IPC-TM-650 EN 2022 试验方法.pdf - 第213页

5.2.3.2 Determination of F__film Prepare a film sample in a vial as standard. Make three injections with venting between injections and equilibration before each one, just as in the case of the liquid. The F__ret__film a…

100%1 / 824
5.1.2
Evaluation
The
volatile content of each specimen is
calculated as follows:
Volatile
Content (%)
=
(W
I
−W
S
)(W
F
−W
S
)
(W
I
−W
S
)
x
100
5.1.3
Report
The
results should be documented and con-
tain the following:
1. Certification that the test was performed in accordance
with this test method.
2. Identification of specimens tested.
3. Volatile content % for each specimen tested and the aver-
age.
Procedure
Method B
5.2 Calibration
5.2.1
Prepare
two liquid calibration mixtures as directed per
vendor instructions. The mixtures do not need to have these
exact amounts but the amount of each ingredient should be
measured precisely to within 0.0002 grams. The mixture
should be stored in a suitable container with a resealable
septa lid. Calibration mixtures should be appropriate for vola-
tiles present in product tested.
5.2.1.2
Make
sure the mixtures are each well mixed, then
determine the density of the mixtures by any acceptable
method.
5.2.1.3
From
each of the mixtures inject a minimum of four
different amounts (0.5 to 3.0 microliters each) into separate,
previously prepared, empty, sealed vials.
5.2.1.4
Allow
each vial to equilibriate in the headspace unit,
then inject each into the chromatograph and obtain an area
for each sample.
5.2.1.5
Plot
the areas vs. the amount (in milligrams) of each
ingredient. The eight points should all lie on the same straight
line.
5.2.1.6
Determine
the response factor Rf (milligrams/area
unit) for each ingredient. If the plot above is not linear or does
not pass through the origin, then it might be necessary to
have different Rf at different amounts.
5.2.2
Sample
5.2.2.1
Cut
and weigh a 25.4 mm square sample of mate-
rial to be tested. Place into sample vial and seal tightly. Equili-
brate sample in the headspace unit and then inject onto the
chromatography column. Allow the chromatogram to run for
eight minutes.
Note 1: Appropriate time and temperature dependent on
product tested.
5.2.2.2
Calculate
the amount of each calibrated peak using
the following equation.
Volatile % (i) = (Area (i) * Rf (i) * 100)/(Wt * F)
where: Area (i) = area generated for ith calibrated peak
Rf (i) = response factor for ith peak
Wa = weight of sample (not including substrate—
subtract out substrate weight by calculating
the area x the density of the substrate under
test (W
s
)
and subtracting the value from the
total sample weight (Wt). Wt −Ws = Wa)
F = fraction of ingredient in film that reaches the
chromatography relative to the same frac-
tion of pure liquid.
5.2.3
Determination of F
5.2.3.1 Determination of F__liq
Prepare
a mixture of
ingredients to be tested. Inject 1 to 2 microliters of the mixture
into a sealed vial. Equilibrate in the head space unit, then
inject into the chromatograph and obtain the area for each
peak of interest.
5.2.3.1.1
Immediately
after the injection is complete, vent
the vial by means of a small syringe needle and replace the
vial to allow it to equilibrate again. Again after equilibration
inject the vial into the chromatograph.
5.2.3.1.2
Vent
the vial a second time and equilibrate and
inject again. There should have been three injections, each
preceded by an equilibration and two ventings, one immedi-
ately after the first injection and the other immediately after the
second injection.
5.2.3.1.3
Calculate
F__ret__liq and F__liq for each ingredi-
ent as follows.
F__ret__liq = (square root (area3/area2) + area2/area1)/2
F__liq = 1 F__ret__liq
IPC-TM-650
Number
2.3.37
Subject
Volatile
Content of Adhesive Coated Dielectric Films
Date
5/98
Revision
B
P
age2of3
电子技术应用       www.ChinaAET.com
5.2.3.2
Determination of F__film
Prepare
a film sample in
a vial as standard. Make three injections with venting between
injections and equilibration before each one, just as in the
case of the liquid. The F__ret__film and F__film are calculated
in exactly the same way as F__ret__liq and F__film, except
that the areas used are generated by the film sample.
