IPC-TM-650 EN 2022 试验方法.pdf - 第493页

Calibrations only last 24 hours, so calibration shall be per- formed within 24 hours of the measurement. See 6.1 for more calibration notes. 5.2 Set up the unit to sweep the target frequency ± 0.5% of the target. 5.3 Mea…

100%1 / 824
1
Scope
This
procedure outlines a test method to deter-
mine the permittivity (dielectric constant or E’r) and loss tan-
gent (dissipation factor or Tanδ) of printed wiring materials at
various frequencies (from 1 MHz to 1.5 GHz) using a single
test fixture for the measurement.
The permittivity and loss tangent are measured using a narrow
sweep of frequency around the target or desired frequency.
The test method is built around the capability of currently
available materials analyzers, which use a capacitance
method to determine permittivity.
This test method is not intended for low loss materials, such
materials may be tested at fixed frequencies using other IPC
test methods.
2
Applicable Documents
HP 4291A-5 Product Note
‘‘Dielectric
Constant Evaluation
of Rough Surface Materials,’’ which describes how to make
accurate measurements using the HP 4291A and HP
16453A.
HP
Application Note 380-1
‘‘Dielectric
Constant Measure-
ments of Solid Materials,’’ which contains a technical back-
ground, suitable for this subject.
3
Test Specimen
3.1
Each
specimen shall be 50 mm x 50 mm by the thick-
ness of the substrate material. Within the limits of the test fix-
ture, the thicker the sample the less error in the measure-
ments. Multilayer samples can be used to increase the
thickness of the sample, but these cannot be simple stacked
layers; they must be physically bonded with no air gaps
between the layers. A target thickness would be 1.0 mm, but
both thinner and thicker samples will work.
3.2
Three
specimens are required for this test.
3.3
All
materials are affected by moisture, including all rein-
forced laminates and most films. Therefore, all samples shall
be conditioned at 23°C ± 2°C and 50% RH ± 5% RH for a
minimum of 24 hours prior to testing. However, if a sample
has recently been etched or exposed to excessive moisture, it
should be dried in an air-circulating oven for two hours at
105°C +5°C, -2°C prior to testing and conditioned at room
temperature as mentioned above.
3.4
Sample Surface Preparation
3.4.1
It
is preferred that the sample be patterned with a
conductive material in the shape and size of the test elec-
trode. This conductive material is preferably 100 angstroms of
vapor deposited copper. Other metals may be used. In all
cases, the conductor on the sample must make good electri-
cal contact with the fixture electrode. Such a conductive pat-
tern eliminates air gaps and other potential sample mounting
errors.
3.4.2
Bare
dielectric materials may be tested with this test
method. The fixture electrodes must be applied with some
level of force to ensure a gap-free contact area. Determining
the correct force setting may require some trial and error test-
ing for each type of sample (see 6.4).
4
Equipment/Apparatus
4.1
The
Hewlett-Packard Impedance Material Analyzer,
model 4291A, or equivalent is recommended.
4.2 Hewlett-Packard
model number 16453A test fixture, or
equivalent
4.3
An
appropriate calibration-verification kit and a fixture-
correction kit as recommended in the instrument’s manual
(i.e., HP4291A Calibration kit). Such a kit usually includes the
following devices:
• OPEN and SHORT for fixture correction
• 50 Ohms impedance
• Dielectric (PTFE) of known characteristic for the purpose of
the calibration verification
4.4
Micrometer,
capable of 0.001 mm resolution
4.5
Circulating
oven capable of 105°C +5°C, -2°C
5
Procedure
5.1
Calibrate
the instrument using the calibration kit accord-
ing to the recommendations of the instrument manufacturer.
The
Institute for Interconnecting and Packaging Electronic Circuits
2215 Sanders Road Northbrook, IL 60062
IPC-TM-650
TEST
METHODS MANUAL
Number
2.5.5.9
Subject
Permittivity
and Loss Tangent, Parallel Plate,
1 MHz to 1.5 GHz
Date
11/98
Revision
Originating Task Group
HDI Test Methods Task Group (D-42a)
Material
in this Test Methods Manual was voluntarily established by Technical Committees of the IPC. This material is advisory only
and its use or adaptation is entirely voluntary. IPC disclaims all liability of any kind as to the use, application, or adaptation of this
material. Users are also wholly responsible for protecting themselves against all claims or liabilities for patent infringement.
Equipment referenced is for the convenience of the user and does not imply endorsement by the IPC.
P
age1of5
电子技术应用       www.ChinaAET.com
Calibrations
only last 24 hours, so calibration shall be per-
formed within 24 hours of the measurement. See 6.1 for more
calibration notes.
