IPC-TM-650 EN 2022 试验方法.pdf - 第491页
6.3.6.2 Probes for Coupled-Signal-Line (Differential) Transmission Line Measurements The probe consider- ations described in 4.3.3 apply for probes used in differential transmission line measurements. However, the necess…

where:
BW
-3dB
is
the 3 dB attenuation bandwidth and t
d
is
the 10 %
to 90 % transition duration of the TDR step response.
Note that this relationship may not accurately represent the
intended operational frequencies of the transmission line
being tested. The bandwidth and risetime characteristics must
be adequate to ensure that the TDR can provide a measure-
ment zone long enough to accurately determine Z
0
for
a
transmission line of a given length. This constant-valued
region corresponds to the round-trip propagation time of the
TDR step on the transmission line being tested. If the TDR
transition duration is too long relative to this propagation time,
the constant-valued region (also called the measurement
zone) will be very short thereby increasing measurement error
and uncertainty. Risetime considerations, however, are not
the best method for determining TDR resolution. It is better to
consider the temporal/spatial resolution of the TDR (see 4.1.2)
than bandwidth/risetime resolution when determining the per-
formance of the TDR measurement system.
6.3.2
Temporal/Spatial Resolution
The
TDR unit may
not be the only limiting factor for temporal resolution. The
probe connecting the TDR unit to the test specimen may also
limit resolution and this needs to be considered. Because of
the nature of TDR, it is easy to include the effects of the TDR
unit and all of the probe devices collectively, by defining t
sys
as
the
fall time of the TDR step that has reflected from a short
circuit placed at the end of the probe and returned to the TDR
head. The mean value of Z
0
is
less susceptible to TDR reso-
lution than are the minimum and maximum values.
6.3.3
Amplitude Scale
If
a coarse vertical scale is used,
quantization error can be significant. Many instruments
change accuracy when their scales are changed, and this can
result in significant but unknown errors that can exceed 3 Ω.
6.3.4
Baseline and Amplitude Drift
The
ability of the TDR
instrument to maintain a constant baseline voltage and con-
stant amplitude step pulse are critical to the repeatability of
the TDR measurement process. TDR step generators and
sampling units are sensitive to time and temperature drifts.
Drift should be minimized and have a value that corresponds
to less than one-tenth the desired impedance uncertainty.
6.3.5
Electrostatic Discharge Damage
ESD
damage to
TDR instrumentation is often not easily detected and may
unknowingly affect measurement accuracy. Therefore, system
calibration should be performed regularly to check for this (see
5.1.1). All cables should have a termination attached to one
end when not in use and while they are being connected to
the TDR instrumentation. The use of a static protection switch
helps eliminate ESD damage to the TDR. Operators should
have anti-static awareness training and should perform all
measurements in anti-static work areas while wearing anti-
static wrist straps.
6.3.6
Probes
Hand-held
(manual) probing solutions are
more sensitive to operator technique than are automated
methods and, consequently, the measurements made using
manual probing methods will exhibit higher variability than
automated methods. Operators should be trained and tested
in their ability to repeatably probe and obtain a consistent
measurement.
6.3.6.1
Probes for Single-Ended Transmission Line
Measurements
The
probe assembly impedance is often
chosen to be 50 Ω to match the impedance of the TDR sys-
tem. Impedance matching minimizes reflections at the inter-
face between the probe and the transmission line under test.
These reflections, which appear at and around the transition
region in the TDR pulse and can extend for some time after
this transition, are perturbations in the TDR waveform and are
undesirable because they affect the length of the measure-
ment zone and may increase measurement uncertainty. When
the characteristic impedance of the transmission line under
test is nominally 50 Ω, these perturbations will normally decay
rapidly. If the impedance of the transmission line under test is
significantly different from 50 Ω, the magnitude of the pertur-
bations can be large and their duration long enough to affect
the measurement zone. The effect of these perturbations
must be taken into account when determining the measure-
ment zone (see 4.1.2). The design and quality of manufacture
of the probe has a large effect on the magnitude and duration
of reflections generated between the TDR system and the
transmission line under test.
When probing non-50 Ω lines, it is possible to separate, in the
TDR waveform, the large signal perturbations caused by the
TDR/probe interface from those caused by the probe/
transmission line interface. To do this, a specially designed
probe is required that is impedance matched to the transmis-
sion line under test and that also has a long propagation delay
between the TDR/probe connection and the probe tip. The
long propagation delay can effectively move the large pertur-
bations at the TDR/probe interface out of the measurement
zone.
