FUJI错误代码查询.pdf - 第1787页

Used by: Cause: A range error occurred during vision processing, and the processing was cancelled. Remedy: A range error occurs when a check is performed on a part but the check point is not on the image. This could be c…

100%1 / 5282
------------------------------
Code: 1D66F???
Display:
Used by:
Cause: There was a logical error in the management program.
Remedy: Check Part data.
------------------------------
Code: 1D670001
Display:
Used by:
Cause: A range error occurred during vision processing.
Remedy: A range error occurs when a check is performed on a part but the check point is not on the
image. This could be caused by
1) a large part being picked incorrectly, resulting in an incomplete image;
2) an incorrect CCD level setting or broken nozzle, resulting in a bad image;
3) an incorrect value entered in Part data (PDG) so that vision processing could not be done;
4) the Proper values CCD_X and CCD_Y are incorrect, resulting in a bad image;
5) a problem with the vision processing system program. Check the above listed conditions.
------------------------------
Code: 1D670002
Display:
Used by:
Cause: The part could not be found.
Remedy: Check the condition of the part feeding system. Check the CCD level and the nozzle sizes
entered in Proper data. When using a 1608 part S nozzle (1.00 mm) check whether or not the normal size
(1.3 mm) is entered. If this is the case and a 1608 part is picked, vision processing might register this as
not a part. If the CCD level is too high, the image will be too dark, and vision processing might register
this as not a part.
------------------------------
Code: 1D670006
Display:
Used by:
Cause: The part was tilted more than 45 deg. and processing was therefore cancelled.
Remedy: Check that the picking conditions have not gone bad.
------------------------------
Code: 1D670007
Display:
Used by:
Cause: A range error occurred during vision processing.
Remedy: None.
------------------------------
Code: 1D670008
Display:
Used by:
Cause: The lead information in Part data (PDG) was checked, and there are not more than two leads on
the appropriate side.
Remedy: Check the following items in PDG: number of leads, lead information and the orientation of the
parts on the feeder.
------------------------------
Code: 1D670009
Display:
第 1786 页
Used by:
Cause: A range error occurred during vision processing, and the processing was cancelled.
Remedy: A range error occurs when a check is performed on a part but the check point is not on the
image. This could be caused by
1) a large part being picked incorrectly, resulting in an incomplete image;
2) an incorrect CCD level setting or broken nozzle, resulting in a bad image;
3) an incorrect value entered in Part data (PDG) so that vision processing could not be done;
4) the Proper values CCD_X and CCD_Y are incorrect, resulting in a bad image;
5) a problem with the vision processing system program. Check the above listed conditions.
------------------------------
Code: 1D67000A
Display: 0010
Used by:
Cause: A range error occurred during vision processing, and the processing was cancelled.
Remedy: A range error occurs when a check is performed on a part but the check point is not on the
image. This could be caused by
1) a large part being picked incorrectly, resulting in an incomplete image;
2) an incorrect CCD level setting or broken nozzle, resulting in a bad image;
3) an incorrect value entered in Part data (PDG) so that vision processing could not be done;
4) the Proper values CCD_X and CCD_Y are incorrect, resulting in a bad image;
5) a problem with the vision processing system program. Check the above listed conditions.
------------------------------
Code: 1D67000B
Display: 0011
Used by:
Cause: The first lead could not be found during vision processing. A range error occurred during vision
processing, and the processing was cancelled.
Remedy: A range error occurs when a check is performed on a part but the check point is not on the
image. This could be caused by
1) a large part being picked incorrectly, resulting in an incomplete image;
2) an incorrect CCD level setting or broken nozzle, resulting in a bad image;
3) an incorrect value entered in Part data (PDG) so that vision processing could not be done;
4) the Proper values CCD_X and CCD_Y are incorrect, resulting in a bad image;
5) a problem with the vision processing system program. Check the above listed conditions.
------------------------------
Code: 1D67000C
Display: 0012
Used by:
Cause: The first lead could not be found during vision processing.
Remedy: Check that the part pickup conditions were OK.
------------------------------
Code: 1D67000D
Display: 0013
Used by:
Cause: Either the bright space exceeded the lead pitch by 50% or the dark space exceeded the lead width
by 50%. The dark space represents the actual lead width.
Remedy: Check the entries in Part data (PDG) for lead width, lead pitch and check the CCD level.
------------------------------
Code: 1D67000E
Display: 0014
Used by:
第 1787 页
Cause: The calculations of the lead locations on a leadless IC resulted in a range error.
Remedy: Calculate the leadless part at the PDG pitch item. Check the PDG lead pitch, lead information,
direction, and the orientation in which the parts are set on the machine.
------------------------------
Code: 1D67000F
Display: 0015
Used by:
Cause: Either the bright space exceeded the lead pitch by 50% or the dark space exceeded the lead width
by 50%. The dark space represents the actual lead width.
Remedy: Check the entries in Part data (PDG) for lead width, lead pitch and check the CCD level. The
Pitch limit item is entered in percentage points. An entry of 0 will give a default of 30%.
------------------------------
Code: 1D670010
Display: 0016
Used by:
Cause: The first lead could not be found during vision processing. A range error occurred during vision
processing, and the processing was cancelled.
Remedy: A range error occurs when a check is performed on a part but the check point is not on the
image. This could be caused by
1) a large part being picked incorrectly, resulting in an incomplete image;
2) an incorrect CCD level setting or broken nozzle, resulting in a bad image;
3) an incorrect value entered in Part data (PDG) so that vision processing could not be done;
4) the Proper values CCD_X and CCD_Y are incorrect, resulting in a bad image;
5) a problem with the vision processing system program. Check the above listed conditions.
------------------------------
Code: 1D670011
Display: 0017
Used by:
Cause: A range error occurred while checking for the ends of the leads.
Remedy: Check the Part data (PDG) lead length, pickup point, and the vision processing conditions.
------------------------------
Code: 1D670012
Display: 0018
Used by:
Cause: The part was moved to where under 60% of the body width has no leads and a range error
occurred.
Remedy: None.
------------------------------
Code: 1D670016
Display: 0022
Used by:
Cause: The machine decides whether or not there are leads on the part. The machine checked half the
part and found no leads, and this error occurred. Alternately, when a bright area is searched for after a
lead, no bright area is found for more than half the part.
Remedy: Check the body size and direction in Part data (PDG).
------------------------------
Code: 1D670017
Display: 0023
Used by:
Cause: When the bumper was recognized the leads were searched for from the center and a range error
occurred.
第 1788 页