LCX_UserManual_en_08_1.pdf - 第225页
Remote control commands R&S ® LCX Series 225 User Manual 1 179.2260.02 ─ 08 Return values: <Primary Measurement V alue> Selects the first measurement parameter . <Secondary Measurement V alue> Selects the…

Remote control commands
R&S
®
LCX Series
224User Manual 1179.2260.02 ─ 08
MEASure:R?................................................................................................................. 225
MEASure:T?..................................................................................................................225
MEASure[:TYPE]?......................................................................................................... 226
MEASure:C? [<Capacitance Type Pair>]
Triggers a new capacitance measurement, reads the resistance value and returns the
result of the specified measurement pair. Without specifying the measurement parame-
ters, the instrument returns the value of the default serial capacitance and dissipation
factor CSD.
Query parameters:
<Capacitance Type
Pair>
DEFault | CSQ | CSD | CSRS | CPQ | CPD | CPRP | CPG | GB |
RX | ZTR | ZTD | YTR | YTD
Return values:
<Primary
Measurement Value>
Selects the first measurement parameter.
<Secondary
Measurement Value>
Selects the second measurement parameter.
Example: See Example "Querying measurement results" on page 223.
Usage: Query only
MEASure:IMPedance?
Triggers a new measurement and queries the impedance and phase angle values
ZTD.
Return values:
<Impedance
Measurement Value>
Selects the impedance measurement parameter.
<Phase Measurement
Degree>
Selects the phase angle measurement parameter.
Example: See Example "Querying measurement results" on page 223.
Usage: Query only
MEASure:L? [<Inductance Type Pair>]
Triggers a new inductance measurement, reads the resistance value and returns the
result of the specified measurement pair. Without specifying the measurement parame-
ters, the instrument returns the value of the default serial inductance and quality factor
LSQ.
Query parameters:
<Inductance Type
Pair>
DEFault | LSQ | LSD | LSRS | LPQ | LPD | LPRP | LPG | RX |
GB | ZTR | ZTD | YTR | YTD
MEASure subsystem

Remote control commands
R&S
®
LCX Series
225User Manual 1179.2260.02 ─ 08
Return values:
<Primary
Measurement Value>
Selects the first measurement parameter.
<Secondary
Measurement Value>
Selects the second measurement parameter.
Example: See Example "Querying measurement results" on page 223.
Usage: Query only
MEASure:R? [<Resistance Type Pair>]
Triggers a new DC resistance measurement, reads the resistance value and returns
the result of the specified measurement pair. Without specifying the measurement
parameters, the instrument returns the value of the default serial resistance and reac-
tance RX.
Query parameters:
<Resistance Type
Pair>
DEFault | RX | RPB | RDC | GB | ZTR | ZTD | YTR | YTD
Return values:
<Primary
Measurement Value>
Selects the first measurement parameter.
<Secondary
Measurement Value>
Selects the second measurement parameter.
Usage: Query only
MEASure:T? [<Transformer Type Pair>]
Triggers a new transformer measurement, reads the resistance value and returns the
result of the specified measurement pair. Without specifying the measurement parame-
ters, the instrument returns the value of the default transformer ratio and phase angle
NTD.
Query parameters:
<Transformer Type
Pair>
DEFault | MTD | NTD | RX | GB | ZTR | ZTD | YTR | YTD
Return values:
<Primary
Measurement Value>
Selects the first measurement parameter.
<Secondary
Measurement Value>
Selects the second measurement parameter.
Example: See Example "Querying measurement results" on page 223.
Usage: Query only
MEASure subsystem

Remote control commands
R&S
®
LCX Series
226User Manual 1179.2260.02 ─ 08
MEASure[:TYPE]? [<Measurement Type Pair>]
Triggers a new measurement and queries the measured values of the specified mea-
surement pair. Without specifying the measurement parameter, the instrument returns
the value of the default impedance measurement ZTD (impedance and phase angle).
Query parameters:
<Measurement Type
Pair>
DEFault | CPD | CPQ | CPG | CPRP | CSD | CSQ | CSRS |
LPD | LPQ | LPG | LPRP | LSD | LSQ | LSRS | RX | RPB | RDC |
MTD | NTD | ZTD | ZTR | GB | YTD | YTR
Return values:
<Primary
Measurement Value>
Selects the first measurement parameter.
<Secondary
Measurement Value>
Selects the second measurement parameter.
Example: See Example "Querying measurement results" on page 223.
Usage: Query only
11.16 READ subsystem
This section describes the remote control commands to read the results of a measure-
ment. These commands are similar to the corresponding commands of the MEASure
subsytem.
READ subsystem