IPC-TM-650
Number
2.5.5.7
Subject
Characteristic
Impedance of Lines on Printed Boards by TDR
Date
03/04
Revision
A
Page
22 of 23
电子技术应用 www.ChinaAET.com

6.3.6.2
Probes for Coupled-Signal-Line (Differential)
Transmission Line Measurements
The
probe consider-
ations described in 4.3.3 apply for probes used in differential
transmission line measurements. However, the necessity to
simultaneously probe two signal lines and one or two refer-
ence plane contacts makes differential probing more difficult
than probing single signal line structures. In a PCB manufac-
turing environment, the use of two probes that were previ-
ously used for single-ended measurements may not be pos-
sible. This is because the operator is required to use both
hands for probing, which leaves them unable to operate the
instrument. Contact your instrument manufacturer for their
probing solutions and advice. Probes from one manufacturer
can also be used with another manufacturer’s TDR if the
impedance values and connectors are compatible.
6.4
Adjustable Measurement Parameters
6.4.1 Sampling Interval (Point Spacing)
The
temporal
resolution of the TDR unit is an issue only if it impacts the
duration of the constant-valued regions in the TDR waveform
(see 4.1.2) that are used for computing Z
0
.
The temporal reso-
lution of the TDR is affected by the transition duration of the
TDR step response, the transition duration of the step
response of all intervening electrical components (connectors,
cables, adapters), measurement jitter, the interval between
sampling instances, and timebase errors. For typical TDR
measurements, timebase errors and sampling intervals should
not be an issue (both are or can be made to be less than 10
ps). The effect of measurement jitter can be modeled by con-
volving the jitter distribution with the TDR step response to
yield an effective TDR step response. The effect of jitter on the
bandwidth of the TDR measurement can be assessed from
the jitter spectrum, which can be described by:
J(ƒ)=e
−2(πσƒ)
2
where:
J is
the jitter spectrum,
f is frequency, and
σ is the rms jitter value
If the effective step response impacts the duration of the mea-
surement zones, then jitter must be reduced. If the jitter has
an observable effect, then the user must reduce the duration
of the measurement zone (by increasing the lower limit and
decreasing the upper limit, (see 5.1.3) from which Z
0
is
com-
puted or reduce the system jitter. Reduction in the duration of
the measurement zone may introduce a bias in the voltage or
reflection coefficient values and this affect the computed value
of Z
0
.
If the rms jitter value is less than 20 % of the transition
duration of the TDR step response, then the jitter is small and
can be ignored. For typical TDR systems, however, rms jitter
is less than 10 ps and will not affect the Z
0
measurements.
Similarly,
the effect of cables, connectors, and adapters on
the measurement can be modeled by convolving their step
responses with that of the TDR unit. If the transition duration
of this new step response meets the requirements of 4.1.2,
then the performance of the cables, connectors, and adapters
is adequate.
6.4.2
Waveform Averaging and Number of Samples in
the Measurement Zone
Waveform
averaging reduces the
effective noise level of the measurement by M
-1/2
,
where M is
the number of acquired waveforms (typically, 8 ≤ M ≤ 256).
Consequently, averaging can reduce measurement noise.
This reduction is limited by the number of bits of the analog-
to-digital converter of the TDR system. However, if the TDR-
system exhibits drift in the timebase, averaging too many
waveforms may result in a reduction of t
sys
and
a commensu-
rate reduction in the temporal/spatial resolution of the TDR.
The number of samples (data points) in the measurement
zone will affect the standard deviation of the computed value
of Z
0
because
this value is the result of averaging all the
samples in the measurement zone. Therefore, the more
samples in the measurement zone, the smaller will be the
standard deviation of the computed Z
0
value.
6.4.3
Selection of Constant-Valued Region (Measure-
ment Zone)
Inconsistency
in defining where the constant-
valued region is located in the TDR waveform may cause a
significant but unknown error than can exceed 5.0 Ω. Speci-
fying the measurement zone improves measurement repeat-
ability of the same or similar samples, and this can improve
assessment of design and fabrication quality and vendor
capability. This measurement zone should be far enough
away from the launch and the open end of the transmission
line under test to minimize the effects of these discontinuities.
The measurement zone is to be given as the separation
between two positions on the transmission line, and these
positions are to be given as a percentage of the transmission
line length referenced from the TDR/transmission line inter-
face. The measurement zone is defined in 5.1.3.
6.5
Acknowledgments
The
majority of the figures used
herein were provided by Mr. Bryan C. Parker of the Introbot-
ics Corporation, Albuquerque, NM.
