LCX_UserManual_en_08_1.pdf - 第212页

Remote control commands R&S ® LCX Series 212 User Manual 1 179.2260.02 ─ 08 Example: See Example "Selecting the measurement function and setting the impedance range" on page 210. Manual operation: See &quot…

100%1 / 265
Remote control commands
R&S
®
LCX Series
211User Manual 1179.2260.02 ─ 08
FUNCtion:TRANsformer:RANGe[:TYPE] N500
FUNCtion:IMPedance:RANGe:AUTO............................................................................... 211
FUNCtion:IMPedance:RANGe:HOLD............................................................................... 211
FUNCtion:IMPedance:RANGe[:VALue].............................................................................211
FUNCtion:IMPedance:SOURce....................................................................................... 212
FUNCtion:IMPedance[:TYPE][:PRIMary].......................................................................... 212
FUNCtion:IMPedance:TYPE:SECondary..........................................................................213
FUNCtion:MEASurement:AUToselect...............................................................................213
FUNCtion:MEASurement:TYPE.......................................................................................214
FUNCtion:TRANsformer:RANGe[:TYPE].......................................................................... 214
FUNCtion:IMPedance:RANGe:AUTO <Auto Range>
Enables automatic impedance range selection.
To set the impedance range manually, use command FUNCtion:IMPedance:
RANGe[:VALue] on page 211.
Parameters:
<Auto Range> *RST: ON
Example: See Example "Selecting the measurement function and setting
the impedance range" on page 210.
Manual operation: See "Range: / [Range]" on page 75
FUNCtion:IMPedance:RANGe:HOLD <Locks selected range>
Freezes the set impedance measurement range.
Parameters:
<Locks selected
range>
*RST: OFF
Example: See Example "Selecting the measurement function and setting
the impedance range" on page 210.
Manual operation: See "[Hold]" on page 76
FUNCtion:IMPedance:RANGe[:VALue] <Range>
Sets the impedance range value.
The manual setting of the impedance range disables the automatic selection, if
enabled with FUNCtion:IMPedance:RANGe:AUTO on page 211.
Parameters:
<Range> Range: 0.3 Ohm to 30E+6 Ohm
*RST: 100 R
Default unit: Ohm
FUNCtion subsystem
Remote control commands
R&S
®
LCX Series
212User Manual 1179.2260.02 ─ 08
Example: See Example "Selecting the measurement function and setting
the impedance range" on page 210.
Manual operation: See "Range: / [Range]" on page 75
FUNCtion:IMPedance:SOURce <Impedance>
Selects the output impedance for the measurement.
Parameters:
<Impedance> LOW | R10 | HIGH | R100
LOW | R10
Sets 10 Ω output impedance.
HIGH | R100
Sets 100 Ω output impedance.
*RST: HIGH
Example: See Example "Selecting the measurement function and setting
the impedance range" on page 210.
Manual operation: See "Source Impedance" on page 80
FUNCtion:IMPedance[:TYPE][:PRIMary] <Impedance Type>
Selects the primary impedance parameter for the measurement corresponding to the
measurement type, see FUNCtion:MEASurement:TYPE on page 214.
Parameters:
<Impedance Type> DEFault | CPD | CPQ | CPG | CPRP | CSD | CSQ | CSRS |
LPD | LPQ | LPG | LPRP | LSD | LSQ | LSRS | RX | RPB | RDC |
MTD | NTD | ZTD | ZTR | GB | YTD | YTR
DEFault
Sets the default setting ZTD (impedance phase angle degree).
CPD | CPQ | CPG | CPRP | CSD | CSQ | CSRS
Capacitive measurement type: Cp (parallel capacitance), Cs
(serial capacitance), D (dissipation factor), Q (quality factor), G
(conductance), Rp (parallel resistance), Rs (serial resistance)
LPD | LPQ | LPG | LPRP | LSD | LSQ | LSRS
Inductive measurement type: Lp (parallel inductance), Ls (serial
inductance), D (dissipation factor), Q (quality factor), G (conduc-
tance), Rp (parallel resistance), Rs (serial resistance), RDc
(direct current resistance)
RX | RPB | RDC | MTD | NTD | ZTD | ZTR | GB | YTD | YTR
Resistance measurement type: R (resistance), X impedance, Rp
(parallel resistance), RDC (direct current resistance), B (suscep-
tance), M (mutual inductance), N (transformer ratio), Z (impe-
dance), G (conductance), Y (admittance), TD (phase angle
degree), TR (phase angle rad)
FUNCtion subsystem
Remote control commands
R&S
®
LCX Series
213User Manual 1179.2260.02 ─ 08
Example: See Example "Selecting the measurement function and setting
the impedance range" on page 210.
Manual operation: See "<measurement pair>" on page 73
FUNCtion:IMPedance:TYPE:SECondary <Impedance Type>
Selects the secondary impedance parameter for the measurement corresponding to
the measurement type, see FUNCtion:MEASurement:TYPE on page 214.
Parameters:
<Impedance Type> DEFault | CPD | CPQ | CPG | CPRP | CSD | CSQ | CSRS |
LPD | LPQ | LPG | LPRP | LSD | LSQ | LSRS | RX | RPB | RDC |
MTD | NTD | ZTD | ZTR | GB | YTD | YTR
DEFault
Sets the default setting ZTD (impedance phase angle degree).
CPD | CPQ | CPG | CPRP | CSD | CSQ | CSRS
Capacitive measurement type: Cp (parallel capacitance), Cs
(serial capacitance), D (dissipation factor), Q (quality factor), G
(conductance), Rp (parallel resistance), Rs (serial resistance)
LPD | LPQ | LPG | LPRP | LSD | LSQ | LSRS
Inductive measurement type: Lp (parallel inductance), Ls (serial
inductance), D (dissipation factor), Q (quality factor), G (conduc-
tance), Rp (parallel resistance), Rs (serial resistance), RDc
(direct current resistance)
RX | RPB | RDC | MTD | NTD | ZTD | ZTR | GB | YTD | YTR
Resistance measurement type: R (resistance), X impedance, Rp
(parallel resistance), RDC (direct current resistance), B (suscep-
tance), M (mutual inductance), N (transformer ratio), Z (impe-
dance), G (conductance), Y (admittance), TD (phase angle
degree), TR (phase angle rad)
Example:
See Example "Selecting the measurement function and setting
the impedance range" on page 210.
Manual operation: See "<measurement pair>" on page 73
FUNCtion:MEASurement:AUToselect
Triggers the automatic selection of the measurement function and the corresponding
default measurement pairs.
Example:
See Example "Selecting the measurement function and setting
the impedance range" on page 210.
Usage: Event
Asynchronous command
Manual operation: See "[Auto]" on page 76
FUNCtion subsystem