IPC-TM-650 EN 2022 试验方法.pdf - 第135页
flushed out during burner ignition and flame adjustment resulting in a false yellow flame tip. Proper flame adjustment becomes very difficult. 6.3 When the flame is correct and the specimens end is at the proper height a…

[257 ± 3.6°F],
then cooled in a desiccator for 4 hours mini-
mum prior to testing.
5.2.3
Reserve
sets of specimens may also be conditioned
concurrently for failure verification purposes.
5.3
Preparation for Testing
5.3.1 Adjustment of Test Flame
The
burner is ignited and
adjusted to produce a blue flame 19 mm [0.75 in] high. The
correct flame is obtained by adjusting the gas supply and the
air ports of the burner until a blue flame with a yellow- tipped
outer cone 19 mm [0.75 in] high is produced. The air supply
is increased slightly by opening the air ports only until the yel-
low tip just disappears and completely blue inner and outer
flame cones are formed. The flame is remeasured for correct
height. The procedure is repeated as necessary until all con-
ditions are met. The burner tube is vertical during adjustment
and testing. (See 6.2, 6.3, and 6.4.)
5.3.2
Specimen Mounting
Each
specimen is mounted in
the test fixture with its longitudinal axis vertical. The clamp
used shall cover no more than the upper 6.4 mm [0.25 in] of
the specimen. The vertical position of the test fixture/
specimen assembly is adjusted so that the lower end of the
specimen is 9.5 mm [0.375 in] above the top of the burner
tube.
5.4
Measurement
5.4.1
The
test flame is placed centrally under the lower end
of the specimen for 10 ± 0.5 seconds. The burner is with-
drawn from the specimen. If active combustion ceases prior
to the specimen being completely consumed, the test flame is
to be immediately placed under the specimen for an additional
10 ± 0.5 seconds, then withdrawn as before.
5.4.2
Evaluation and Report
Steps
5.5.1, 5.5.2, and
5.5.3 shall be performed for each test condition.
5.5
Recording Data
The
following data is to be observed
and recorded.
5.5.1
Duration
of specimen burning to the nearest second
after the first test flame application for each specimen.
5.5.2
Duration
of specimen burning to the nearest second
after the second test flame application for each specimen.
5.5.3
Duration
of specimen burning plus glowing to the
nearest second after the second test flame application for
each specimen, only if required by the specification.
5.5.4 If any specimen burns up to the holding clamp on any
ignition.
5.6
Calculation
5.6.1
Calculate
the total specimen burning time and the
average burning time based on ten ignitions per set of five
specimens.
5.6.2
Calculation
of the glowing time for each specimen, if
required by the specification.
5.7
Report
The
material shall be reported as out of compli-
ance for one or more of the following test results, unless oth-
erwise specified.
5.7.1
More
than one specimen per set burns up to the hold-
ing clamp on any ignition.
5.7.2
More
than one specimen per set burns for a period of
time longer than allowed by the specification for a single
specimen.
5.7.3
The
total specimen burning time or the average speci-
men burning time as applicable exceeds the maximum
allowed by the specification and is beyond the tolerance
specified in 5.5.4.
5.7.4
More
than one specimen glows for a period of time
greater than allowed by the specification (when applicable).
5.8
Retest
If
only one specimen per set fails to comply with
the requirements, the reserve set of specimens shall be
tested. In the case of total or average specimen burning time,
the reserve set shall be tested only if these calculated values
exceed the specification maximum by five seconds or less. All
specimens from the reserve set shall comply with the require-
ments.
6.0 Notes
6.1
Most
laminates covered by this test method do not drip
molten
or flaming material, and therefore provisions for this evalu-
ation have not been described. If necessary, refer to UL-STD-94.
6.2 The
inside of the burner barrel should be cleaned fre-
quently. Specimen combustion by-products can collect
around and inside the barrel tip. These deposits can be
IPC-TM-650
Number
2.3.10
Subject
Flammability
of Laminate
Date
12/94
Revision
B
P
age2of3
电子技术应用 www.ChinaAET.com

flushed
out
during
burner ignition and flame adjustment resulting
in a false yellow flame tip. Proper flame adjustment becomes very
difficult.
6.3
When
the flame is correct and the specimens end is at
the proper height above the burner 9.5 mm [0.375 in], the
inner blue come of the flame will just meet the end of the
specimen. The specimen will then be ignited by the hottest
area of the flame.
6.4
Accurate
centering of the flame under the specimen is
essential for consistent testing.
6.5
Desiccator Conditions
The
Test Methods Task Group
determined that a great majority of test laboratories are unable
to consistently hold the Relative Humidity in a desiccator to
less than 20%. Based on data from participating company lab
management, the lowest practically feasible RH for use with
the affected IPC Test Methods is 30% maximum.
IPC-TM-650
Number
2.3.10
Subject
Flammability
of Laminate
Date
12/94
Revision
B
P
age3of3
电子技术应用 www.ChinaAET.com

