11.2.11 LP- Mapping
Choose "LP- Mapping..."
Fig. 11.2 - 12 Menu LP- Mapping
Description for LP- Mapping see chapter 11.2.15
11.2.12 Head- Mapping
Choose "Head- Mapping..."
Fig. 11.2 - 13 Menu Head- Mapping
Description for Head- Mapping see chapter 11.2.16
1 - 25
Student Guide SIPLACE X
Edition 09/2005 11 Sitest
25
11.2.13 Basically description of all calibration steps
Machine zero point: 11
The PCB-camera center is the reference at the gantry. All positions at the incremental encoder
of X/Y-axis refer to this camera center.
A drilling is optically centered with the PCB-camera on a defined position at the calibration tool
support.
Then the Zero point correction of the gantry axes are changed that if the PCB camera is above
this hole and the Positioncounter shows exactly the value of:
MA nullpunkt_x_PG1 631300 / MA nullpunkt_y_PG1 1298000. ( pg means gantry group)
MA nullpunkt_x_PG2 1368700 / MA nullpunkt_y_PG2 702000 ( See Fig. 11.2 - 2)
PCB camera: 11
The Pixel size of the CCD sensors of the camera is determined in µm. Measured and calcu-
lated with Ax/Bx/Ay/By calibration values. Saved in KAM_DAT.MA as:XU_Pixel / YU_Pixel
(in 11600 nm Standard-PCB camera SST 5), (in 9900 nm Multicolorillum. PCB camera SST 18)
The camera center is determined.
This camera center is now the reference for all Position and Offsets of the gantry!!
The Mounting angle of the CCD-chip in the camera to the Ma-coordinate system. Saved as
‘Kamera_winkel’ at the Data bloc of the PCB camera in KAMDAT.MA.
Calibration tool position:(optional) 11
Calibrate the X and Y pick up position of the calibration tool.
Travel range:(optional) 11
For travel range calibration move the respective axis: -to the Zero-pulse
Then to the Hardware-limit switch
Measure the position value at the position counter
Calculate the position for SW-limit switch (Y +/- 1.5 mm , X +/- 0,5 mm)
65mm
Turning for reference run
speed monitoring
Speed monitoring
HW travel range
HW travel range
Y-Axis
45mm
Turning for re fere nc e run
Speed monitoring
HW tra vel ra nge
HW tra vel ra ng e
Speed monitoring
X-Axis
1 - 26
Student Guide SIPLACE X
11 Sitest Edition 09/2005
26
Component camera: 11
The Pixel size of the CCD sensors of the camera is determined in µm. Measured and calcu-
lated with Ax/Bx/Cx/Ay/ByCy calibration values.Saved in KAM_DAT.MA as:XU_Pixel /
YU_Pixel (50000 nm 12 nozzle-comp.-camera SST 12) (81000 nm 6 nozzle-comp.-camera
SST 13) (27500 nm DCA-camera SST14)
The camera center is determined.
The Mounting angle of the CCD-chip in the camera to the turning level of the placement star
is measured.Saved as ‘Kamera_winkel’ at the Data bloc of the respective comp. camera in KA-
MDAT.MA.
Sequence segment offset bottom (II): 11
Sequence at one nozzle:
Is the calibration tool picked with a Nozzle under 0 degree; optically centered and placed with
the PCB-camera is the exact placement position determined (in µm).
Is the calibration tool picked with a Nozzle under 90 degree; optically centered and placed with
the PCB-camera is the exact placement position determined (in µm).
Is the calibration tool picked with a Nozzle under 180 degree; optically centered and placed
with the PCB-camera is the exact placement position determined (in µm).
Is the calibration tool picked with a Nozzle under 270 degree; optically centered and placed
with the PCB-camera is the exact placement position determined (in µm).
This sequence is repeated. From the 8 placement positions is the average value of the place.
Deviation calculated and taken for the seg. offset.
The calibration tool do not turn at this sequence.