2-26 2 Basic operation 8 Secure the syringe in the head. While holding the lock nut by hand, turn the adapter holder in the direction of arrow to clamp the syringe into position in the head. Clamping the syringe Lock nut…

2-25
2
Basic operation
5
Attach a nozzle to the nozzle joint.
Use a hex wrench (1.5) to attach the nozzle to the nozzle joint.
6
Attach a nozzle to each nozzle.
1. Hold the nozzle directly by hand or place the nozzle on the supplied air tube piece with the ring nut
passed along it as shown below.
2. Align the notch of the nozzle with the guide on the head shaft, then insert the nozzle into the head
shaft from the bottom.
3. Tighten the ring nut to secure the nozzle while pressing up the air tube against the nozzle so that it
does not drop.
Attaching a nozzle
Nozzle
Ring nut
Air tube piece (supplied)
63203-N7-00
7
Set the syringe.
1. Before installing the syringe in the head, attach the syringe adapter to the top of the syringe.
2. Insert the syringe into the adapter holder as shown.
3. Insert the syringe into the head.
4. Insert the air hose into the air fitting.
Setting the syringe in the adapter holder
Syringe adapter
Adapter holder
Syringe
Air hose
Air fitting
63204-N7-10
c
CAUTION
When a low-liquid sensor is used, if the product label affixed to the syringe is large, then peel if off at least the half from
the syringe tip. If the product label covers the lower part of the syringe where the low-liquid sensor is aligned, the
sensor cannot detect whether the liquid has run out or not.

2-26
2
Basic operation
8
Secure the syringe in the head.
While holding the lock nut by hand, turn the adapter holder in the direction of arrow to clamp the
syringe into position in the head.
Clamping the syringe
Lock nut
Heater
Adapter holder
Muffler
63205-N7-10
c
CAUTION
• The heater in the head may be hot, so use caution.
• The muffler is likely to break, so do not touch it.
9
Cancel emergency stop.
Close the safety cover, release the emergency stop button, and press the [READY] button.
0
Press the [Close] button to close the "Cylinder exchange" window.
You will return to the Setup screen.
n
NOTE
After the syringe of a head is replaced with a new syringe, press the [Init Dispense] button for the applicable head to
reset the liquid amount count value.

2-27
2
Basic operation
3.5 Starting and warming up the machine
After setting up the head unit, perform warm-up operation. A warm-up of about 10 minutes is usually adequate.
If less than two hours has elapsed after the machine was last used for production, no warm-up is necessary.
1
Press the [Warm Up] button on the Setup screen.
The Warm-up screen then appears.
2
Enter a warm-up time.
Press the [Stop after time over] button and enter the warm-up time. Warm-up will automatically stop
when the warm-up time you set has elapsed. Normally, about a 10-minute warm-up is recommended.
Warm-up time setting screen
Press [Warm Up] button to open "Warm Up" screen.
Setup screen Press this button to set warmup time.
64207-N7-00