IPC-TM-650 EN 2022 试验方法.pdf - 第222页
The measurement location can then be observed on the video camera image and adjusted if necessary. The collimator area indicated on the video image should fit entirely within the test area of the sample specimen as seen …

5 Procedure
5.1 Instrument Setup
Prior to the purchase of the Certi-
fied Reference Materials (CRMs), confirm with the XRF
manufacturer that the instrument is capable of measuring
phosphorus content and obtain details of the recommended
machine set-up and operational procedures.
Instrument setups usually contain a product file that contains
the required measurement specific hardware and software
settings for the application. In addition, the product file con-
tains a calibration file which defines the calibration settings
and certified reference material to be used.
5.2 Typical Instrument setup conditions and measuring
ranges are as follows:
• Aperture Size: 1 mm for both 10kV and 50kV applications.
• Anode Current (I): I=1 mA for 10kV and I=0.15 mA for 50kV
(Anode current setup maximizing achievable instrument
count rates will yield best instrument repeatability, reference
5.3).
• Primary Beam Filter: NO filter for 10 kV and Ni Filter for
50 kV.
• Measurement Time: 120 s for 10kV and 20 s for 50kV.
5.3 Instrument Calibration Calibration shall be per-
formed with CRM’s according to the instrument manufacturer
instructions. The CRM’s shall be traceable to national labora-
tories. The structure of the reference material shall be similar
to the samples under investigation, i.e., NiP/Cu/PCB, Au/NiP/
Cu/PCB or Au/Pd/NiP/Cu/PCB. Individual calibration foils
shall not be used for multilayer coatings. The certified refer-
ence standards shall have compositions and thicknesses
similar to the samples to be measured. If desired, it is possible
to calibrate an instrument over the full (low to high) phospho-
rous range. However, optimum accuracy can be achieved by
calibrating each phosphorous range (low, mid, and high)
separately. Each phosphorous content range should be cali-
brated with no less than 4 standards per range. No less than
3 measurements per calibration standard shall be performed.
Calibration checks should be performed after each calibration
and periodically by re-measuring the calibration standards. If
the results are within the measurement uncertainty of the
standards and the uncertainty of the measurement itself, no
action is required. If not, a recalibration of the instrument is
required. Typical CRM standards used and results obtained
are summarized in Table 1.
5.4 Sample Placement There are some basic rules for
positioning specimens. For each measurement, it shall be
ensured that the X-ray fluorescence radiation can reach the
detector without obstruction. For flat, unpopulated PCB
boards, this is not a problem.
If populated boards are being measured, the operator shall
note the position of the detector and position the sample such
that no components are present in locations that would
prevent the radiation emanating from the measurement loca-
tion from reaching the detector, as illustrated schematically in
Figure 2.
The area measured should be flat and not tilted.
5.5 Measurement XRF equipment operation is instrument
specific and shall be in accordance with the instrument
manufacturer’s instructions. Always ensure that the correct
measurement file is selected for the application to be mea-
sured. Typically, instruments will slide the measuring stage out
of the instrument when the measurement chamber is opened.
The test sample is then positioned on the programmable X-Y
stage such that the laser pointer points at the measurement
location. When the measurement chamber is closed, the
stage will automatically retract into the chamber.
IPC-2344-2
Figure 2 Sample Placement with Respect to Detector
IPC-TM-650
Number
2.3.44
Subject
Determination of Thickness and Phosphorus Content in
Electroless Nickel (EN) Layers by X-Ray Fluorescence (XRF)
Spectrometry
Date
03/16
Revision
Page3of4

