Remote control commands R&S ® LCX Series 226 User Manual 1 179.2260.02 ─ 08 MEASure[:TYPE]? [<Measurement T ype Pair>] T riggers a new measurement and queries the measured values of the specified mea- surement …

100%1 / 265
Remote control commands
R&S
®
LCX Series
225User Manual 1179.2260.02 ─ 08
Return values:
<Primary
Measurement Value>
Selects the first measurement parameter.
<Secondary
Measurement Value>
Selects the second measurement parameter.
Example: See Example "Querying measurement results" on page 223.
Usage: Query only
MEASure:R? [<Resistance Type Pair>]
Triggers a new DC resistance measurement, reads the resistance value and returns
the result of the specified measurement pair. Without specifying the measurement
parameters, the instrument returns the value of the default serial resistance and reac-
tance RX.
Query parameters:
<Resistance Type
Pair>
DEFault | RX | RPB | RDC | GB | ZTR | ZTD | YTR | YTD
Return values:
<Primary
Measurement Value>
Selects the first measurement parameter.
<Secondary
Measurement Value>
Selects the second measurement parameter.
Usage: Query only
MEASure:T? [<Transformer Type Pair>]
Triggers a new transformer measurement, reads the resistance value and returns the
result of the specified measurement pair. Without specifying the measurement parame-
ters, the instrument returns the value of the default transformer ratio and phase angle
NTD.
Query parameters:
<Transformer Type
Pair>
DEFault | MTD | NTD | RX | GB | ZTR | ZTD | YTR | YTD
Return values:
<Primary
Measurement Value>
Selects the first measurement parameter.
<Secondary
Measurement Value>
Selects the second measurement parameter.
Example: See Example "Querying measurement results" on page 223.
Usage: Query only
MEASure subsystem
Remote control commands
R&S
®
LCX Series
226User Manual 1179.2260.02 ─ 08
MEASure[:TYPE]? [<Measurement Type Pair>]
Triggers a new measurement and queries the measured values of the specified mea-
surement pair. Without specifying the measurement parameter, the instrument returns
the value of the default impedance measurement ZTD (impedance and phase angle).
Query parameters:
<Measurement Type
Pair>
DEFault | CPD | CPQ | CPG | CPRP | CSD | CSQ | CSRS |
LPD | LPQ | LPG | LPRP | LSD | LSQ | LSRS | RX | RPB | RDC |
MTD | NTD | ZTD | ZTR | GB | YTD | YTR
Return values:
<Primary
Measurement Value>
Selects the first measurement parameter.
<Secondary
Measurement Value>
Selects the second measurement parameter.
Example: See Example "Querying measurement results" on page 223.
Usage: Query only
11.16 READ subsystem
This section describes the remote control commands to read the results of a measure-
ment. These commands are similar to the corresponding commands of the MEASure
subsytem.
READ subsystem
Remote control commands
R&S
®
LCX Series
227User Manual 1179.2260.02 ─ 08
Example: Reading measurement results
The example represents the remote control commands for triggering and querying
measurement pair readings.
*************************************************
// Request information on the measurement result
// query recently measured impedance values
READ:IMPedance?
// Response: returns Z and Theta d values
// query the recently measured values of measurement pair 2
READ[:TYPe]? CPRP
// Response: returns the measurement values of the
// parallel capacitance and parallel resistance
// parameter pair
READ[:TYPe]?
// Response: returns the measurement values of the
// primary parameter pair
READ:IMPedance?........................................................................................................ 227
READ[:TYPE]?.............................................................................................................. 227
READ:IMPedance?
Queries the impedance measurement results.
Return values:
<Impedance
Measurement Value>
Selects the impedance measurement parameter.
Default unit: Ohm
<Phase Measurement
Value>
Selects the phase angle measurement parameter.
Example: See Example "Reading measurement results" on page 227.
Usage: Query only
READ[:TYPE]? [<Impedance Type Pair>]
Triggers a new measurement and queries the measured values of the specified mea-
surement pair.
Without specifying the measurement parameter, the instrument returns the value of the
first measurement pair.
Query parameters:
<Impedance Type
Pair>
DEFault | CPD | CPQ | CPG | CPRP | CSD | CSQ | CSRS |
LPD | LPQ | LPG | LPRP | LSD | LSQ | LSRS | RX | RPB | RDC |
MTD | NTD | ZTD | ZTR | GB | YTD | YTR
READ subsystem