1
Scope
This
test method is intended to evaluate the rela-
tive flame resistance characteristics of a permanent solder
mask coating on specified printed wiring laminates. It is
intended for use on laminate 0.5 mm thick and greater. It
should indicate to what degree, if any, that the coating may
diminish the flammability of the laminate. The performance
level of a material by these methods should not be assumed
to correlate with its performance in end-use applications.
2
Applicable Documents
IPC-TM-650 Test Method Manual
2.3.6
Etching, Ammonium Persulfate Method
2.3.7 Etching, Ferric Chloride Method
2.3.7.1 Cupric Chloride Etching Method
3
Test Specimen
3.1
Test
specimens shall be prepared using vendor recom-
mendations or manufacturers’ production process and fabri-
cated parallel to either the warp or the fill direction as conve-
nient.
3.2 Test
specimens shall be 130 mm±5mminlength and
13 mm ± 0.5 mm in width by the thickness being tested.
Metal cladding shall be removed by etching, using IPC-TM-
650, Test Method 2.3.6, 2.3.7, or 2.3.7.1.
3.3
The
solder mask coating shall be applied evenly at both
the minimum and maximum anticipated coating thickness
over both the minimum and maximum laminate thickness
desired.
3.4
All
of these specimen sets, as well as one specimen set
of uncoated laminate, for each laminate thickness, should
come from the same laminate lots. In addition, if other solder
masks are to be compared, all should be applied to the same
lots, or results can vary widely. Pretesting of laminate lots is
recommended.
3.5
Edges
may be smoothed after fabrication, providing that
any radius imparted to the corners does not exceed 12.5 mm.
3.6 Five
specimens shall be prepared for each condition
required (two conditions—see specimen conditioning).
Reserve duplicates of five specimens are also required in the
event a retest is necessary. Total specimens per set are there-
fore equal to 20. Total sets required, including uncoated sub-
strate, would be six.
Note:
Additional
coatings tests would not have to include the
uncoated sets if the laminate lots are the same and submitted
simultaneously.
4
Apparatus
4.1 Test Chamber
A
laboratory hood, totally enclosed,
with a heat resistant glass window for observing the test, shall
be used. The exhaust fan shall be turned off during the test,
but may be turned on periodically to clear out the fumes and
carbonized airborne particles between tests.
4.2
Clamping Devices
A
clamping device, adjustable for
vertical positioning of the test specimen, shall be provided
within the test chamber so the specimen will hang with its
length in a vertical position approximately coincident with the
central vertical axis of the test chamber.
4.3
Laboratory Burner
A
Bunsen or Tirrell Burner shall be
used having a tube length of 100 mm and an inside diameter
of 9.4 mm ± 1.5 mm. The burner shall not be equipped with
end attachments.
4.4
Gas Supply
The
gas supply shall be regulated and
metered for uniform flow. The standard gas shall be technical
grade methane. Natural gas having a nominal heat content of
1000 BTU per cubic foot may be substituted. Other fuel
gases, such as butane, propane, and acetylene may be used;
however, technical grade methane will be used for referee
testing.
4.5
Timing Device
A
stop watch or other suitable timing
device with a precision of 0.5 seconds minimum
4.6
Desiccator
A
desiccator containing dried silica gel or
calcium chloride
4.7
Circulating Oven
A
conditioning oven of circulating
draft type capable of maintaining 70°C ± 1°C
The
Institute for Interconnecting and Packaging Electronic Circuits
2215 Sanders Road • Northbrook, IL 60062-6135
IPC-TM-650
TEST
METHODS MANUAL
Number
2.3.10.1
Subject
Flammability
of Soldermask on Printed Wiring
Laminate
Date
8/98
Revision
Originating Task Group
N/A
Material
in this Test Methods Manual was voluntarily established by Technical Committees of the IPC. This material is advisory only
and its use or adaptation is entirely voluntary. IPC disclaims all liability of any kind as to the use, application, or adaptation of this
material. Users are also wholly responsible for protecting themselves against all claims or liabilities for patent infringement.
Equipment referenced is for the convenience of the user and does not imply endorsement by the IPC.
P
age1of4
电子技术应用 www.ChinaAET.com