Instrument functions R&S ® LCX Series 1 13 User Manual 1 179.2260.02 ─ 08 T able 8-2: Pin assignment Pin Signal Direction Pin Signal Direction 1 - 8 Digital output sig- nals 1 to 8. The digital outputs are active low…

Instrument functions
R&S
®
LCX Series
112User Manual 1179.2260.02 ─ 08
Contents of the *.csv file:
●
Up to 2 measurement parameters
●
The measured value range of each of the two parameters for the individual bin.
●
The value ranges of the measurement parameters defined for the corresponding
categories (up to 8 bins).
●
How the limits are specified:
– Minimum and maximum values
– Nominal value and percentage deviation
– Nominal and deviation value
●
Additional signals you want to have assigned at the output, i.e. end of measure-
ment, out of bin, both).
With the binning function enabled, the R&S LCX switches to measurement mode "trig-
gered" automatically. The measurement starts with the first trigger signal. You can
either trigger the measurement manually, using the [Trig.] key at the front or by sending
the remote command *TRG. It is also possible to apply an external trigger signal to the
"Trigger Input" connector at the rear of the instrument.
During the binning measurements, the R&S LCX is permanently in the triggered mode
and cannot be disabled.
How to: see Section 8.8.5.5, "Configuring and using the binning function",
on page 116.
For details on the trigger modes, see the specifications document, section "Digital trig-
ger and control interfaces".
8.8.5.2 Assignment of the binning interface
The digital I/O port of the binning interface is a 15-pole D-Sub female socket, including
eight data lines. The remaining pins provide digital input and output of control signals.
Figure 8-26: Digital I/O connector, front view
For details on the interface specification, see section "Digital trigger and control interfa-
ces" in the specifications document.
Specific instrument functions

Instrument functions
R&S
®
LCX Series
113User Manual 1179.2260.02 ─ 08
Table 8-2: Pin assignment
Pin Signal Direction Pin Signal Direction
1 - 8 Digital output sig-
nals 1 to 8.
The digital outputs
are active low.
OUT 11 - 13 GND OUT
9 Trigger signal input.
The falling edge of
an externally
applied signal ini-
tiates the trigger.
IN 14 + 3.3 V OUT
10 Digital input IN 15 + 12 V OUT
8.8.5.3 Binning settings
In the "Binning" menu, you can upload the binning file, and start the measurement.
Access:
1. Press the [settings] key at the front panel.
2. In the "Functions" tab, select "Binning".
Figure 8-27: Binning menu
In the initial softkey menu, you can load the binning file and activate the measure-
ment.
During the process, the main screen indicates the parameters related to the mea-
surement, and access to the bin configuration.
The required remote commands are described in Section 11.12, "HANDler subsystem",
on page 214.
Enabled....................................................................................................................... 113
Load From File............................................................................................................ 113
Enabled
Enables the sorting function.
The measurement starts with a trigger signal.
Remote command:
HANDler[:STATe] on page 216
Load From File
Accesses the file manager function to upload the binning configuration file.
Specific instrument functions

Instrument functions
R&S
®
LCX Series
114User Manual 1179.2260.02 ─ 08
Remote command:
HANDler:CONFig:PATH on page 216
8.8.5.4 Binning measurement settings
When you have enabled the binning measurement, the R&S LCX shows the binning
specific measurement results window on the screen.
Access:
1. Press the [settings] key at the front panel.
2. In the "Functions" tab, select "Binning".
3. Load the corresponding binning file as described in "Configuring a binning file"
on page 116 and "Loading the binning file" on page 117.
Figure 8-28: Binning measurement results window
This view shows the configuration parameters of the measurement and the bins for
sorting the measured components. A bar graph displays the distribution of the
measurement results graphically.
You can access the settings or reset the counter on the screen directly.
Binning.........................................................................................................................114
Settings softkey........................................................................................................... 115
<filename>...................................................................................................................115
Counter........................................................................................................................115
Primary - <param>.......................................................................................................115
Secondary - <param>..................................................................................................115
Bins..............................................................................................................................115
└ EOM..............................................................................................................115
└ OUT...............................................................................................................115
└ --....................................................................................................................115
└ B1 to B6 (B8).................................................................................................115
Histogram.................................................................................................................... 115
Binning
Indicates the measurement mode.
Specific instrument functions