IPC-TM-650 EN 2022 试验方法.pdf - 第365页

11.1 The report shall include the following: 11.1.1 Designation of the material, including the name of the manufacturer and information on composition when known. 11.1.2 Method of preparation of the test specimen. 11.1.3…

100%1 / 824
Note
2—
Instruments
from duPont and Perkin Elmer have
been found suitable.
6.0
Test Specimens
6.1
The test specimen shall be between .05 and 0.3 inches
thick. This thickness may be as received or may be laminated
by the user from pre-impregnated ‘‘B’’ stage and copper free
‘‘C’’ stage material. It laminated by the user, the user shall be
responsible to contact the manufacturer for the exact layup
and process parameters used for quality acceptance at the
manufacturers facility.
Note
3—
Repeatability
of Test Results will vary with layup,
bake out, laminating pressure/ramp speed, press time, etc.
6.2 Specimens
should be between 0.3 and 0.4 inches in
height and have flat and parallel upper and lower surfaces.
The surfaces to be measured shall be perpendicular to the
fiber fillers and the identity of the direction of the fiber fillers
shall be maintained throughout the test. The upper and lower
surfaces shall be polished with 600 grit paper to remove burrs
or strands of fiber filler. The specimens shall then be cleaned
using isopropyl alcohol, and dried for 1 hour at 10°C above
the maximum specified temperature of the run.
Note
4—
The
1 hour prebake may be eliminated if Condition
(7.), is performed immediately after final polish.
6.3
There
shall be three specimens prepared from the same
piece of material for each direction to be measured.
7.0
Conditioning
7.1
Conditioning
of test specimen shall include immersion in
isopropyl alcohol with agitation for 20 seconds, followed by
Condition E-1/110 and C
1
40/23/50
in accordance with
D-618.
8.0
Calibration
8.1
Calibrate
the apparatus in accordance with the instru-
ment manufacturer’s recommendations.
9.0
Procedure
9.1
Measure
the height of the specimen.
9.2 Place
the specimen in the specimen holder under the
probe. The thermocouple or other means for sensing speci-
men temperature should be in contact with the specimen, or
as near to the specimen as possible.
9.3
Assemble
the furnace to the specimen holder. If mea-
surements at subambient temperatures are to be made, cool
the specimen holder and furnace to at least 20°C below the
lowest temperature of interest, using procedures as given by
the instrument manufacturer. The refrigerant used for cooling
shall not come into direct contact with the specimen.
Note
5—
The
temperature range to be tested shall be speci-
fied by the user, so that the manufacturer and user will test
over the same temperature range. If tested over different tem-
perature ranges, the repeatability may be unacceptable.
9.4
Place
weights on the sensing probe to ensure that the
probe is in contact with the specimen witha1to3-gload.
9.5
Increase
the furnace temperature at 5 = 0.5°C/min.
over the desired temperature range.
9.6
Record
the specimen temperature and change in speci-
men height using appropriate ranges on the X-Y recorder.
Note
6—
A
gas purge may be used to replace the air around
the specimen for measurement of expansion in different atmo-
spheres.
9.7
Test
at least three specimens of the same material.
Retest of a specimen may be used only as reference and shall
not be treated as an independent test of a new specimen.
10.0
Calculation
10.1
Calculate
the average coefficient of thermal expan-
sions, α, over the temperature intervals of interest as follows:
α =( H/T)/H
where:
H = original height of specimen,
H = change in height of the specimen (in the same units)
over the temperature interval T, and
T = temperature interval, °C (see Figure 1).
Note
7—
H
and T may on some instruments be read
directly from the recorder chart. On other instruments con-
stant factors may need to be applied to the chart readings to
obtain these values.
11.0
Report
IPC-TM-650
Number
2.4.41
Subject
Coefficient
of Linear Thermal Expansion of Electrical Insulating
Materials
1
Date
3/86
Revision
P
age2of3
电子技术应用       www.ChinaAET.com
11.1
The
report shall include the following:
11.1.1
Designation
of the material, including the name of the
manufacturer and information on composition when known.
11.1.2
Method
of preparation of the test specimen.
11.1.3
Specimen
orientation with respect to original sample,
if applicable.
11.1.4
Sample
size.
11.1.5
Temperatures
between which the coefficient of linear
thermal expansion has been determined.
11.1.6
Average
coefficient of linear thermal expansion per
degree Celsius.
11.1.7
Transition
temperatures, if noted.
11.1.8
Instrument
manufacturer and model number.
11.1.9
Purge
gas, if used, and rate of gas flow, and
11.1.10
X-Y
chart record.
