Remote control commands R&S ® LCX Series 195 User Manual 1 179.2260.02 ─ 08 QSPot Correction is enabled with spot correction data of measurement settings other than the current settings. Example: See Example "Pe…

Remote control commands
R&S
®
LCX Series
194User Manual 1179.2260.02 ─ 08
SPOT
Correction is enabled with the spot correction data of the current
measurement settings.
QSPot
Correction is enabled with spot correction data of measurement
settings other than the current settings.
Example: See Example "Performing open/short/load correction"
on page 191.
Usage: Query only
CORRection:OPEN:STATe
<Open Correction State>
Enables the open correction function.
Parameters:
<Open Correction
State>
1 | 0 | ON | OFF
*RST: OFF
Example: See Example "Performing open/short/load correction"
on page 191.
Manual operation: See "Open/Short/Load Correction Enabled" on page 83
CORRection:OPEN[:EXECute]
Executes an open correction on all frequencies.
Example:
See Example "Performing open/short/load correction"
on page 191.
Usage: Event
Asynchronous command
Manual operation: See "Start Short/Open Correction (Full)" on page 84
CORRection:SHORt:MODE?
Queries the state of short correction.
Return values:
<Short Correction
Mode>
OFF | UNCorrected | SPOT | QSPot | FULL
OFF
The instrument disabled correction data.
UNCorrected
Correction is enabled, but the correction data does not match
the current measurement settings or is not available.
SPOT
Correction is enabled with the spot correction data of the current
measurement settings.
CORRection subsystem

Remote control commands
R&S
®
LCX Series
195User Manual 1179.2260.02 ─ 08
QSPot
Correction is enabled with spot correction data of measurement
settings other than the current settings.
Example: See Example "Performing open/short/load correction"
on page 191.
Usage: Query only
CORRection:SHORt:STATe <Short Correction State>
Enables the short correction function.
Parameters:
<Short Correction
State>
1 | 0 | ON | OFF
*RST: OFF
Example: See Example "Performing open/short/load correction"
on page 191.
Manual operation: See "Open/Short/Load Correction Enabled" on page 83
CORRection:SHORt[:EXECute]
Executes a short correction on all frequencies.
Example:
See Example "Performing open/short/load correction"
on page 191.
Usage: Event
Asynchronous command
Manual operation: See "Start Short/Open Correction (Full)" on page 84
CORRection:SPOT:LOAD:STANdard <Reference value for primary>, <Reference
value for secondary>
Defines a working point for load correction.
Assign the working point number and the primary and secondary reference values.
Parameters:
<Reference value for
primary>
Sets the primary standard value as reference, e.g. the value of a
calibration resistor.
Default unit: Ohm
<Reference value for
secondary>
Sets the secondary standard value.
Default unit: Ohm
Example: See Example "Performing open/short/load correction"
on page 191.
Manual operation: See "Reference Load R" on page 85
See "Reference Load X" on page 85
CORRection subsystem

Remote control commands
R&S
®
LCX Series
196User Manual 1179.2260.02 ─ 08
CORRection:SPOT:LOAD[:EXECute]
Performs a load correction at a dedicated working point.
Example: See Example "Performing open/short/load correction"
on page 191.
Usage: Event
Asynchronous command
Manual operation: See "Start Load Correction (Spot)" on page 85
CORRection:SPOT:OPEN[:EXECute]
Performs an open correction at a dedicated working point.
Example:
See Example "Performing open/short/load correction"
on page 191.
Usage: Event
Asynchronous command
Manual operation: See "Start Short/Open Correction (Spot)" on page 84
CORRection:SPOT:SHORt[:EXECute]
Performs a short correction at a dedicated working point.
Example:
See Example "Performing open/short/load correction"
on page 191.
Usage: Event
Asynchronous command
Manual operation: See "Start Short/Open Correction (Spot)" on page 84
CORRection:STORage:RESTore <Storage Slot>
Recalls the correction data set that has been stored before in a memory slot.
Setting parameters:
<Storage Slot> Range: 1 to 10
Example: See Example "Performing open/short/load correction"
on page 191.
Usage: Setting only
Manual operation: See "Restore Corrections" on page 86
CORRection:STORage:SAVE <Storage Slot>
Saves the correction data set of a full open or short correction.
CORRection subsystem