IPC-TM-650 EN 2022 试验方法.pdf - 第492页
1 Scope This procedure outlines a test method to deter- mine the permittivity (dielectric constant or E’r) and loss tan- gent (dissipation factor or Tan δ ) of printed wiring materials at various frequencies (from 1 MHz …

6.3.6.2
Probes for Coupled-Signal-Line (Differential)
Transmission Line Measurements
The
probe consider-
ations described in 4.3.3 apply for probes used in differential
transmission line measurements. However, the necessity to
simultaneously probe two signal lines and one or two refer-
ence plane contacts makes differential probing more difficult
than probing single signal line structures. In a PCB manufac-
turing environment, the use of two probes that were previ-
ously used for single-ended measurements may not be pos-
sible. This is because the operator is required to use both
hands for probing, which leaves them unable to operate the
instrument. Contact your instrument manufacturer for their
probing solutions and advice. Probes from one manufacturer
can also be used with another manufacturer’s TDR if the
impedance values and connectors are compatible.
6.4
Adjustable Measurement Parameters
6.4.1 Sampling Interval (Point Spacing)
The
temporal
resolution of the TDR unit is an issue only if it impacts the
duration of the constant-valued regions in the TDR waveform
(see 4.1.2) that are used for computing Z
0
.
The temporal reso-
lution of the TDR is affected by the transition duration of the
TDR step response, the transition duration of the step
response of all intervening electrical components (connectors,
cables, adapters), measurement jitter, the interval between
sampling instances, and timebase errors. For typical TDR
measurements, timebase errors and sampling intervals should
not be an issue (both are or can be made to be less than 10
ps). The effect of measurement jitter can be modeled by con-
volving the jitter distribution with the TDR step response to
yield an effective TDR step response. The effect of jitter on the
bandwidth of the TDR measurement can be assessed from
the jitter spectrum, which can be described by:
J(ƒ)=e
−2(πσƒ)
2
where:
J is
the jitter spectrum,
f is frequency, and
σ is the rms jitter value
If the effective step response impacts the duration of the mea-
surement zones, then jitter must be reduced. If the jitter has
an observable effect, then the user must reduce the duration
of the measurement zone (by increasing the lower limit and
decreasing the upper limit, (see 5.1.3) from which Z
0
is
com-
puted or reduce the system jitter. Reduction in the duration of
the measurement zone may introduce a bias in the voltage or
reflection coefficient values and this affect the computed value
of Z
0
.
If the rms jitter value is less than 20 % of the transition
duration of the TDR step response, then the jitter is small and
can be ignored. For typical TDR systems, however, rms jitter
is less than 10 ps and will not affect the Z
0
measurements.
Similarly,
the effect of cables, connectors, and adapters on
the measurement can be modeled by convolving their step
responses with that of the TDR unit. If the transition duration
of this new step response meets the requirements of 4.1.2,
then the performance of the cables, connectors, and adapters
is adequate.
6.4.2
Waveform Averaging and Number of Samples in
the Measurement Zone
Waveform
averaging reduces the
effective noise level of the measurement by M
-1/2
,
where M is
the number of acquired waveforms (typically, 8 ≤ M ≤ 256).
Consequently, averaging can reduce measurement noise.
This reduction is limited by the number of bits of the analog-
to-digital converter of the TDR system. However, if the TDR-
system exhibits drift in the timebase, averaging too many
waveforms may result in a reduction of t
sys
and
a commensu-
rate reduction in the temporal/spatial resolution of the TDR.
The number of samples (data points) in the measurement
zone will affect the standard deviation of the computed value
of Z
0
because
this value is the result of averaging all the
samples in the measurement zone. Therefore, the more
samples in the measurement zone, the smaller will be the
standard deviation of the computed Z
0
value.
6.4.3
Selection of Constant-Valued Region (Measure-
ment Zone)
Inconsistency
in defining where the constant-
valued region is located in the TDR waveform may cause a
significant but unknown error than can exceed 5.0 Ω. Speci-
fying the measurement zone improves measurement repeat-
ability of the same or similar samples, and this can improve
assessment of design and fabrication quality and vendor
capability. This measurement zone should be far enough
away from the launch and the open end of the transmission
line under test to minimize the effects of these discontinuities.
The measurement zone is to be given as the separation
between two positions on the transmission line, and these
positions are to be given as a percentage of the transmission
line length referenced from the TDR/transmission line inter-
face. The measurement zone is defined in 5.1.3.
