• Sweep Frequency Generator Mainframe H.P. 8350B • RF Plug-In, 0.01 to 20 GHz H.P. 83592A. A plug-in of nar- rower frequency range (X-band) may be selected at consid- erable cost savings. 83545A 5.9 12.4 GHz. • Power Spl…

100%1 / 824
from
the base boards to the top of the 25.4 mm x 51 mm area
to which clamping pressure is applied. The minimum horizon-
tal dimension must be enough to extend at least 6.5 mm
beyond the center line of the vertical portion of the probe line
on either side. For the pattern card of Figure 4 and fixture of
Figure 12, these minimums are 38.1 mm x 68.6 mm. For the
smaller size, the clamp force in 6.1 or Table 1 is not changed
because the effective area over which the force is applied is
not reduced. The test fixture is designed to accommodate a
total specimen thickness of either 3.18 mm ± 0.22 mm or
2.54 mm ± 0.18 mm from an even number of layers.
NOTE: Testing of built-up specimens introduces error, which
can exceed 5% due to air gaps. Exact correlation factors and
techniques must be agreed upon or other methods of test
used. The 1 MHz method of IPC-TM-650, Method 2.5.5.3,
can be used with a correction factor based on tests of
samples of the nominal thickness of Table 1 using both tech-
niques.
With some material types not based on woven fabric rein-
forcement, it is possible to machine specimens to achieve the
nominal thickness for test.
4.0
Suggested Electronic Apparatus
The
principal com-
ponents required for the test setup consist of the test fixture
described in 5.0, a microwave signal source, an accurate
means of measuring the signal frequency, an accurate means
for detecting power level, and an accurate method of deter-
mining frequency values above and below the resonant fre-
quency at the half-power level for the test fixture loaded with
the specimen.
The microwave signal source must be capable of providing an
accurate signal. During the required time period and range of
frequency needed to make a permittivity and loss tangent
measurement, the source must provide a leveled power out-
put that falls within a 0.1 dB range. When the source is set for
a particular frequency, the output must be capable of remain-
ing within 5 MHz of the set value for the time required to make
a measurement.
The means for measuring frequency shall have a resolution of
5 MHz or less and an accuracy of 8 MHz or less. An error of
+8 MHz in measurement of a resonant frequency for a mate-
rial with nominal permittivity of 2.50 represents a -0.004 error
in permittivity.
The means for detecting the power level shall have a resolu-
tion of 0.1 dB or less and be capable of comparing power
levels withina3dBrange with an accuracy of 0.1 dB. An error
of 0.1 dB in estimating half power frequency points can result
in an error in the loss tangent of about 0.0001 for a material
with 2.5 permittivity. See 7.2, equation 5.
4.1
Manual Test Setup
The
method of determining the
half-power points depends partly on the type of signal source
used. If the power input to the test fixture is maintained con-
stant as the frequency is varied, then an SWR meter may be
used to determine the half-power points at the output of the
test fixture. This may be accomplished by using a leveled
sweep generator or by using a tunable klystron (at a consid-
erable savings) and manually adjusting the power input to the
test fixture to a prescribed level by use of a variable attenua-
tor. A typical equipment list is shown below. Equivalent makes
and models of equipment may be substituted where it can be
shown that equivalent results are obtained. For example, if a
leveling system is not used and the power output of the
source varies widely with frequency, a ratiometer may be sub-
stituted for the two SWR meters. If only permittivity is desired,
it is not necessary to level the input.
The following equipment, or equivalent, may be used.
• Sweep Frequency Generator H.P. 8350B or 8620C
• X-Band Frequency Plug in Unit H.P. 83545A or 86251A
• Frequency Meter H.P. X532B
• Crystal Detector (2) H.P. 423B (Neg)
• Matched Load Resistor for one Crystal Detector H.P.
11523A, opt. 001
• SWR Meter (2) H.P. 415E
• Directional Coupler HP 779D
• 10 dB Attenuator H.P. 8491B
• 8.9 kN Dillon Force Gauge, Compression Model X, Part
Number 381612301, with ± 1% full scale accuracy
• Vise or press that is able to exert controlled 4.45 kN force
on the test fixture and that opens at least 127 mm to accept
the force gauge and test fixture
• Semi-rigid Coaxial Cable and Connectors
• Waveguide to Coaxial Adapters (2) H.P. X281A
• The measuring equipment shall be connected as shown on
Figure 1.
4.2
A Test Setup for Computer Automation of Data
The
following
components or equivalent, properly interconnected,
can be used most effectively with a computer control program
for automated testing.
