Remote control commands R&S ® LCX Series 207 User Manual 1 179.2260.02 ─ 08 Return values: <AC Current Measurement V alue> Default unit: A Example: See Example "Querying measurement values" on page 20…

Remote control commands
R&S
®
LCX Series
206User Manual 1179.2260.02 ─ 08
FETCh:ACCuracy?.........................................................................................................206
FETCh:CURRent:DC?....................................................................................................206
FETCh:CURRent[:AC]?.................................................................................................. 206
FETCh:IMPedance?.......................................................................................................207
FETCh:VOLTage:DC?.................................................................................................... 207
FETCh:VOLTage:TRANsformer:SECondary?....................................................................207
FETCh:VOLTage:TRANsformer[:PRIMary]?......................................................................208
FETCh:VOLTage[:AC]?...................................................................................................208
FETCh[:TYPE]?.............................................................................................................208
FETCh:ACCuracy?
Queries the accuracy of the last measurement.
The R&S LCX returns the accuracy of the impedance (δ|Z|) in percent, and the phase
angle (δΘ) in degrees.
If an invalid value is returned, the instrument reports error code -230.
Return values:
<Magnitude Accuracy
Percentage>
Selects the magnitude accuracy.
<Phase Accuracy
Degree>
Selects the phase accurracy.
Example: See Example "Querying measurement values" on page 205.
Usage: Query only
Manual operation: See "<measurement accuracy>" on page 73
FETCh:CURRent:DC?
Applies to DC measurements only.
Queries the DC current measurement value of the last measurment.
If an invalid value is returned, the instrument reports error code -230.
Return values:
<DC Current
Measurement Value>
Default unit: A
Example: See Example "Querying measurement values" on page 205.
Usage: Query only
FETCh:CURRent[:AC]?
Applies to AC measurements.
Queries the AC current measurement value of the last measurement.
If an invalid value is returned, the instrument reports error code -230.
FETCh subsystem

Remote control commands
R&S
®
LCX Series
207User Manual 1179.2260.02 ─ 08
Return values:
<AC Current
Measurement Value>
Default unit: A
Example: See Example "Querying measurement values" on page 205.
Usage: Query only
FETCh:IMPedance?
Queries the recently measured impedance values Z and Θd.
If an invalid value is returned, the instrument reports error code -230.
Return values:
<Primary
Measurement Value>
Selects the impedance measurement parameter.
Default unit: Ohm
<Secondary
Measurement Value>
Selects the phase angle measurement parameter.
Example: See Example "Acquiring measurement readings" on page 205.
Usage: Query only
FETCh:VOLTage:DC?
Queries the most recent DC voltage measurement value.
If an invalid value is returned, the instrument reports error code -230.
Return values:
<DC Voltage
Measurement Value>
Default unit: V
Example: See Example "Querying measurement values" on page 205.
Usage: Query only
FETCh:VOLTage:TRANsformer:SECondary?
Queries the recently measured secondary transformer voltage value.
If an invalid value is returned, the instrument reports error code -230.
Return values:
<Secondary Voltage
Measurement Value>
Default unit: V
Example: See Example "Querying measurement values" on page 205.
Usage: Query only
FETCh subsystem

Remote control commands
R&S
®
LCX Series
208User Manual 1179.2260.02 ─ 08
FETCh:VOLTage:TRANsformer[:PRIMary]?
Queries the recently measured primary transformer voltage value.
If an invalid value is returned, the instrument reports error code -230.
Return values:
<Primary Voltage
Measurement Value>
Default unit: V
Example: See Example "Querying measurement values" on page 205.
Usage: Query only
FETCh:VOLTage[:AC]?
Queries the most recent AC voltage measurement value.
If an invalid value is returned, the instrument reports error code -230.
Return values:
<AC Voltage
Measurement Value>
Default unit: V
Example: See Example "Querying measurement values" on page 205.
Usage: Query only
FETCh[:TYPE]? [<Measurement Type Pair>]
Queries the recently measured values for the specified measurement pair.
Without specifying the measurement pair, the instrument returns the values of the pri-
mary measurement pair.
If an invalid value is returned, the instrument reports error code -230.
Query parameters:
<Measurement Type
Pair>
DEFault | CPD | CPQ | CPG | CPRP | CSD | CSQ | CSRS |
LPD | LPQ | LPG | LPRP | LSD | LSQ | LSRS | RX | RPB | RDC |
MTD | NTD | ZTD | ZTR | GB | YTD | YTR
Return values:
<Primary
Measurement Value>
Selects the first measurement parameter.
<Secondary
Measurement Value>
Selects the second measurement parameter.
Example: See Example "Acquiring measurement readings" on page 205.
Usage: Query only
FETCh subsystem