5.2 Set up the unit to sweep the target frequency ± 0.5% of
the target.
5.3
Measure
the sample thickness with the micrometer and
insert into the test fixture. The sample must make good con-
tact with the fixture electrodes (see 6.1 concerning the proper
force to be applied). The sample must not touch the back wall
of the fixture. The sample electrode placement and thickness
measurement shall be obtained from the same area of the
sample.
5.4
Run
the test and record the average permittivity and loss
over the narrow frequency range sweep. The scanned data
may also be saved on disk. See 6.2 for comments on
expected behavior for permittivity as a function of frequency.
5.5
Repeat
5.1 through 5.4 for all desired frequencies.
5.6
Report
the average permittivity and loss at the frequen-
cies requested.
6 Notes
6.1
Correct
calibration and operation of the test equipment
is required to obtain accurate measurement of permittivity and
loss. Proper sample preparation is also very important for
obtaining useful data from this test. Calibrate the materials
analyzer in accordance with manufacturer’s instructions. An
automatic program has been developed for the HP 4291A,
which will ease calibration and setup (see 6.6).
6.2
The
permittivity should decrease slightly with increasing
frequency. If it increases greatly or decreases more than 0.2
units from approximately 20 MHz to 1.2 GHZ, reposition
(reset) the sample in the fixture and measure again (check for
debris between the electrodes; blow out with air).
6.3
Testing
at temperatures and humidities other than room
temperature may be performed with this instrument, as a spe-
cialized fixture can be placed in a temperature chamber. A
temperature chamber must be used when testing with this
fixture under conditions where condensation might contami-
nate the electrodes, as such contamination gives spurious
results.
6.4
The
pressure of the test fixture on the specimen affects
the measured permittivity and loss values, in particular for
un-metallized test specimens. Too light of pressure reduces
the area of electrode/sample contact, thus leaving air gaps,
which result in erroneous measurements. If the pressure is too
high, the sample can be reduced in thickness and the mea-
sured values would be incorrect because the thickness is
unknown. Making measurements while adjusting the force
should lead the operator to a force setting where the mea-
sured value is independent of the force applied.
6.5
The
HP 4291 Materials Analyzer and related fixtures and
calibration kits are available from Hewlett Packard, (800) 452-
4844.
6.6
Reference Program for Automatic Calibration and
Operation
This
automatic calibration and operation pro-
gram was developed for the HP 4291A and is published in
this method as a reference. Although the program listed in this
section has been tested and used, it is given here for ‘‘refer-
ence only.’’
6.6.1
Procedure Using Automatic Program
Turn
on the
analyzer with the program/calibration disk in the drive and fol-
low the user friendly calibration instructions, which appear on
the monitor. The calibration on the HP4291 lasts about 24
hours; after that, it begins to drift and provides slightly higher
values as a function of time. The calibration procedure in
6.6.1.1 through 6.6.1.10 should therefore be performed, at a
minimum, on a daily basis. During the calibration procedure,
the line traces on the monitor should be observed during each
step. Noisy or erratic traces are an indication of external inter-
ference. If noise is observed, the calibration procedure should
be aborted and rerun. Clean the electrode on the test head,
standards, and fixture connections and electrodes on a regu-
lar (weekly) basis. Blow dry.
6.6.1.1
Allow
at least 30 minutes for the unit to warm up
and stabilize.
6.6.1.2
As
the unit is a frequency sweeping unit, enter the
start and stop frequency (in megahertz) of the test (single fre-
quency tests use close start/stop frequencies, which should
be ± 0.5% of the target frequency).
6.6.1.3 Place
the OS (open) calibration standard on the test
head as prompted on the unit’s monitor and press Return on
the unit’s keyboard or ‘‘x1’’ on the unit.
IPC-TM-650
Number
2.5.5.9
Subject
Permittivity
and Loss Tangent, Parallel Plate,
1 MHz to 1.5 GHz
Date
11/98
Revision
P
age2of5
电子技术应用       www.ChinaAET.com
6.6.1.4
Place
the 0 (short) calibration standard on the test
head as prompted on the unit’s monitor and press Return on
the unit’s keyboard or ‘‘x1’’ on the unit.
6.6.1.5
Place
the 50 calibration standard on the test head
as prompted on the unit’s monitor and press Return on the
unit’s keyboard or ‘‘x1’’ on the unit.
6.6.1.6 Place
the HP 16453A test fixture or equivalent on
the test head as specified in accordance with the manufactur-
er’s instructions as prompted on the unit’s monitor.
6.6.1.7 Open
the fixture gap and lock it open as prompted
on the monitor and press Return on the unit’s keyboard or
‘‘x1’’ on the unit.