6.5
Acknowledgments
The
majority of the figures used
herein were provided by Mr. Bryan C. Parker of the Introbot-
ics Corporation, Albuquerque, NM.
IPC-TM-650
Number
2.5.5.7
Subject
Characteristic
Impedance of Lines on Printed Boards by TDR
Date
03/04
Revision
A
Page
23 of 23
电子技术应用 www.ChinaAET.com

1
Scope
This
procedure outlines a test method to deter-
mine the permittivity (dielectric constant or E’r) and loss tan-
gent (dissipation factor or Tanδ) of printed wiring materials at
various frequencies (from 1 MHz to 1.5 GHz) using a single
test fixture for the measurement.
The permittivity and loss tangent are measured using a narrow
sweep of frequency around the target or desired frequency.
The test method is built around the capability of currently
available materials analyzers, which use a capacitance
method to determine permittivity.
This test method is not intended for low loss materials, such
materials may be tested at fixed frequencies using other IPC
test methods.
2
Applicable Documents
HP 4291A-5 Product Note
‘‘Dielectric
Constant Evaluation
of Rough Surface Materials,’’ which describes how to make
accurate measurements using the HP 4291A and HP
16453A.
HP
Application Note 380-1
‘‘Dielectric
Constant Measure-
ments of Solid Materials,’’ which contains a technical back-
ground, suitable for this subject.
3
Test Specimen
3.1
Each
specimen shall be 50 mm x 50 mm by the thick-
ness of the substrate material. Within the limits of the test fix-
ture, the thicker the sample the less error in the measure-
ments. Multilayer samples can be used to increase the
thickness of the sample, but these cannot be simple stacked
layers; they must be physically bonded with no air gaps
between the layers. A target thickness would be 1.0 mm, but
both thinner and thicker samples will work.
3.2
Three
specimens are required for this test.
3.3
All
materials are affected by moisture, including all rein-
forced laminates and most films. Therefore, all samples shall
be conditioned at 23°C ± 2°C and 50% RH ± 5% RH for a
minimum of 24 hours prior to testing. However, if a sample
has recently been etched or exposed to excessive moisture, it
should be dried in an air-circulating oven for two hours at
105°C +5°C, -2°C prior to testing and conditioned at room
temperature as mentioned above.
3.4
Sample Surface Preparation
3.4.1
It
is preferred that the sample be patterned with a
conductive material in the shape and size of the test elec-
trode. This conductive material is preferably 100 angstroms of
vapor deposited copper. Other metals may be used. In all
cases, the conductor on the sample must make good electri-
cal contact with the fixture electrode. Such a conductive pat-
tern eliminates air gaps and other potential sample mounting
errors.
3.4.2
Bare
dielectric materials may be tested with this test
method. The fixture electrodes must be applied with some
level of force to ensure a gap-free contact area. Determining
the correct force setting may require some trial and error test-
ing for each type of sample (see 6.4).
4
Equipment/Apparatus
4.1
The
Hewlett-Packard Impedance Material Analyzer,
model 4291A, or equivalent is recommended.
4.2 Hewlett-Packard
model number 16453A test fixture, or
equivalent
4.3
An
appropriate calibration-verification kit and a fixture-
correction kit as recommended in the instrument’s manual
(i.e., HP4291A Calibration kit). Such a kit usually includes the
following devices:
• OPEN and SHORT for fixture correction
• 50 Ohms impedance
• Dielectric (PTFE) of known characteristic for the purpose of
the calibration verification
4.4
Micrometer,
capable of 0.001 mm resolution
4.5
Circulating
oven capable of 105°C +5°C, -2°C
5
Procedure
5.1
Calibrate
the instrument using the calibration kit accord-
ing to the recommendations of the instrument manufacturer.
The
Institute for Interconnecting and Packaging Electronic Circuits
2215 Sanders Road • Northbrook, IL 60062
IPC-TM-650
TEST
METHODS MANUAL
Number
2.5.5.9
Subject
Permittivity
and Loss Tangent, Parallel Plate,
1 MHz to 1.5 GHz
Date
11/98
Revision
Originating Task Group
HDI Test Methods Task Group (D-42a)
Material
in this Test Methods Manual was voluntarily established by Technical Committees of the IPC. This material is advisory only
and its use or adaptation is entirely voluntary. IPC disclaims all liability of any kind as to the use, application, or adaptation of this
material. Users are also wholly responsible for protecting themselves against all claims or liabilities for patent infringement.
