LCX_UserManual_en_08_1.pdf - 第130页

General instrument functions R&S ® LCX Series 130 User Manual 1 179.2260.02 ─ 08 Assigning the function to the "User Button" T o assign the screenshot function to the "User Button": 1. Press the […

100%1 / 265
General instrument functions
R&S
®
LCX Series
129User Manual 1179.2260.02 ─ 08
Access:
1. Press the [settings] key at the front panel.
2. In the "Device" tab, select "Screenshot".
Figure 9-7: Screenshot
The "Screenshot" menu provides access to dialogs and editors for managing and
saving instrument settings.
Save Location............................................................................................................. 129
Save Location
Opens a dialog where you can select how to determine the location for the file.
"Auto"
Selects the target folder automatically.
"Manual"
Enables you to select and define the target folder on the USB flash
drive.
9.4.2 Using the screenshot function
A screenshot is done with the [ (User)] key, provided the function is assigned to the
key.
The following instructions show how to configure function for use, and how to save a
captured image, e.g., in a user-defined target folder.
Without a USB memory stick connected, the R&S LCX saves the file internally under /
int/screenshot.
See Section 9.1, "File and data management", on page 120 for information on how to
access the memory and manage file export.
Screenshot
General instrument functions
R&S
®
LCX Series
130User Manual 1179.2260.02 ─ 08
Assigning the function to the "User Button"
To assign the screenshot function to the "User Button":
1. Press the [settings] key at the front panel.
2. In the "Device" tab, select "User Button".
3. Select "User Button Action".
4. Select "Screenshot".
5.
Return to the "Device" settings tab with
.
The (User) now captures the screen when pressed.
Setting the screenshot file location manually
To determine a user-defined target folder for the screenshot files, the USB memory
stick must be connected:
1. Connect the USB memory stick to the USB interface at the front panel of the
R&S LCX.
2. Press the [settings] key at the front panel.
3. In the "Device" tab, select "Screenshot".
4. Set "Save Location" > "Manual".
5. Select "Target Folder".
6. In the selection editor, select the memory location, e.g. "USB1A".
7. "Close" the window.
The R&S LCX saves the hardcopy file in the predefined screenshots folder on the
selected memory location.
Dragging screenshots
To capture the current screen, e.g., a measurement result:
1. Press the (User) at the front panel.
The R&S LCX drags a screenshot, generates a filename and saves the file in the
target folder, as defined.
When completed, it shows the "Screenshot" dialog with information on file path and
filename.
2. Confirm with "OK".
The R&S LCX returns to the measurement.
For further use of the screenshots, you can open the files in any application that
supports *.png file format.
For information on how to access the memory and manage file export, see Sec-
tion 9.1, "File and data management", on page 120.
Screenshot
General instrument functions
R&S
®
LCX Series
131User Manual 1179.2260.02 ─ 08
9.5 CSV settings
The R&S LCX enables you to record measurement readings and save the data in a file
for subsequent evaluation and validation. Recorded data consists of character strings
that list general metadata, as, e.g., date, device name, etc., the parameter names and
the acquired measurement values.
The standard *.csv file format the R&S LCX saves the data, allows you to use a
spreadsheet program application for numerical or graphical analysis and evaluation.
A *.csv file saves tabular data (numbers and text) in plain text. Each line of the file is
a data record and each record consists of one or more fields, separated by a field
delimiter. In the "CSV Settings" dialog, you can set the special characters to set the
delimiters.
9.5.1 Settings for CSV data
Access:
1. Press the [settings] key at the front panel.
2. In the "Device" tab, select "CSV Settings".
Figure 9-8: CSV settings menu
The dialog provides the parameters for setting the data separators in the *.csv
file. The error designator enables you to fill up empty cells.
Field Delimiter............................................................................................................. 131
Decimal Separator...................................................................................................... 131
Error Designator..........................................................................................................132
Line End Marker..........................................................................................................132
Field Delimiter
Selects the line separator between individual parameters or values.
"Semicolon"
Separates the data entries by a semicolon.
"Comma"
Uses a comma.
Decimal Separator
Selects the decimal separator for values.
CSV settings