LCX_UserManual_en_08_1.pdf - 第188页

Remote control commands R&S ® LCX Series 188 User Manual 1 179.2260.02 ─ 08 VOL T age[:LEV el] <V oltage Level> Sets the test signal voltage in RMS (root mean square). Parameters: <V oltage Level> <num…

100%1 / 265
Remote control commands
R&S
®
LCX Series
187User Manual 1179.2260.02 ─ 08
DEFault
Uses the default setting "SHORt".
LONG
Sets the measurement time 0.500 s.
MEDium
Sets the measurement time 0.100 s.
SHORt
Sets the measurement time .015 s.
SSHort
Sets the measurement time 0.004 s.
*RST:
LONG
Example: See Example "Setting key parameters" on page 186.
Manual operation: See "Speed" on page 80
CURRent[:LEVel]
<Current Level>
Sets the test signal current in RMS (root mean square).
Parameters:
<Current Level> Range: 0.0001 A to 0.2 A (depends on measurement set-
tings)
Increment: 0.00001 A
Default unit: A
Example: See Example "Setting key parameters" on page 186.
Manual operation: See "Level: / [Level]" on page 74
FREQuency[:CW] <Test Signal Frequency>
Sets the frequency of the test signal.
Parameters:
<Frequency> float
Range:
4 to 300 * 10
3
(depends on instrument model, the
equipped options and measurement settings)
Increment: 0.1 Hz
*RST: 1E+3 Hz
Default unit: Hz
Example: See Example "Setting key parameters" on page 186.
Manual operation: See "Frequency: / [Freq.]" on page 74
Test signal commands
Remote control commands
R&S
®
LCX Series
188User Manual 1179.2260.02 ─ 08
VOLTage[:LEVel] <Voltage Level>
Sets the test signal voltage in RMS (root mean square).
Parameters:
<Voltage Level> <numeric>
Sets the value.
MIN | MINimum
Queries the lower limit of the signal level.
MAX | MAXimum
Queries the upper limit of the signal level.
DEF | DEFault
Queries the signal level the instrument sets by default.
Range: 0.01 V to 10 V (depends on measurement settings)
Increment: 0.001 V
*RST: 1 V
Default unit: V
Example: See Example "Setting key parameters" on page 186.
Manual operation: See "Level: / [Level]" on page 74
11.5 BIAS subsystem
The BIAS subsystem contains the commands for bias configuration.
Example: Bias settings
The example represents the remote control commands for applying either an external
or an internal DC bias level.
// Assign an internally generated voltage level and enable bias.
BIAS:VOLTage[:LEVel] 1
BIAS:CURRent[:LEVel] 0.1
BIAS:STATe 1
// Query an externally applied voltage level and enable bias.
BIAS:EXTernal:MEASure:VOLTage?
// Response: 0.2
BIAS:EXTernal[:VOLTage][:STATe]ON
// Response: 1
BIAS:CURRent[:LEVel]...................................................................................................188
BIAS:EXTernal:MEASure:VOLTage?................................................................................ 189
BIAS:EXTernal[:VOLTage][:STATe]...................................................................................189
BIAS:STATe...................................................................................................................189
BIAS:VOLTage[:LEVel]....................................................................................................189
BIAS:CURRent[:LEVel] <Bias Current Level>
Sets the internal bias current value.
BIAS subsystem
Remote control commands
R&S
®
LCX Series
189User Manual 1179.2260.02 ─ 08
To activate the bias, use command BIAS:STATe on page 189.
Parameters:
BiasCurrent Range: 0 to 0.2 A
Increment: 0.001 A
*RST: 0 A
Default unit: A
Example: See Example "Bias settings" on page 188.
Manual operation: See "Current Bias: / [Bias Level]" on page 75
BIAS:EXTernal:MEASure:VOLTage?
Queries the value of the externally applied voltage.
Parameters for setting and query:
<voltage> Default unit: V
Example: See Example "Bias settings" on page 188.
Usage: Query only
BIAS:EXTernal[:VOLTage][:STATe] <External Voltage Bias State>
Enables the externally supplied bias voltage.
Parameters:
<External Voltage
Bias State>
1 | 0 | ON | OFF
*RST: OFF
Example: See Example "Bias settings" on page 188.
BIAS:STATe <Bias State>
Enables the internal DC bias.
To set the corresponding bias voltage or current value, use commands BIAS:
CURRent[:LEVel] on page 188, or BIAS:VOLTage[:LEVel] on page 189.
Parameters:
<Bias State> 1 | 0 | ON | OFF
*RST: OFF
Example: See Example "Bias settings" on page 188.
Manual operation: See "[Bias Enable]" on page 75
See "Voltage Bias" on page 81
BIAS:VOLTage[:LEVel] <Bias Voltage Level>
Sets the internal DC bias voltage value.
To enable the bias, use command BIAS:STATe on page 189.
BIAS subsystem