12.3 Setting FFT Analysis Conditio ns 251 11 Chapter 12 FFT Function 12 About the frequency range a nd number of analysis point s ___________________ • The settings for the frequency r ange and number of ana lysis points…

100%1 / 428
12.3 Setting FFT Analysis Conditions
250
F
Select the data to be used for FFT analysis.
There are two analysis methods: analysis using new measurements and analysis of data measured
using the memory function.
12.3.2 Selecting the Data Source for Analysis
1
Select the input data source.
Move the flashing cursor to the
[Reference] item.
Select
2
When finished making settings, click [START].
For the [New Data] case
Measurement starts to acquire data for the number of analysis points
specified as the [Sampling Point], and FFT analysis is performed.
For the [From Memory] case
Analysis is performed on the number of specified points from data
previously recorded in memory (Memory function data) .
The analysis starting point can also be specified.
See: "12.7.1 Analyzing after Specifying an Analysis Starting Point" (p.273)
The frequency range is selected automatically.
See: "Relationship Between Frequency Range, Resolution and Number of
Analysis Points" (p.252)
New Data Acquire a new waveform for analysis.
From Memory
Calculates data measured using the memory function.
Procedure
To open the screen: Right-click and select [STATUS] [Status] sheet
See: To set from the Waveform screen (p.265)
When the input data [Reference] is
[From Memory]
Analysis is performed until the specified
number of FFT analysis points have been
processed, then the data is shifted by that
amount and analysis repeats until all of the
previously acquired data has been pro-
cessed. (If the amount of data is less than
the specified number of FFT analysis
points, no analysis occurs.)
When no trace is displayed after start-
ing measurement
Analysis is impossible if [From Memory]
is selected as the input data source and
no recorded data exists in the instrument's
memory.
Either select [New Data] as the input data
source, or load the data to be analyzed
before clicking [START] again.
1
When the [Reference] is set to [From Memory], the frequency is automatically
set.
The setting cannot be changed.
12.3 Setting FFT Analysis Conditions
251
11
Chapter 12 FFT Function
12
About the frequency range and number of analysis points ___________________
The settings for the frequency range and number of analysis points determine
the input signal acquisition time and frequency resolution.
The frequency range setting for the FFT function corresponds to the timebase
setting of the Memory function. Changing the frequency range also changes
the data sampling period.
See: "Relationship Between Frequency Range, Resolution and Number of Analysis
Points" (p.252)
The cut-off frequency of the anti-aliasing filter is the same as the frequency
range setting.
See: "Relationship Between Frequency Range, Resolution and Number of Analysis
Points" (p.252)
The set number of analysis points specifies the amount of data to be analyzed
with each measurement. Increasing the number of analysis points increases
the frequency resolution, but also increases the time required for calculations.
See: "Number of Analysis Points" (p.A15)
12.3.3 Setting the Frequency Range and Number of
Analysis Points
When the [Reference] is [From Memory], the frequency is automatically
set.
The setting cannot be changed.
To control sampling by an external signal,
select [External]
When the input data [Reference] is
[From Memory]
The frequency range is set automatically
when analysis is started.
1
Set the number of FFT analysis points.
Move the flashing cursor to the [Sampling Point] item.
Select
See: "Number of Analysis Points" (p.A15)
2
Select the frequency range.
Move the flashing cursor to the [Frequency Range] item.
Select
See: "Relationship Between Frequency Range, Resolution and Number of
Analysis Points" (p.252)
1000 (default setting), 2000, 5000, 10000
8 MHz(default setting), 4 MHz, 2 MHz, 800 kHz, 400 kHz, 200 kHz, 80
kHz, 40 kHz, 20 kHz, 8 kHz, 4 kHz , 2 kHz, 800 Hz, 400 Hz, 200 Hz, 80
Hz, 40 Hz, 20 Hz, 8 Hz, 4 Hz, 1.33 Hz, 800 mHz, 667 mHz, 400 mHz, 333
mHz, 133 mHz, External
Procedure
To open the screen: Right-click and select [STATUS] [Status] sheet
See: To set from the Waveform screen (p.265)
Frequency Resolution
(during acquisition)
The resolution is affected by settings of
frequency range and the number of
analysis points. Not displayed for exter-
nal sampling.
