MR8740、MR8741_user_manual_eng_20191016H.pdf - 第71页

2.7 Performing Calibration (When Mounting MR8990) 59 2 Chapter 2 Measurement Prep arations This procedure compen sates for MR8990 Digital Voltmeter Unit differences. The compensation procedure is performed for all channe…

100%1 / 428
2.6 Adjusting the Zero Position (Zero-Adjust)
58
This procedure compensates for module differences and sets the reference potential of the instru-
ment to 0 V.
The compensation procedure is performed for all channels and ranges.
Perform zero-adjust in the following cases.
When an module was changed.
When power was turned off and on again.
When settings were initialized (system reset).
When DC/RMS is switched at the 8971Current Unit (model MR8740 only),
8972 DC/RMS Unit, or U8974 High Voltage Unit
When measurement mode has been switched on Model U8979 Charge Unit.
When the ambient temperature has changed significantly.
Zero-position drift
*
may occur.
* Drift: This refers to spurious output caused by a shift in the operating point of an opera-
tional amprifier. Drift can occur due to changes in temperature and due to compo-
nent aging over a period of use.
2.6 Adjusting the Zero Position (Zero-Adjust)
To obtain precise measurements, provide about 30 minutes warm-up after turning power on to allow
the internal temperature of the modules to stabilize.
Note that zero-adjust cannot be performed during a measurement.
During zero-adjust, the mouse operation is disabled. (The procedure may take several seconds.)
Before starting zero-adjust
Procedure
To open the screen: Right-click and select [CHAN] [Unit List] sheet
1
Move the flashing cursor to the [Zero-Adjust]
item.
2
Select [Exec Zero-Adjust].
The zero-adjust procedure is carried out.
Zero-adjust has no effect on the 8969 and U8969 Strain Unit.
(Perform zero-adjust using Auto Balance.)
See:"7.9.4 Setting Model 8969 and U8969 Strain Unit" (p.165)
MR8990 Digital Voltmeter Unit performs calibration when zero adjustment is
performed.
See:"2.7 Performing Calibration (When Mounting MR8990)" (p.59)
2.7 Performing Calibration (When Mounting MR8990)
59
2
Chapter 2 Measurement Preparations
This procedure compensates for MR8990 Digital Voltmeter Unit differences.
The compensation procedure is performed for all channels and ranges.
In the following cases, calibration should be executed again.
When a module was changed.
When power was turned off and on again.
When settings were initialized (system reset).
When the ambient temperature has changed significantly.
2.7 Performing Calibration (When Mounting
MR8990)
Before performing this procedure, allow the equipment to warm up for about 30 minutes after the
power is turned on to enable the internal temperature of the modules to stabilize.
Note that calibration cannot be performed during measurement. However, if the calibration setting is
set to ON, calibration is performed at the start of measurement.
Mouse operations are not accepted during calibration. (The procedure may take several seconds.)
Before starting calibration
Procedure
To open the screen: Right-click and select [CHAN] [Unit List] sheet
1
Move the flashing cursor to the [Zero-Adjust].
2
Select [Execute Zero-Adjust].
Calibration is executed.
2.7 Performing Calibration (When Mounting MR8990)
60