IPC-TM-650
Number
2.5.5.7
Subject
Characteristic
Impedance of Lines on Printed Boards by TDR
Date
03/04
Revision
A
Page
23 of 23
电子技术应用 www.ChinaAET.com

1
Scope
This
procedure outlines a test method to deter-
mine the permittivity (dielectric constant or E’r) and loss tan-
gent (dissipation factor or Tanδ) of printed wiring materials at
various frequencies (from 1 MHz to 1.5 GHz) using a single
test fixture for the measurement.
The permittivity and loss tangent are measured using a narrow
sweep of frequency around the target or desired frequency.
The test method is built around the capability of currently
available materials analyzers, which use a capacitance
method to determine permittivity.
This test method is not intended for low loss materials, such
materials may be tested at fixed frequencies using other IPC
test methods.
2
Applicable Documents
HP 4291A-5 Product Note
‘‘Dielectric
Constant Evaluation
of Rough Surface Materials,’’ which describes how to make
accurate measurements using the HP 4291A and HP
16453A.
HP
Application Note 380-1
‘‘Dielectric
Constant Measure-
ments of Solid Materials,’’ which contains a technical back-
ground, suitable for this subject.
3
Test Specimen
3.1
Each
specimen shall be 50 mm x 50 mm by the thick-
ness of the substrate material. Within the limits of the test fix-
ture, the thicker the sample the less error in the measure-
ments. Multilayer samples can be used to increase the
thickness of the sample, but these cannot be simple stacked
layers; they must be physically bonded with no air gaps
between the layers. A target thickness would be 1.0 mm, but
both thinner and thicker samples will work.
3.2
Three
specimens are required for this test.
3.3
All
materials are affected by moisture, including all rein-
forced laminates and most films. Therefore, all samples shall
be conditioned at 23°C ± 2°C and 50% RH ± 5% RH for a
minimum of 24 hours prior to testing. However, if a sample
has recently been etched or exposed to excessive moisture, it
should be dried in an air-circulating oven for two hours at
105°C +5°C, -2°C prior to testing and conditioned at room
temperature as mentioned above.
3.4
Sample Surface Preparation
3.4.1
It
is preferred that the sample be patterned with a
conductive material in the shape and size of the test elec-
trode. This conductive material is preferably 100 angstroms of
vapor deposited copper. Other metals may be used. In all
cases, the conductor on the sample must make good electri-
cal contact with the fixture electrode. Such a conductive pat-
tern eliminates air gaps and other potential sample mounting
errors.
3.4.2
Bare
dielectric materials may be tested with this test
method. The fixture electrodes must be applied with some
level of force to ensure a gap-free contact area. Determining
the correct force setting may require some trial and error test-
ing for each type of sample (see 6.4).
4
Equipment/Apparatus
4.1
The
Hewlett-Packard Impedance Material Analyzer,
model 4291A, or equivalent is recommended.
4.2 Hewlett-Packard
model number 16453A test fixture, or
equivalent
4.3
An
appropriate calibration-verification kit and a fixture-
correction kit as recommended in the instrument’s manual
(i.e., HP4291A Calibration kit). Such a kit usually includes the
following devices:
• OPEN and SHORT for fixture correction
• 50 Ohms impedance
• Dielectric (PTFE) of known characteristic for the purpose of
the calibration verification
4.4
Micrometer,
capable of 0.001 mm resolution
4.5
Circulating
oven capable of 105°C +5°C, -2°C
5
Procedure
5.1
Calibrate
the instrument using the calibration kit accord-
ing to the recommendations of the instrument manufacturer.
The
Institute for Interconnecting and Packaging Electronic Circuits
2215 Sanders Road • Northbrook, IL 60062
IPC-TM-650
TEST
METHODS MANUAL
Number
2.5.5.9
Subject
Permittivity
and Loss Tangent, Parallel Plate,
1 MHz to 1.5 GHz
Date
11/98
Revision
Originating Task Group
HDI Test Methods Task Group (D-42a)
Material
in this Test Methods Manual was voluntarily established by Technical Committees of the IPC. This material is advisory only
and its use or adaptation is entirely voluntary. IPC disclaims all liability of any kind as to the use, application, or adaptation of this
material. Users are also wholly responsible for protecting themselves against all claims or liabilities for patent infringement.
Equipment referenced is for the convenience of the user and does not imply endorsement by the IPC.
P
age1of5
电子技术应用 www.ChinaAET.com