IPC-TM-650
Number
2.5.5.5
Subject
Stripline
Test for Permittivity and Loss Tangent (Dielectric Constant
and Dissipation Factor) at X-Band
Date
3/98
Revision
C
P
age2of25
电子技术应用       www.ChinaAET.com
Sweep Frequency Generator Mainframe H.P. 8350B
• RF Plug-In, 0.01 to 20 GHz H.P. 83592A. A plug-in of nar-
rower frequency range (X-band) may be selected at consid-
erable cost savings. 83545A 5.9 12.4 GHz.
• Power Splitter H.P. 11667A
• Automatic Frequency Counter H.P. 5343A
• Source Synchronizer H.P. 5344A. Obtained as an intercon-
nected assembly with the counter.
• Coaxial cables and adapters.
• 10 dB Attenuator H.P. 8491B
• 8.9 kN Dillon Force Gauge, Compression Model X, Part
Number 381612301, with ±1% full scale accuracy.
• Vise or press that is able to exert controlled 4.45 kN force
on the test fixture and that opens at least 127 mm to accept
the force gauge and test fixture.
• Programmable Power Meter H.P. 436A
• Power Sensor H.P. 8484A with 70 to 10 dBm range.
• IEEE 488 (GPIB) cables
• Controlling computer with GPIB interface.
The above equipment is connected as follows as illustrated in
Figure 2:
RF connections. The power splitter connects directly to the
RF plug-in output. One output of the splitter connects by RF
cable to the counter input. The other output is connected by
RF cable to the attenuator which connects to one of the test
fixture probe lines.
Control connections. Connections between counter and
synchronizer are provided as specified by the manufacturer.
The FM output from the synchronizer connects by BNC to
the FM input on the sweeper. GPIB cables connect in par-
allel to sweeper, synchronizer, power meter, and computer
interface.
Other connections. The power sensor is connected to the
other probe of the fixture and its special cable connects into
the power meter.
• A synthesized CW generator can be used to replace the
sweeper, plug-in, power splitter, connector, and source syn-
chronizer for the simpler set-up shown in Figure 3.
4.3
Automated Network Analyzer for the Test Setup
The
instrumentation described in 4.1 or 4.2 may be replaced
with either a scalar or vector network analyzer with test cables
connected to the test fixture of 5.0 as the device under test
(DUT). Examples of automated network analyzers known to
be suitable include the Hewlett-Packard 8510 vector network
analyzer or the Wiltron Model 561 scalar network analyzer.
These or equivalent may be used.
Such instruments may be operated either manually or under
computer control with suitable programming to locate the
resonant frequency and the frequencies above and below
resonance where transmitted power is 3 dB below that at
resonance. Network analyzers have several advantages over
the instrumentation described in 4.1 and 4.2. Data collection
is rapid and may be continuously refreshed with averaging.
The log magnitude response curve for ratio of transmitted to
incident power (the S21 parameter) as dB versus frequency is
visible on a screen for easy verification of a valid resonance. A
large number of dB frequency data points near the resonance
are readily available for optional use of non-linear regression
analysis techniques to determine the frequency and Q values
with statistically better degrees of uncertainty than those
attainable by the three point (f
r
,f
1
and
f
2
)
method in either
section 6.2 or 6.3.
5.0
Test Fixture
5.1 Recommended Fixture Design
An
improved test fix-
ture design is shown that facilitates changing test pattern
cards and lends itself to control of temperature. The test fix-
ture shall be constructed as shown in Figure 4 through Figure
14.
The resonator circuit shown in Figure 4 is an example of a test
pattern designed for a material with a permittivity of 2.20. For
other permittivity values, different pattern dimensions will be
required as outlined in Table 1. It shall be defined on one side
of a material of similar type to that being tested, a laminate
with dielectric thickness of 0.216 mm ± .018 mm. The clad-
ding thickness is normally specified as MF-150F designation 1
copper (nominal thickness of 0.036 mm but designation down
to Q (0.010 mm) may also be used. Designation Q is preferred
for high permittivity materials as covered in 4.2 and 9.7, Note.
The reverse side of the circuit board has all copper removed.
The copper foil shall be of IPC-MF-150, type 1, electro-
deposited, type 5, wrought, or type 7, wrought-annealed. The
type of copper foil and the treatment for adhesion will affect
the Q measurement. The 1/Qc values in Table 1 do not take
into account surface treatments or higher resistivity values for
the conductor that are encountered with the specified foil
types.
