Instrument functions R&S ® LCX Series 84 User Manual 1 179.2260.02 ─ 08 Remote command: CORRection:SHORt:STATe on page 195 CORRection:OPEN:STATe on page 194 CORRection:LOAD:STATe on page 193 Start Short/Open Correcti…

Instrument functions
R&S
®
LCX Series
83User Manual 1179.2260.02 ─ 08
3. Select "Open/Short/Load Correction".
Figure 8-3: Open/Short/Load correction dialog
In this menu, you can enable the corresponding corrections directly, and access
the subdialogs to configure and start the corrections. You can save the correction
data of "Open/Short (FULL)" for later reuse.
The required remote commands are described in: Section 11.6, "CORRection sub-
system", on page 190
How to: see Section 8.6.3, "Performing open/short/load correction", on page 86.
If you need to perform the open and short correction frequently, you can assign these
functions to the [★ (User)] key and perform them with just one keystroke, see Sec-
tion 9.3, "User button", on page 127.
Settings
Open/Short/Load Correction Enabled...........................................................................83
Start Short/Open Correction (Spot)...............................................................................84
Start Short/Open Correction (Full)................................................................................ 84
Start Load Correction (Spot)......................................................................................... 85
Reference Load R.........................................................................................................85
Reference Load X......................................................................................................... 85
Correction storage.........................................................................................................85
Save Corrections...........................................................................................................85
Restore Corrections...................................................................................................... 86
Open/Short/Load Correction Enabled
Enables the corresponding correction function.
To start a correction process, select the corresponding Start Short/Open Correction
(Full) or Start Short/Open Correction (Spot) entry of the respective correction.
Measurement parameters

Instrument functions
R&S
®
LCX Series
84User Manual 1179.2260.02 ─ 08
Remote command:
CORRection:SHORt:STATe on page 195
CORRection:OPEN:STATe on page 194
CORRection:LOAD:STATe on page 193
Start Short/Open Correction (Spot)
Figure 8-4: Start correction at one working point by the example of the short correction dialog
Performs the correction at a defined working point.
The function requires that you have enabled the state, see "Open/Short/Load Correc-
tion Enabled" on page 83, and set the test fixture to open, or short accordingly.
How to: see "Aligning the instrument on a dedicated working point" on page 88
Remote command:
CORRection:SPOT:OPEN[:EXECute] on page 196
CORRection:SPOT:SHORt[:EXECute] on page 196
Start Short/Open Correction (Full)
Figure 8-5: Start correction on all frequencies by the example of the open correction dialog
Performs the corresponding correction over the entire measurement range.
The function requires that you have enabled the state, see "Open/Short/Load Correc-
tion Enabled" on page 83, and set the test fixture to open, or short accordingly.
How to: see "Aligning the instrument on all frequencies" on page 87
Remote command:
CORRection:OPEN[:EXECute] on page 194
CORRection:SHORt[:EXECute] on page 195
Measurement parameters

Instrument functions
R&S
®
LCX Series
85User Manual 1179.2260.02 ─ 08
Start Load Correction (Spot)
Figure 8-6: Start load correction at one working point with a specified reference value
Performs the load correction at a defined working point.
The function requires that you have enabled the state, see "Open/Short/Load Correc-
tion Enabled" on page 83 and specified the real resistance and the complex reactance
of the reference component. The R&S LCX prompts you to connect the specified load.
How to: see "Aligning the instrument using load correction" on page 89.
Remote command:
CORRection:SPOT:LOAD[:EXECute] on page 196
Reference Load R
Sets the resistance value as the refence for the load correction.
Remote command:
CORRection:SPOT:LOAD:STANdard on page 195
Reference Load X
Sets the impedance value as a reference for the load correction.
Remote command:
CORRection:SPOT:LOAD:STANdard on page 195
Correction storage
Provides access to save and restore the data of "Open/Short (FULL)" correction.
Figure 8-7: Correction storage menu
The R&S LCX provides 10 internal memory slots to save and restore a correction data
set.
How to: see "Saving/recalling full correction data" on page 90
Save Corrections
Opens an on-screen editor for selecting a memory slot for saving correction data.
Measurement parameters