCurrent Range: 0 to 0.2 A
Increment: 0.001 A
*RST: 0 A
Default unit: A
Example: See Example "Bias settings" on page 188.
Manual operation: See "Current Bias: / [Bias Level]" on page 75
BIAS:EXTernal:MEASure:VOLTage?
Queries the value of the externally applied voltage.
Parameters for setting and query:
<voltage> Default unit: V
Example: See Example "Bias settings" on page 188.
Usage: Query only
BIAS:EXTernal[:VOLTage][:STATe] <External Voltage Bias State>
Enables the externally supplied bias voltage.
Parameters:
<External Voltage
Bias State>
1 | 0 | ON | OFF
*RST: OFF
Example: See Example "Bias settings" on page 188.
BIAS:STATe <Bias State>
Enables the internal DC bias.
To set the corresponding bias voltage or current value, use commands BIAS:
CURRent[:LEVel] on page 188, or BIAS:VOLTage[:LEVel] on page 189.
Parameters:
<Bias State> 1 | 0 | ON | OFF
*RST: OFF
Example: See Example "Bias settings" on page 188.
Manual operation: See "[Bias Enable]" on page 75
See "Voltage Bias" on page 81
BIAS:VOLTage[:LEVel] <Bias Voltage Level>
Sets the internal DC bias voltage value.
To enable the bias, use command BIAS:STATe on page 189.
BIAS subsystem