Instrument functions R&S ® LCX Series 76 User Manual 1 179.2260.02 ─ 08 [Auto] ............................................................................................................................. 76 [Hold] .…

100%1 / 265
Instrument functions
R&S
®
LCX Series
75User Manual 1179.2260.02 ─ 08
"Voltage Level"
Sets the signal voltage.
"Current Level"
Sets the signal current.
Remote command:
VOLTage[:LEVel] on page 188
CURRent[:LEVel] on page 187
Range: / [Range]
Selects the measurement range.
Remote command:
FUNCtion:IMPedance:RANGe:AUTO on page 211
FUNCtion:IMPedance:RANGe[:VALue] on page 211
Current Bias: / [Bias Level]
Sets the internal the current preload.
Remote command:
BIAS:CURRent[:LEVel] on page 188
Voltage Bias: / [Bias Level]
Sets the value of the internal voltage preload.
Remote command:
BIAS:VOLTage[:LEVel] on page 189
[Comp.]
Accesses the "Open/Short/Load Correction" dialog, see Section 8.6.1, "Measurement
parameter settings", on page 78.
[Bias Enable]
Enables the internal voltage or current bias.
Remote command:
BIAS:STATe on page 189
8.5 Measurement control functions
With the measurement control hardkeys at the front panel, you can control a running
measurement.
Access:
Press the corresponding key ([Auto], [Hold] or [Trig.]) at the front panel.
Repeated pressing the [Hold] key turns the function on and off (toggle). The [Auto]
and the [Trig.] keys trigger the corresponding event.
The required remote commands are described in Section 11.11, "FUNCtion subsys-
tem", on page 209.
Measurement control functions
Instrument functions
R&S
®
LCX Series
76User Manual 1179.2260.02 ─ 08
[Auto].............................................................................................................................76
[Hold].............................................................................................................................76
[Trig.]............................................................................................................................. 76
[Auto]
Selects measurement function and the related default measurement pairs automati-
cally, according to the connected sample component.
For an overview of the impedance measurement parameters the R&S LCX provides,
see Section 5.2, "Impedance measurement parameters of the R&S LCX", on page 50.
Remote command:
FUNCtion:MEASurement:AUToselect on page 213
[Hold]
Freezes the currently set range.
When enabled, the R&S LCX displays the prefix "Hold:" at the range value.
Remote command:
FUNCtion:IMPedance:RANGe:HOLD on page 211
[Trig.]
Triggers one measurement cycle manually.
Remote command:
*TRG on page 185
TRIGger[:IMMediate] on page 243
8.6 Measurement parameters
The settings described in this section refer to parameters that you can configure for
your specific measurement in addition to the test signal, measurement mode, or mea-
surement functions.
About measurement parameters
Aspects on the measurement parameters:
Speed
The measurement speed defines the number of measurements in one second
when triggered continuously.
Impact of measurement speed and accuracy:
The measurement accuracy decreases with increasing speed, i.e. the higher the
number of measurements per second you perform, the lower the measurement
accuracy you can expect.
At low test signal frequencies, the R&S LCX reduces the measurement speed
automatically.
See "Speed" on page 80.
Cable length
Any cable covers inductive and capacitive parts, which depend in particular on the
cable length. As these characteristics are considered as parasitics, they distort a
Measurement parameters
Instrument functions
R&S
®
LCX Series
77User Manual 1179.2260.02 ─ 08
measurement result. The cable length correction compensates these parasitic
transmission characteristics.
As a rule, make sure to keep the connection between the R&S LCX and the com-
ponent under test as short as possible, even more, the length must not exceed
1 m.
When using test fixtures with cables, the cable length correction function reduces
the impact of cable losses to a minimum, and thus compensates the parasitic
transmission characteristics. See "Cable Length" on page 80 for the correspond-
ing setting.
Test signal type
The R&S LCX provides either the signal voltage or the signal current for setting the
test signal level, see "Test Signal Type" on page 80. If the signal current is
defined, the set value corresponds to the maximum current that flows when short-
circuited.
External voltage bias
See "Voltage Bias" on page 81.
Source impedance
By default, the R&S LCX measures with an output impedance of 100 Ω over the
entire measurement range ("High Z"). For certain applications, e.g., the measure-
ment of low-impedance components, you can set the output impedance to 10 Ω
("Low Z").
See "Source Impedance" on page 80.
Open/Short/Load correction
The numerous types of passive components with a wide variety of shapes require
suitable test fixtures, which in turn are equipped with or without test leads, clips or
probes, etc. As a result, measurement errors due to the test setup and used equip-
ment can occur. In particular, residual impedance, admittance, stray capacities or
environmental factors can impact the measurement.
Using the open/short/load correction technique, the R&S LCX compensates such
impacts.
For example, the open/short correction offsets residuals of test fixtures. For com-
plex errors, you can use the load correction function, which refers to a reference
component, e.g., a reference resistor. We recommend that you perform Open/Short
and Load correction when the measurement setup covers cable lengths other than
standard cables, user-specific test fixtures or additional passive circuits or compo-
nents, etc.
Open/Short/Spot corrections are always performed with the current measurement
settings (frequency, voltage, range, low Z/high Z, etc.). The measurement settings
are not changed for the measurement.
Spot correction
Note: To achieve the most accurate spot correction, we recommend that you per-
form all spot corrections in the "Auto" measurement range. The "Auto" range set-
ting selects the best measurement range for reading spot correction data with the
connected DUT. Running a spot correction with a manually selected measurement
range, or "Auto" range and "Hold", requires that you know the best measurement
range. Selecting an incorrect measurement range results in inaccurate correction
data and distorts the measurement results.
Measurement parameters