249 E690023 “Somethi ng is w rong w ith the measured volume” [Cause] T he condens er capac ity gotten fr om the tester is irregular. • The objec t(com ponent) of the m easurem ent is not contacting the pr obe well. • Com…

100%1 / 1361
248
E690018 “A time-out occurred while reading voltage in”
[Cause] The tester does not start outputting data when starting to measure the voltage.
Component Verification is broken.
[Treatment] Please turn off the power supply and re-boot your machine.
If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690019 “The measured resistance does not get stable. | Please check the
probe status”
[Cause] The resistance gotten from the tester does not get stable.
The object(component) of the measurement is not contacting the probe well.
Component Verification is broken.
[Treatment] Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690020 “The measured volume does not get stable. | Please check the
probe status”
[Cause] The condenser capacity gotten from the tester does not get stable.
The object(component) of the measurement is not contacting the probe well.
Component Verification is broken.
[Treatment] Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690021 “The measured voltage does not get stable. | Please check the
probe status”
[Cause] The voltage gotten from the tester does not get stable.
The object(component) of the measurement is not contacting the probe well.
Component Verification is broken.
[Treatment] Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690022 “Something is wrong with the measured resistance”
[Cause] The resistance gotten from the tester is irregular.
The object(component) of the measurement is not contacting the probe well.
Component Verification is broken.
[Treatment] Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
249
E690023 “Something is wrong with the measured volume”
[Cause] The condenser capacity gotten from the tester is irregular.
The object(component) of the measurement is not contacting the probe well.
Component Verification is broken.
[Treatment] Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690024 “Something is wrong with the measured voltage”
[Cause] The voltage gotten from the tester is irregular.
The object(component) of the measurement is not contacting the probe well.
Component Verification is broken.
[Treatment] Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690025 “Something is wrong with the measured resistance(unit)”
[Cause] The unit for the resistance gotten from the tester is irregular.
The object(component) of the measurement is not contacting the probe well.
Component Verification is broken.
[Treatment] Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690026 “Something is wrong with the measured volume(unit)”
[Cause] The unit for the condenser capacity gotten from the tester is irregular.
The object(component) of the measurement is not contacting the probe well.
Component Verification is broken.
[Treatment] Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690027 “Something is wrong with the measured voltage(unit)
[Cause] The unit for the condenser capacity gotten from the tester is irregular.
The object(component) of the measurement is not contacting the probe well.
Component Verification is broken.
[Treatment] Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
250
E690028 “Resistance can no be measured. Please check if probe shuts
down properly”
[Cause] The tester overflowed data when measuring the resistance.
The object(component) of the measurement is not contacting the probe.
The contact between the object(component) of the measurement and the
probe is not good.
The resistance of the object(component) of the measurement is exceeding the
measurement range.
The Component Verification is broken.
[Treatment] Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
Please check the object.
If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690029 “Volume can not be measured. Please check if probe shuts down
properly”
[Cause] The tester overflowed data when measuring the condenser capacity.
The object(component) of the measurement is not contacting the probe.
The contact between the object(component) of the measurement and the
probe is not good.
The resistance of the object(component) of the measurement is exceeding the
measurement range.
The Component Verification is broken.
[Treatment] Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
Please check the object.
If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690030 “Voltage can not be measured. Please check if probe shuts down
properly”
[Cause] The tester overflowed data when measuring the diode polarity.
The object(component) of the measurement is not contacting the probe.
The contact between the object(component) of the measurement and the
probe is not good.
The resistance of the object(component) of the measurement is exceeding the
measurement range.
The Component Verification is broken.
[Treatment] Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
Please check the object.
If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E691001 “Input/Output buffer status of CCL-48M does not get ready”
[Cause] The input/output buffer status of the backup table counter(CCL-48M) does not get
ready.
Backup table counter is broken.
[Treatment] Please turn off the power supply and re-boot your machine.
If the error occurs even after the re-boot, there is a possibility that the Safety
board is broken. Please contact our technical service dept. for information.