IPC-TM-650
Number
2.5.5.5
Subject
Stripline
Test for Permittivity and Loss Tangent (Dielectric Constant
and Dissipation Factor) at X-Band
Date
3/98
Revision
C
P
age3of25
电子技术应用       www.ChinaAET.com
The
test pattern card shall have a permittivity equal to the
nominal value of the type being tested with a tolerance of ±
2.5% of the nominal value (measured by stacking sufficient
plies to the total thickness requirement of a specimen as
above. Use a photo resist and etching method capable of
reproducing circuit dimensions with ± 0.025 mm tolerance. All
copper shall be removed from the other side of the test pat-
tern card. See 9.7, Note, for special treatment of ceramic-
PTFE substrate types.
The pattern card of Figure 4 is 68.6 mm wide by 55.4 mm
high and is designed for the fixture hardware in Figure 5
through Figure 14. The length is cut so that when the pattern
card is clamped for the lap joint with the striplines on the base
card, the resonator is centered in the 51 mm high area above
the base plates of the fixture. For materials with permittivity
values higher than the nominal 2.50 shown in Table 1, please
see 5.2 for a discussion of recommended fixture modifica-
tions.
Probe line widths are based on ground plane spacing taken
as twice the nominal thickness of the two specimens plus
thickness of the pattern card and its 0.034 mm copper foil
pattern and computed as if the stripline were centered
between ground planes
(1,3)
.
Chamfer
values are based on published design curves
(2)
.
The
length of the four node resonator is given in Table 1.
Resonators of lower node values for the purpose of measur-
ing L according to 6.1, will be proportionately shorter with
the probe lengths modified so that the gap is the same.
The values for conductor loss, 1/Q
c
,
in Table 1 are calculated
from known properties of copper, the test frequency, the cal-
culated characteristic impedance of the section of stripline
comprising the resonator, and its cross-sectional geometry
using published formulas
(1)
.
The values shown are usually
biased low giving a high bias to loss tangent results, because
conductor actually used may not have a smooth surface and
may include oxides, microvoids, or other sources of higher
resistivity.
5.2
Fixture Modifications for High Permittivity Materi-
als
Modification
of the fixture design of Figure 5 through Fig-
ure 14 and pattern card dimensions in Figure 4 are recom-
mended to overcome problems experienced with extraneous
transmissions and resonances at frequencies near the desired
resonant peak.
5.2.1 Replace
the coax-stripline launcher shown in Figure 7.
The part suggested has a tab width of 1.27 mm and may be
replaced with Omni-Spectra Part No. 2070-5029-02, or
equivalent, intended for 1.57 mm ground plane spacing and
with a tab width of 0.635 mm. A further acceptable alternative
is to redesign the base plates to accept another type of
coaxial fitting such as a flange mount jack, which can be
modified to provide a smooth, low-reflection transition from
3.0 mm semirigid cable with Z
0
=
50 Ohm, low permittivity
insulation into stripline with Z
0
=
50 Ohm, and high permittiv-
ity insulation in the fixture.
5.2.2
If
the stripline launcher in 5.2.1 is used, the edge at
the step to accommodate the launcher body on the base
plate should be machined with a slight undercut for an acute
included angle of about 80°. This, combined with a means to
press the launcher body axially against the edge, will assure a
well-defined ground connection from coax to stripline. A
poorly defined ground connection with ground current path
length varying or longer than that of the signal conductor has
been found to give rise to scattering, reflections, and reso-
nances in the open ended probe line that are evident as extra-
neous fixture transmissions that may distort the resonant peak
to be measured.
5.2.3
Omit
the conductor lap joints but keep the extended
base cards in the fixture assembly. See figures 13 and 14.
With high permittivity materials, the lap joint also gives rise to
unwanted scattering, reflections, and resonances in the open-
ended probe line, as discussed in 5.2.2. For this purpose the
resonator pattern card will have a longer vertical dimension to
extend down to the launcher pin replacing the spacer board
in Figure 13. It should still center the resonator in the clamp-
ing block area. The base dielectric boards will be etched free
of metal. The ground plane foils will also extend down to the
launcher.
The feature of extending the base dielectric boards upward
above the base plates is to be retained as a means to prevent
premature damage to the resonator pattern card with
repeated loading and unloading of the fixture. The base plate
with the deeper step will be on the side toward which the
resonator pattern faces to avoid straining the offset launcher
tab during assembly.
5.2.4
Scale
down the fixture dimensions to move remaining
probe line resonances away from the resonant frequency of
interest. For ε
r
=
10.5 material, the following dimensions were
found effective.
IPC-TM-650
Number
2.5.5.5
Subject
Stripline
Test for Permittivity and Loss Tangent (Dielectric Constant
and Dissipation Factor) at X-Band
Date
3/98
Revision
C
P
age4of25