The measurement location can then be observed on the video
camera image and adjusted if necessary. The collimator area
indicated on the video image should fit entirely within the test
area of the sample specimen as seen in Figure 3. The sample
image is then focused with the autofocus feature of the instru-
ment.
A minimum of 5 measurements shall be made per measure-
ment location (0.060 in x 0.060 in pad). Using a polycapillary
instrument, the 5 measurements shall be made at different
locations on the pad or the instrument should be used in a
scanning mode across the pad. On each side of the test
specimen on which an electroless nickel coating has been
applied, three pads shall be measured. The Measurement
Report shall include as a minimum:
• Instrument used
• Size of the collimator
• Measurement time
• Excitation conditions
• Individual measurement results
• Statistical measurement parameters such as mean, stan-
dard deviation and relative standard deviation
• Specification Limits as required
• Operator, time and date
6 Notes
6.1 Measurement Results:
Table 2 demonstrates the excellent standard deviation
achievable (0.4 wt.-% for 60 s measuring time) for measure-
ment of P-concentration. It should be noted that a high total
spectral intensity of more than 50,000 cps is the result of very
high flux excitation by an instrument using a polycapillary X-ray
optic emitted from a relatively small measuring spot of less
than 50 µm Ø.
In the case of standard aperture beam collimation, the total
measuring time for similar precision is expected to be a factor
of 2-3 X longer.
IPC-2344-2
Figure 3 Positioning ofa1mmØCollimator on a 1.5 mm
x 1.5 mm [0.060 in x 0.060 in] ENEPIG Pad
Table 1 Measurement of Coating
Thickness and Phosphorus Content:
Typical achievable results for NiP/Cu/PCB samples
(1 mm collimator; measuring time 120 s)
CRM
Specified Values Measured Values
d (µm) %P d (µm) %P
CRM 1 5.20 (0.1) 0 5.29 (0.1) 0.0 (0.3)
CRM 2 7.35 (0.2) 0 7.43 (0.1) 0.1 (0.3)
CRM 3 1.2 (0.1) 8 (0.4) 1.2 (0.1) 7.7 (0.3)
CRM 6 2.89 (0.1) 10.6 (0.4) 2.88 (0.1) 10.7 (0.4)
CRM 4 6.9 (0.2) 9.0 (0.4) 6.5 (0.1) 8.9 (0.3)
CRM 5 5.90 (0.2) 11.2 (0.4) 5.7 (0.1) 11.1 (0.3)
CRM 7 11.20 (0.2) 11.3 (0.4) 11.2 (0.1) 11.4 (0.3)
Table 2 Evaluation of Measurement
Repeatability using a Polycapillary XRF:
5 Readings of the ENEPIG Sample shown in Figure 3.
Au Pd NiP wt-%P
Mean 0.049 µm 0.096 µm 3.2 µm 9.3
Standard
deviation
0.002 µm 0.002 µm 0.026 µm 0.413
50 nm Au/96 nm Pd/3.2 µm NiP9.3/Cu/PCB
(Small spot polycapillary instrument, measuring time 60 s)
IPC-TM-650
Number
2.3.44
Subject
Determination of Thickness and Phosphorus Content in
Electroless Nickel (EN) Layers by X-Ray Fluorescence (XRF)
Spectrometry
Date
03/16
Revision
Page4of4

1
Scope
This
test method uses pressure sensitive tape to
determine the adhesion quality of platings, marking inks or
paints, and other materials used in conjunction with Printed
Boards.
2
Applicable Documents
Commercial Item Description (CID) A-A-113
Tape,
Pres-
sure Sensitive, Adhesive.
3
Test Specimens
Any
preproduction, first article, or pro-
duction printed board. A minimum of three tests should be
performed for each evaluation.
4
Apparatus or Material
4.1 Tape
A
roll of pressure sensitive tape 3M Brand 600
1/2 inch wide or a tape as described in (CID AA-113), Type 1,
Class B, except that the tape may be clear. The shelf life of
the tape is one year.
5
Procedure
5.1 Test
Press
a strip of pressure sensitive tape, 50 mm
[2.0 in] minimum in length, firmly across the surface of the test
area removing all air entrapment. The time between applica-
tion and removal of tape shall be less than one minute.
Remove the tape by a rapid pull force applied approximately
perpendicular (right angle) to the test area. An unused strip of
tape must be used for each test.
5.2
Evaluation
Visually
examine tape and test area for evi-
dence of any portion of the material tested having been
removed from the specimen.
5.3
Report
The
report should note any evidence of material
removed by this test.
6 Notes
6.1
If
plating overhang breaks off (slivers) and adheres to the
tape, it is evidence of overhang but not an adhesion failure.
6.2
If
foreign material (oil, grease, etc.) is present on the test
surface the results may be affected.
6.3
Certification
of 3M Brand 600 1/2 inch tape to CID-A-A-
113 is not required. The 3M Brand 600 1/2 inch tape is avail-
able through most office supply stores.
2215
Sanders Road
Northbrook, IL 60062-6135
IPC-TM-650
TEST
METHODS MANUAL
Number
2.4.1
Subject
Adhesion,
Tape Testing
Date
05/04
Revision
E
Originating Task Group
Rigid Printed Board Performance Task Group
(D-33a)
Material
in this Test Methods Manual was voluntarily established by Technical Committees of IPC. This material is advisory only
and its use or adaptation is entirely voluntary. IPC disclaims all liability of any kind as to the use, application, or adaptation of this
material. Users are also wholly responsible for protecting themselves against all claims or liabilities for patent infringement.
Equipment referenced is for the convenience of the user and does not imply endorsement by IPC.
P
age1of1
ASSOCIA
TION CONNECTING
ELECTRONICS INDUSTRIES
®
电子技术应用 www.ChinaAET.com