MR8740、MR8741_user_manual_eng_20191016H.pdf - 第104页
4.2 Saving Data 92 Save Operations (When setting the save destination to [LAN]) _________ The file is saved accord ing to the method (on th e right fig- ure) specified under [ Save Settings] by using Model 9333 LAN Commu…

4.2 Saving Data
91
4
Chapter 4 Saving/Loading Data & Managing Files
Auto Save Operations (When setting the save destination to the recording media)
Save To:
USB:\HIOKI_MR
8740
Save method:
Normal Save
Folder to
save:
No
HIOKI_MR8740
0000AUTO.MEM
0001AUTO.MEM
0001HIOKI_MR8740
(A new folder is created automatically when 5,000 files have
been saved in the HIOKI_MR8740 folder.)
Up to 5,000 files
Example 1: Saving Files to the Topmost Directory of the Storage Media
(A folder named "HIOKI_MR8740 " is created and file is saved there)
0000AUTO.MEM
When the storage media becomes full:
Automatic saving stops.
USB
File name will be "4-digit number
+ file name".
Create a folder named "TEST" on the USB mem-
ory stick beforehand.)
Save To:
USB:\HIOKI_MR
8740 \TEST
Save method:
Normal Save
Folder to
save:
No
HIOKI_MR8740
TEST
0000AUTO.MEM
0001AUTO.MEM
0001TEST
(A new folder is created automatically when 5,000 files have
been saved in the TEST folder.)
Up to 5,000 files
Example 2: Saving Files to a Folder on Storage Media
0000AUTO.MEM
When the storage media becomes full: Automatic saving stops.
Folder name created automatically will be
"4-digit number + folder name".
Save To:
USB:\HIOKI_MR
8740
Save method:
Normal Save
Folder to
save:
Yes
HIOKI_MR8740
AUTO014530
0000AUTO.MEM
0001AUTO.MEM
0001AUTO014530
(A new folder is created automatically when 5,000 files have been
saved in the AUTO014530 folder.)
Up to 5,000 files
Example 3: Create folder on media automatically and save data there
0000AUTO.MEM
"AUTO + hours/minutes/seconds" is created
automatically. (Example: 1:45:30)
Folder name created next will be "4-dig-
it number + folder name".
When the storage media becomes full:
Automatic saving stop.
Save To:
USB:\HIOKI_MR
8740 or specified
folder in
"HIOKI_MR8740"
folder
Save method:
Delete Save
Folder to
save:
No
HIOKI_MR8740
AUTO014530
0000AUTO.MEM
0001AUTO.MEM
Up to 5,000 files
Example 4: Using Delete/Save to Automatically Save Data
Folder name will be "AUTO + hours/min-
utes/seconds". (Example: 1:45:30)
When the number of files in the folder has reached 5,000, or when the storage media has become full, files in the
AUTO014530 folder (or specified folder) will automatically be deleted in sequence, starting from the oldest file, and
replaced by new files. (Only waveform files are deleted.)

4.2 Saving Data
92
Save Operations (When setting the save destination to [LAN]) _________
The file is saved according to the method (on the right fig-
ure) specified under [Save Settings] by using Model 9333
LAN Communicator.
For more information about how to configure the save set-
tings, see the instruction manual of Model 9333 LAN Com-
municator.
The destination folders and files are named in the previ-
ously-determined formats as follows:
Create a folder named "TEST" on the
USB memory stick beforehand.)
Save To:
USB:\HIOKI_MR
8740\TEST
Save method:
Normal Save
Folder to
save:
No
HIOKI_MR8740
TEST
0000AUTO.IDX
0001AUTO.MEM
0001TEST
Up to 5,000 files
Example 5: Saving data in an existing folder on the media using divided saving
0000AUTO.IDX
When the storage media becomes full:
Automatic saving stops.
0000AUTO
0001AUTO
Up to 5,000 folders
0000AUTO.MEM
0000AUTO.MEM
Folder name created au-
tomatically will be "4-digit
number + folder name".
0000AUTO
(A new folder is created automatically when 5,000 folders have
been saved in the TEST folder.)
Save To:
USB:\HIOKI_M
R8740
Save method:
Normal Save
Folder to
save:
Yes
HIOKI_MR8740
AUTO014530
0000AUTO.IDX
0001AUTO.MEM
0001AUTO014530
(A new folder is created automatically when 5,000 folders have been
saved in the 0001AUTO folder.)
Up to 5,000 files
Example 6: Saving data in an automatically created folder on the media using divided saving
0000AUTO.IDX
When the storage media becomes full:
Automatic saving stop.
0000AUTO
0001AUTO
Up to 5,000 folders
0000AUTO.MEM
0000AUTO.MEM
"AUTO + hours/minutes/seconds" is
created automatically. (Example: 1:45:30)
Folder name created next will be "4-
digit number + folder name".
0000AUTO
192.168.1.1 14-01-10
14-01-11
HIOKI 9333
Folder name will be
"WAVE + hours/
minutes/seconds".
14-01-20
14-01-21
Folder name will be
"WAVE + hours/
minutes/seconds".
WAVE124012.BMP
Numerical calculation files
PARAMETER.TXT
PARAMETER_1.TX
WAVE123015.MEM
WAVE123245.MEM
WAVE123245.TXT
Specified
folder
Folder name
will be Model
MR9947 IP
address.
File names will be
"PRAMETER_ +
number".
Folder names
will be "YY-
MM-DD".
Waveform files

4.2 Saving Data
93
4
Chapter 4 Saving/Loading Data & Managing Files
To perform quick saving by clicking [SAVE] in the right-click menu, you need to set the saving con-
ditions beforehand. The type of data to be saved are as following. (Settings data, waveform data,
display screens, waveform screen, numerical calculation results)
4.2.3 Saving Data Selectively (SAVE)
Procedure
To open the screen: Right-click and select [SYSTEM] [File Save] sheet
1
Set the save method for when [SAVE] is clicked.
Move the flashing cursor to the [Select at save] item, and select
[No].
Select
When [Yes] is selected, the following setting determines which dialog box appears
when clicking [SAVE]. (Middle screen at right)
However, this is not applicable when other dialogs such as “Folder Reference Di-
alog” are shown on the screen.
2
Set the save destination.
Move the flashing cursor to the [Save To], and select [Edit].
The Browse folders dialog box appears (at bottom right).
Move the flashing cursor to the save target media
*
and confirm
the setting with [Confirm].
Select
When the root directory (topmost folder on media) was selected, a folder named
"HIOKI_MR8740" or "HIOKI_MR8741" is created automatically. (If the media was
initialized in the instrument, the folder will already have been created.) This folder
is then used as save target.
To create a new folder, select [New Folder]. When [LAN] is set
as the save destination,
[New Folder] setting is ignored and the
folder named with the current date is created.
3
Set the file name.
Move the flashing cursor to the [Save Name], and enter a save
name.
See: "7.1.3 Alphanumeric Input" ( p.141)
When [LAN] is set as the save destination, [Save Name] setting is
ignored and the files to be saved are named in the previously-
determined format.
See:
"Save Operations (When setting the save destination to [LAN])"
(p.92)
Yes After clicking [SAVE], set the data to save in the dialog box, then
save the data. (default setting)
See: "Selection Save"(p.87)
No
The preset data is saved upon clicking [SAVE].
See: "Quick Save"(p.87)
USB1 Automatically save the waveform data on the USB memory stick.
LAN
Automatically save the waveform data on the PC connected via
LAN. Model 9333 LAN Communicator is required to be installed.
1
2
3
File Name
The maximum number of characters for the [Save Name] string is 123. The