7.5 Variable Function (Setting the Waveform Display Freely)
156
Make the Valiable Function Setting per channel _____________________
Procedure
To open the screen: Right-click and select [CHAN] [Each Ch] sheet
1
2
3
1
Enable the Variable function.
Move the flashing cursor to the [Variable], and select [On].
2
Set the display range per division
Move the flashing cursor to the [Range(div)], and enter numeri-
cal value.
(Measurement units depend on the measurement mode of the module.)
(When this is changed, the upper/lower limit values for the display also change ac-
cordingly.)
3
Set the waveform zero position to display on the vertical
axis (vertical axis).
Move the flashing cursor to the [Zero pos%], and enter numeri-
cal % value.
(When this is changed, the upper/lower limit values for the display also change ac-
cordingly.)
4
(When setting upper and lower values)
Move the flashing cursor to the [Upper] and [Lower], and specify the
values.
(When this is changed, the display range and zero position values also change ac-
cordingly.)
4
To reset the settings
Select [Reset] to return the set-
tings to the default values.
When upper and lower values are set, waveforms can be displayed at full
span on the screen.
Depending on the scaling setting, the upper and lower display values may be less
than 1. In such cases, set [Variable] to [On] and then select [Auto Set]. Easy-
to-read upper and lower limit values are set based on the currently set values.
For information on numeric input, see "7.1.3 Alphanumeric Input" (p.141).
•The [Unit List] sheet accessed from the Channel screen also lets you turn the
Variable function On or Off individually for each channel.
By using the Scaling and Variable functions together, the full span of a sen-
sor's output can be displayed. (p.157)
When Scaling is enabled, values are displayed in scaling units. When these
settings are changed, the numerical values indicating the display range on the
level monitor are changed accordingly.
7.5 Variable Function (Setting the Waveform Display Freely)
157
6
Chapter 7 Utility Functions
7
Displaying All Channels for Making the Variable Function Setting ______
Description
When setting combined use of the Scaling and Variable functions
When Auto-Correction of the Variable function is enabled (On, default set-
ting) (p.311)
The Variable function settings change according to Scaling and vertical axis
(voltage axis) range settings. Set Scaling before setting the Variable function.
If you change Scaling settings after enabling the Variable function, the Variable
setting voltage is automatically corrected so that the displayed size of waveforms
is unchanged.
When Auto-Correction of the Variable function is disabled (Off)
Set the Variable function after setting Scaling.
If setting the Variable function first, enter post-scaling values (converted physical
values).
To display the full span of output from a sensor
By using the Scaling function in combination, voltage from a sensor can be con-
verted to the physical units of the measurement object.
Procedure
To open the screen: Right-click and select [DISP] Waveform screenRight-click and select [CH.SET] 
Display range window
1
Enable the Variable function.
Move the flashing cursor to the [Variable], and select [On].
2
Set the upper and lower limits.
Move the flashing cursor to the [Upper] and [Lower], and
enter numerical value.
12
A
A
A
A
Example:
Set Scaling as follows:
Scaling: Decimal or exponent, Two-Point Setting
Units: A
Sensor Output (Input 1): 1.23 [V]
(Scale 1): 0 [A]
Sensor Output (Input 2): 5.78 [V]
(Scale 2):10 [A]
(with Variable function Off)
Voltage from the sensor is displayed as voltage.
It is displayed with the vertical axis (voltage axis) range
and at the zero position set on the Channel settings win-
dow (
[Analog] sheet).
The Variable function is set as follows:
Variable: On, Set Upper/Lower Limits
Lower Limit: 0 [A] Upper Limit: 10 [A]
The full span of output from the sensor is displayed.
7.6 Fine Adjustment of Input Values (Vernier Function)
158
Fine adjustment of input voltage can be performed arbitrarily on the Waveform screen. When
recording physical values such as noise, temperature and acceleration using sensors, amplitude
can be adjusted to facilitate calibration.
7.6 Fine Adjustment of Input Values (Vernier
Function)
Normal Display Vernier Function Enabled
1.2 V input is displayed as 1.0 V
1.2 V
1.0 V
Procedure
To open the screen: Right-click and select [DISP] Waveform screen Right-click and select [CH.SET]
Channel settings window ([Analog] sheet)
1
Move the flashing cursor to the [] vernier setting item of
the channel to adjust.
2
Make the adjustment while watching the waveform.
Vernier Magnify waveform
Vernier
Compress waveform
Vernier Reset
Return waveform to original position
The adjustment range is 50 - 200% of the original waveform. The magnifica-
tion/compression ratio is not displayed.
Vernier adjustments cannot be verified on waveforms.
The waveform data (saved file data) is adjusted by the Vernier function.