Equipment referenced is for the convenience of the user and does not imply endorsement by the IPC.
P
age1of5
电子技术应用 www.ChinaAET.com

Calibrations
only last 24 hours, so calibration shall be per-
formed within 24 hours of the measurement. See 6.1 for more
calibration notes.
5.2 Set up the unit to sweep the target frequency ± 0.5% of
the target.
5.3
Measure
the sample thickness with the micrometer and
insert into the test fixture. The sample must make good con-
tact with the fixture electrodes (see 6.1 concerning the proper
force to be applied). The sample must not touch the back wall
of the fixture. The sample electrode placement and thickness
measurement shall be obtained from the same area of the
sample.
5.4
Run
the test and record the average permittivity and loss
over the narrow frequency range sweep. The scanned data
may also be saved on disk. See 6.2 for comments on
expected behavior for permittivity as a function of frequency.
5.5
Repeat
5.1 through 5.4 for all desired frequencies.
5.6
Report
the average permittivity and loss at the frequen-
cies requested.
6 Notes
6.1
Correct
calibration and operation of the test equipment
is required to obtain accurate measurement of permittivity and
loss. Proper sample preparation is also very important for
obtaining useful data from this test. Calibrate the materials
analyzer in accordance with manufacturer’s instructions. An
automatic program has been developed for the HP 4291A,
which will ease calibration and setup (see 6.6).
6.2
The
permittivity should decrease slightly with increasing
frequency. If it increases greatly or decreases more than 0.2
units from approximately 20 MHz to 1.2 GHZ, reposition
(reset) the sample in the fixture and measure again (check for
debris between the electrodes; blow out with air).
6.3
Testing
at temperatures and humidities other than room
temperature may be performed with this instrument, as a spe-
cialized fixture can be placed in a temperature chamber. A
temperature chamber must be used when testing with this
fixture under conditions where condensation might contami-
nate the electrodes, as such contamination gives spurious
results.
6.4
The
pressure of the test fixture on the specimen affects
the measured permittivity and loss values, in particular for
un-metallized test specimens. Too light of pressure reduces
the area of electrode/sample contact, thus leaving air gaps,
which result in erroneous measurements. If the pressure is too
high, the sample can be reduced in thickness and the mea-
sured values would be incorrect because the thickness is
unknown. Making measurements while adjusting the force
should lead the operator to a force setting where the mea-
sured value is independent of the force applied.
6.5
The
HP 4291 Materials Analyzer and related fixtures and
calibration kits are available from Hewlett Packard, (800) 452-
4844.
6.6
Reference Program for Automatic Calibration and
Operation
This
automatic calibration and operation pro-
gram was developed for the HP 4291A and is published in
this method as a reference. Although the program listed in this
section has been tested and used, it is given here for ‘‘refer-
ence only.’’
6.6.1
Procedure Using Automatic Program
Turn
on the
analyzer with the program/calibration disk in the drive and fol-
low the user friendly calibration instructions, which appear on
the monitor. The calibration on the HP4291 lasts about 24
hours; after that, it begins to drift and provides slightly higher
values as a function of time. The calibration procedure in
6.6.1.1 through 6.6.1.10 should therefore be performed, at a
minimum, on a daily basis. During the calibration procedure,
the line traces on the monitor should be observed during each
step. Noisy or erratic traces are an indication of external inter-
ference. If noise is observed, the calibration procedure should
be aborted and rerun. Clean the electrode on the test head,
standards, and fixture connections and electrodes on a regu-
lar (weekly) basis. Blow dry.
6.6.1.1
Allow
at least 30 minutes for the unit to warm up
and stabilize.
6.6.1.2
As
the unit is a frequency sweeping unit, enter the
start and stop frequency (in megahertz) of the test (single fre-
quency tests use close start/stop frequencies, which should
be ± 0.5% of the target frequency).
6.6.1.3 Place
the OS (open) calibration standard on the test
head as prompted on the unit’s monitor and press Return on
the unit’s keyboard or ‘‘x1’’ on the unit.
IPC-TM-650
Number
2.5.5.9
Subject
Permittivity
and Loss Tangent, Parallel Plate,
1 MHz to 1.5 GHz
Date
11/98
Revision
P
age2of5
电子技术应用 www.ChinaAET.com