NOTE
The
preceding test method was originally ASTM
D3386-75, until modified for use by IPC for round-robin test-
ing of organic substrate materials. Upon completion of the test
program, recommendations for revision will be made to
ASTM.
IPC-2441-1
Figure
1 Specimen height versus temperature
IPC-TM-650
Number
2.4.41
Subject
Coefficient
of Linear Thermal Expansion of Electrical Insulating
Materials
1
Date
3/86
Revision
P
age3of3
电子技术应用       www.ChinaAET.com
1.0
Scope
1.1
To
describe the vitreous silica dilatometer method for
determining the linear thermal expansion of laminated materi-
als within the temperature range of –55°C to 100°C. Inorganic
substrates (non-laminated) shall be tested within a range of
–55° to 150°C.
2.0
Applicable Documents
ASTM-E-228
Standard
Test Method for Linear Thermal
Expansion of Solid Materials with a Vitreous Silica Dilatometer
ASTM-D-696
Test
for Coefficient of Linear Thermal Expan-
sion of Plastics
ASTM-E-831
Test
for Linear Thermal Expansion of Solid
Materials by Thermodilatometry
ASTM-E-77
Verification
and Calibration of Liquid-in-Glass
Thermometers
ASTM-E-220
Calibration
of Thermocouples by Comparison
Techniques
ASTM-E-644
Testing
Industrial Resistance Thermometers
3.0
Test Specimen
3.1
Laminated
materials which may or may not contain
metal layers.
3.2
Nominal
test specimen dimensions shall be 1/4 inch
wide x 2 inch –4 inch long x 1/8 inch minimum thickness. End
surfaces shall be ground parallel. Any deviation from nominal
should recognize thermal gradients of the temperature cham-
ber, thermal lag of specimen and any bending of specimen.
Thicknesses under 1/8 inch shall be supported by adequate
clamping devices unless it is certain that the specimen will
remain straight during testing.
4.0 Apparatus
4.1
Vitreous
silica dilatometer of either the tube or push rod
type to determine the change in length of a solid material as a
function of temperature. The temperature is controlled at a
constant heating or cooling rate. The linear thermal expansion
and the coefficients of linear thermal expansion (CTE) are cal-
culated from the recorded data.
This device measures the difference in thermal expansion
between a test specimen and the vitreous silica parts of the
dilatometer (Figure 1).
4.2
Specimen
holder (tube) and probe shall be made of vit-
reous silica. The probe contact shall be flat or be rounded to
approximately a 10 mm radius.
4.3
Chamber
for uniformly heating and cooling the speci-
men. The specimen temperature change rate shall be con-
trolled. The temperature gradient in the specimen shall not
exceed 0.5°C/cm.
4.4
Transducer,
for measuring the difference in length
between the specimen and the specimen holder with an
accuracy of at least ± 0.5µm. The transducer shall be pro-
tected or mounted so that temperature changes will not affect
the readings by more than 1.0µm.
4.5 Micrometer,
for measuring the reference length, L
o
,o
f
the specimen with an accuracy of at least ± 25µm.
4.6
Thermocouple,
types E, K, or T, for measurement of the
specimen temperature. (Type E is NiCr versus constantan,
type K is NiCr versus NiAl and Type T is Cu versus constan-
tan.)
4.7
Recorder
or data logger for collecting temperatures and
lengths.
5.0
Procedure
5.1 Sample Preparation
Rough
cut with a band saw or
metallurgical cut-off wheel and finish machining by grinding.
Care must be exercised to remove roughness from specimen
ends. The ends shall be parallel to ± .001 inch/inch.
5.2
Sample
condition (only for laminated, organic speci-
mens).
5.2.1
The
specimen shall be immersed in isopropyl alcohol
and agitated for twenty seconds.
The
Institute for Interconnecting and Packaging Electronic Circuits
2215 Sanders Road Northbrook, IL 60062-6135
IPC-TM-650
TEST
METHODS MANUAL
Number
2.4.41.1
Subject
Coefficient
of Thermal Expansion by the Vitreous
Silica (Quartz) Dilatometer Method
Date
8/97
Revision
A
Originating Task Group
N/A
Material
in this Test Methods Manual was voluntarily established by Technical Committees of the IPC. This material is advisory only
and its use or adaptation is entirely voluntary. IPC disclaims all liability of any kind as to the use, application, or adaptation of this
material. Users are also wholly responsible for protecting themselves against all claims or liabilities for patent infringement.
Equipment referenced is for the convenience of the user and does not imply endorsement by the IPC.
P
age1of3
电子技术应用       www.ChinaAET.com