6.6.1.8
Short
the fixture gap as prompted on the monitor
and press Return on the unit’s keyboard or ‘‘x1’’ on the unit.
6.6.1.9
Place
the pure PTFE specimen in the fixture in close
contact with electrodes and with the PTFE specimen not
touching the back wall of the fixture, then press Return on the
unit’s keyboard or ‘‘x1’’ on the unit. Pull out the PTFE and
recheck. Permittivity should be within ± 0.01 or ± 0.02 units.
If not, recalibrate starting at ‘‘Fixture.’’ If it is still off, redo the
calibration from beginning.
6.6.1.10
Test
the sample in accordance with Section 5.
6.6.2
The
following calibration/test program is the latest ver-
sion of what was developed for the P H4291A. Using a text
editor (no embedded coding), enter this program on a DOS-
formatted disk and name it ‘‘AUTOST.’’ Place the disk in the
unit and turn it on (If you have any improvements PLEASE for-
ward them to the IPC): The program disk will calibrate and
configure the instrument’s screen. Two windows will appear.
The top window should be permittivity (E’r) and the bottom
window should be loss tangent (δ). Both windows should be
set to the frequency range entered in 6.3.1.2. If they are not,
return to 6.3.1 and repeat the process.
100 !**** PLEASE SET PTFE THICKNESS ON LINE 1610 AND
2600 ****
110 !**** PLEASE SET PTFE PERMITTIVITY ON LINE 2580
****
120 !**** PLEASE SET PTFE LOSS TANGENT ON LINE 2590
****
130 !****
200 !**** SAVE THIS PROGRAM ON A DOS FORMATTED
210 !**** DISK USING A TEXT EDITOR (LINES 100-5000
220 !**** NAME THE PROGRAM ‘‘AUTOST’’ WITH NO
EXTENSION
230 !**** PLACE DISK IN UNITS DRIVE AND TURN UNIT ON
240 !**** PROGRAM WILL AUTOMATICALLY RUN. TO
RESET
250 !**** UNIT FOR NEW FREQUENCY RANGE PRESS
‘‘STOP’’
260 !**** ON UNITS FACE THEN ‘‘START’’ ON UNITS FACE.
270 !**** IF YOU WANT TO RE-CALIBRATE ANSWER YES [1]
280 !**** WHEN PROMPTED
290 !****
1000 !**** INITIALIZE UNIT FOR TEST ***
1010 ASSIGN @Hp4291 TO 800
1020 Scode=8
1030 CLEAR @Hp4291
1040 ABORT Scode
1050 OUTPUT @Hp4291;‘‘DISP:ALL BST’’
1070 !
1100 !**** CALIBRATION STATE CHECKING ****
1110 OUTPUT @Hp4291;‘‘SENS:CORR1?’’
1120 ENTER @Hp4291;Stat
230 IF Stat THEN
1130 INPUT ‘‘RE-CALIBRATE NOW? [1] Yes; [0] No
[RETURN OR x1]’’,Ans$
1140 IF Ans$<>‘‘1’’ THEN GOTO Test_loop
1150 END IF
1160 !
1200 !**** SETTING UP START/STOP FREQUENCY ****
1210 INPUT ‘‘Enter Start Frequency in MHz, [RETURN OR
x1]’’,A
1220 B=A*1000000
1230 OUTPUT @Hp4291;‘‘SENS:FREQ:STAR ’’;B
1240 INPUT ‘‘Enter Finish Frequency in MHz, [RETURN OR
x1]’’,A
1250 B=A*1000000
1260 OUTPUT @Hp4291;‘‘SENS:FREQ:STOP ’’;B
1270 !
1300 !**** SETTING ENABLE REGISTER FOR SRQ ****
1310 OUTPUT @Hp4291;‘‘STAT:INST:ENAB 256’’
1320 OUTPUT @Hp4291;‘‘*SRE 4’’
1330 !
1400 !**** SETTING TRIGGER SOURCE INTERNAL, CON-
TINUOUS OFF ****
1410 OUTPUT @Hp4291;‘‘TRIG:SOUR INT’’
1420 OUTPUT @Hp4291;‘‘INIT:CONT OFF’’
1430 !
1500 !**** SETTING CAL MODE TO FIXED/USER ****
1510 OUTPUT @Hp4291;‘‘SENS:CORR1:COLL:FPO USER’’
1520 OUTPUT @Hp4291;‘‘SENS:CORR2:COLL:FPO USER’’
IPC-TM-650
Number
2.5.5.9
Subject
Permittivity
and Loss Tangent, Parallel Plate,
1 MHz to 1.5 GHz
Date
11/98
Revision
P
age3of5
电子技术应用       www.ChinaAET.com