Remote control commands R&S ® LCX Series 201 User Manual 1 179.2260.02 ─ 08 DIMeasure:INT erval:STEPsize <Interval stepsize> Sets the step size within the measurement range for interval type DIMeasure: INTerval…

Remote control commands
R&S
®
LCX Series
200User Manual 1179.2260.02 ─ 08
*************************************************
// Cancel a running dynamic impedance measurement
DIMeasure:ABORt
DIMeasure:ABORt......................................................................................................... 200
DIMeasure:EXECute...................................................................................................... 200
DIMeasure:INTerval:POINts............................................................................................ 200
DIMeasure:INTerval:STEPsize.........................................................................................201
DIMeasure:INTerval:TYPE.............................................................................................. 201
DIMeasure:SWEep:MAXimum.........................................................................................201
DIMeasure:SWEep:MINimum..........................................................................................201
DIMeasure:SWEep:PARameter....................................................................................... 202
DIMeasure:ABORt
Stops a running dynamic impedance measurement.
Example:
See Example "Using the DIMeasure function" on page 199.
Usage: Event
Options: Option: R&S LCX-K106
Manual operation: See "Abort" on page 109
DIMeasure:EXECute
Starts the dynamic impedance measurement with the selected parameters.
Return values:
<DIM execution
state>
Example: See Example "Using the DIMeasure function" on page 199.
Options: Option: R&S LCX-K106
Manual operation: See "Measurement" on page 107
DIMeasure:INTerval:POINts <Interval points>
Sets the number of measurement points within the measurement range for interval
type DIMeasure:INTerval:TYPE > Number of Points.
Parameters:
<Interval points> Range: 1 to 65536 points
Example: See Example "Using the DIMeasure function" on page 199.
Options: Option: R&S LCX-K106
Manual operation: See "Number of Points" on page 108
DIMeasure subsystem

Remote control commands
R&S
®
LCX Series
201User Manual 1179.2260.02 ─ 08
DIMeasure:INTerval:STEPsize <Interval stepsize>
Sets the step size within the measurement range for interval type DIMeasure:
INTerval:TYPE > Step Size.
Parameters:
<Interval stepsize>
Example: See Example "Using the DIMeasure function" on page 199.
Options: Option: R&S LCX-K106
Manual operation: See "Step Size" on page 108
DIMeasure:INTerval:TYPE
<Interval parameter>
Selects the mode to determine the measurement steps within the sweep range
(DIMeasure:SWEep:MINimum to DIMeasure:SWEep:MAXimum).
Parameters:
<Interval parameter> STEPsize | POINts
STEPsize
Measures in defined step sizes within the sweep range, set with
DIMeasure:INTerval:STEPsize.
POINts
Measures in increments calculated by the number of sweep
points (DIMeasure:INTerval:POINts on page 200) within
the sweep range.
Example: See Example "Using the DIMeasure function" on page 199.
Options: Option: R&S LCX-K106
Manual operation: See "Interval Type" on page 108
DIMeasure:SWEep:MAXimum <Sweep stop value>
Sets the stop value for the selected sweep parameter.
The value must be at least > DIMeasure:SWEep:MINimum. The maximum value
depends on the instrument model and installed options.
Parameters:
<Sweep stop value>
Example: See Example "Using the DIMeasure function" on page 199.
Options: Option: R&S LCX-K106
Manual operation: See "Maximum" on page 108
DIMeasure:SWEep:MINimum <Sweep start value>
Sets the start value for the selected sweep parameter.
DIMeasure subsystem

Remote control commands
R&S
®
LCX Series
202User Manual 1179.2260.02 ─ 08
The value depends on the instrument model and installed options. The maximum value
must be at least < DIMeasure:SWEep:MAXimum.
Parameters:
<Sweep start value>
Example: See Example "Using the DIMeasure function" on page 199.
Options: Option: R&S LCX-K106
Manual operation: See "Minimum" on page 108
DIMeasure:SWEep:PARameter <Sweep parameter>
Selects the measurement parameter that varies in defined steps within the sweep
range (DIMeasure:SWEep:MINimum.
Parameters:
<Sweep parameter> VOLTage | FREQuency | VBIas | IBIas
VOLTage
Sweeps the test signal voltage.
FREQuency
Sweeps the test signal frequency.
VBIas
Sweeps the voltage bias.
IBIas
Sweeps the current bias.
Example: See Example "Using the DIMeasure function" on page 199.
Options: Option: R&S LCX-K106
Manual operation: See "Sweep Parameter" on page 107
11.9 DISPLay subsystem
The DISPlay subsystem contains the commands for customizing the display appear-
ance, and enables you to close messages.
DISPLay subsystem