2
1
12.3 Setting FFT Analysis Conditions
252
Relationship Between Frequency Range, Resolution and Number of
Analysis Points
Range
[Hz]
Sampling
frequency
[Hz]
Timebase
[/div]
(MEM)
Sampling
period
Number of FFT Analysis Points
1,000 2,000 5,000 10,000
Resolu-
tion [Hz]
Acquisi-
tion
interval
Resolu-
tion [Hz]
Acquisi-
tion
interval
Resolu-
tion [Hz]
Acquisi-
tion
interval
Resolu-
tion [Hz]
Acquisi-
tion
interval
8 M *
1
20 M 5 s 50 ns 20 k 50 s 10 k 100 s 4 k 250 s 2 k 500 s
4 M *
1
10 M 10 s 100 ns 10 k 100 s5 k200 s 2 k 500 s1 k1 ms
2 M *
1
5 M 20 s 200 ns 5 k 200 s 2.5 k 400 s 1 k 1 ms 500 2 ms
800 k *
1
2 M 50 s 500 ns 2 k 500 s 1 k 1 ms 400 2.5 ms 200 5 ms
400 k *
1
1 M 100 s1 s 1 k 1 ms 500 2 ms 200 5 ms 100 10 ms
200 k *
1
500 k 200 s2 s 500 2 ms 250 4 ms 100 10 ms 50 20 ms
80 k *
1
200 k 500 s5 s 200 5 ms 100 10 ms 40 25 ms 20 50 ms
40 k
100 k 1 ms 10
s 100 10 ms 50 20 ms 20 50 ms 10 100 ms
20 k
50 k 2 ms 20
s 50 20 ms 25 50 ms 10 100 ms 5 200 ms
8 k
20 k 5 ms 50
s 20 50 ms 10 100 ms 4 250 ms 2 500 ms
4 k
10 k 10 ms 100
s 10 100 ms 5 200 ms 2 500 ms 1 1 s
2 k
5 k 20 ms 200 s 5 200 ms 2.5 400 ms 1 250 ms 500 m 2 s
800
2 k 50 ms 500 s 2 500 ms 1 1 s 400 m 2.5 s 200 m 5 s
400
1 k 100 ms 1 ms 1 1 s 500 m 2 s 200 m 5 s 100 m 10 s
200
500 200 ms 2 ms 500 m 2 s 250 m 4 s 100 m 10 s 50 m 20 s
80
200 500 ms 5 ms 200 m 5 s 100 m 10 s 40 m 25 s 20 m 50 s
40
100 1 s 10 ms 100 m 10 s 50 m 20 s 20 m 50 s 10 m 100 s
20
50 2 s 20 ms 50 m 20 s 25 m 40 s 10 m 100 s 5 m 200 s
8 *
2
20 5 s 50 ms 20 m 50 s 10 m 100 s 4 m 250 s 2 m 500s
4 *
2
10 10 s 100 ms 10 m 100 s 5 m 200s 2 m 500 s 1 m 1 ks
1.33 *
2
3.33 30 s 300 ms 3.33 m 300 s 1.66 m 600s 666 1.5 ks 333 3 ks
800 m *
2
2 50 s 500 ms 2 m 500 s 1 m 1 ks 400 2.5 ks 200 5 ks
667 m *
2
1.67 60 s 600 ms 1.66 m 600 s 833 1.2 ks 333 3 ks 166 6 ks
400 m *
2
1 100 s 1 s 1 m 1 ks 500 2 ks 200 5 ks 100 10 ks
333 m *
2
833 m 120 s 1.2 s 833 1.2 ks 416 2.4 ks 166 6 ks 83.3 12 ks
133 m *
2
333 m 300 s 3 s 333 3 ks 166 6 ks 66.6 15 ks 33.3 30 ks
The cut-off frequency of the anti-aliasing filter is the same as the frequency range.
*1. The anti-aliasing filter is turned off.
*2. Cut-off frequency is 20 Hz.