5.2.3.3
F__liq
and F__film are determined at least eight
times, preferably over a period of several days. Approximately
the same number of liquids as films are tested each day. The
values of F__liq and F__film are averaged separately.
5.2.3.5
Once
determined, F should not change unless the
ingredients change or unless the equilibrium conditions
change.
5.2.3.6
Different
materials will have different values of F.
6 Notes
Multiple
Headspace Extraction—A Procedure for Quantita-
tive Analysis of Volatile Compounds in Solid Samples
,
Kolb,
Auer, and Pospisil, ANGEWANTE CHROMATOGRAPHIE—
APPLIED CHROMATOGRAPHY, January, 1981.
IPC-TM-650
Number
2.3.37
Subject
Volatile
Content of Adhesive Coated Dielectric Films
Date
5/98
Revision
B
P
age3of3
电子技术应用       www.ChinaAET.com
1.0
Scope
This
test method establishes a procedure for
determining the thermal decomposition temperature of
organic films using thermogravimetric analysis (TGA).
2.0
Applicable Documents
ASTM D 618
Standard
Practice for Conditioning Plastics and
Electrical Insulating Materials for Testing
ASTM
D 3850
Standard
Test Method for Rapid Thermal
Degradation of Solid Electrical Insulating Materials by Thermo-
gravimetric Analysis
3.0
Test Specimen
The
free films are cut to fit into the
TGA sample pan giving an initial weight of nominally 2-20 mg.
4.0
Apparatus or Material
Thermal
gravimetric analysis
equipment equipped with air, nitrogen purge, capable of
maintaining 5°C heating rate from ambient to 800°C, such as
TA Instruments or Perkin Elmer TGA instruments.
5.0
Procedure
5.1
The
test specimens should be conditioned at 23 ± 2°C
and 50 ± 5% relative humidity for not less than 24 hours prior
to testing. Refer to ASTM D 618 and ASTM D 3850.
5.2
Follow
the manufacturer’s recommendations for equip-
ment startup and calibration.
5.3
Set
the dry nitrogen purge rate at 55 cc/min.
5.4
Allow
the instrument to purge for 30 min.
5.5
Heat
at a rate of 5°C/min from ambient to 150°C.
5.6
Hold
the temperature for 15 min to remove any moisture
from the specimen.
5.7
Heat
at a rate of 5°C/min to 800°C.
5.8
Record
the temperature at which 5% additional weight
loss occurs based on the weight after the 150°C hold.
5.9 Repeat
steps 5.3-5.6 with dry air instead of nitrogen
using a new sample.
5.10
Report
the thermal decomposition temperature, T
max
,
in
degrees Celsius (°C), followed by the evaluation criteria
(e.g., 5% wt. loss), and the atmosphere under which the test
was conducted (e.g., nitrogen), e.g. 185°C (5% wt. loss, nitro-
gen).
6.0
Notes
6.1
Calibration
of the instrument must be carried out
according to the manufacturer’s recommendations. The rate
of flow of the gas in the cell will have a significant effect on the
calibration, therefore the instrument must be calibrated with
the same flow rate used during the test. After the instrument
has been calibrated, the flow rate should not be changed.
6.2
The
surface area of the test specimen will affect the
results. If two specimens of the same material are tested, one
with a large surface area and one with a small surface area,
both of the same mass, the specimen with the smaller surface
area will normally lose weight at a slower rate.
6.3
The
specimen may initially gain weight due to buoyancy
effects.
The
Institute for Interconnecting and Packaging Electronic Circuits
2215 Sanders Road Northbrook, IL 60062-6135
IPC-TM-650
TEST
METHODS MANUAL
Number
2.3.40
Subject
Thermal
Stability
Date
7/95
Revision
Originating Task Group
Deposited Dielectric Task Group (C-13a)
Material
in this Test Methods Manual was voluntarily established by Technical Committees of the IPC. This material is advisory only
and its use or adaptation is entirely voluntary. IPC disclaims all liability of any kind as to the use, application, or adaptation of this
material. Users are also wholly responsible for protecting themselves against all claims or liabilities for patent infringement.
Equipment referenced is for the convenience of the user and does not imply endorsement by the IPC.
P
age1of1
电子技术应用       www.ChinaAET.com