semi合集-English.pdf - 第255页

SEMI E43-0301 © SEMI 1995 , 2001 3 7.2.1.3 Ch ar ge the reference calibratio n c apacito r to 1 volt with a charging source (power su pply). Calculate the am o unt of charge on t he capacitor by multiplying the v o ltage…

100%1 / 7923
SEMI E43-0301 © SEMI 1995, 2001 2
operation as their surfaces may be at elevated potentials
that represent a shock hazard to the operator.
7 Equipment and Performan ce Verification
Methods
7.1 Equipment
7.1.1 Electrostatic Locator/Field Sensor/Field Meter
— An electrostatic fieldmeter measures the value of the
electrostatic field at its sensor. Electrostatic fieldmeters
are calibrated and recommended for use at a particular
distance from the charged object. Fieldmeters are best
suited for making general surveys or audits, for making
measurements of surfaces at very high potentials
(charge levels), and for making measurements when
long-term stability is not important. They are not well
suited for measurements of surfaces with very low
potentials or when high spatial resolution of the surface
potential is needed.
7.1.1.1 The electrostatic locator/field sensor/field
meter will henceforth be referred to as “the fieldmeter.”
Note that for measurements to be taken in the presence
of air ionization, a chopper stabilized fieldmeter is
required. The fieldmeter must be capable of making
field measurements at a distance of 2.54 centimeters
(cm) = 1 inch or less, from the field source to the sensor
for this guide, as written. However, see Section 7.2.5
for fieldmeters that are operated at fixed distance(s),
and adjust values in this document where applicable.
7.1.2 Electrostatic Voltmeter An electrostatic
voltmeter nulls the electrostatic field at its sensor
(probe). An electrostatic voltmeter indicates the
presence and approximate level of the charge(s)
creating the electrostatic field. Under appropriate
conditions, electrostatic voltmeters provide a better
approximation of the charge level as compared to
electrostatic fieldmeters. Electrostatic voltmeters are
relatively free of drift and more environmentally stable
as compared to fieldmeters.
7.1.2.1 Electrostatic voltmeters are well suited for
fixed installation in equipment. Electrostatic voltmeters
exhibit a high degree of accuracy that is independent of
the distance from the charge. Thus, they are considered
better suited for making more accurate and repeatable
measurements as compared to fieldmeters. The probe
can be located very close to a charged surface without
arc-over, and, under appropriate conditions, can resolve
a small spatial area on a surface.
7.1.2.2 Electrostatic voltmeters are best suited for
making measurements of surfaces at potentials below
20kV, or when a calibrated or fixed distance from the
probe to the surface cannot be maintained. They are
also best suited for measuring low surface potentials, or
when it is desired to resolve a small area on the surface.
Electrostatic voltmeters are unsuitable for measuring
surfaces at very high potentials, such as above 20kV.
7.1.2.3 The electrostatic voltmeter will henceforth be
referred to as “the voltmeter.”
7.1.3 Electrometer An electrometer is a contact
voltmeter with a very high input impedance. Ideally,
this input impedance would be infinite. In practice, it is
limited by intrinsic physical materials properties of
insulators and by stray leakage paths between the input
terminals. Low voltage electrometers (below 200 Volt)
have typical input resistances of 10
14
ohms, accuracies
better than 0.1%, and can resolve microVolt type
potentials. High voltage electrometers (Kilovolts)
usually rely on resistive voltage dividers and have
typical input resistances in the 10
11
ohms range with
accuracies in the 1% range. It is important to evaluate
and understand the burden that the input impedance of
an electrometer represents when measuring voltage
potentials on very small charged structures.
7.1.4 Charged Plate Monitor A charged plate
monitor is an instrument typically used to monitor the
performance of air ionization equipment. Monitoring is
done with an electrically isolated 15 cm × 15 cm (6
inches × 6 inches) metal plate, henceforth referred to as
“the plate.” The instrument typically provides a means
to charge the plate to a known voltage (1000 or 5000
volts of either polarity), a plate sensor to determine the
voltage on the plate, and timing circuitry to determine
the time required to discharge the plate to a percentage
of its initial charge. For the purposes of this guide, the
charged plate monitor, or a separate isolated plate
assembly, can be used for performance verification
purposes as explained in Section 7.2.
7.2 Equipment Performance Verification (Confidence
Test)
7.2.1 Performance Verification of a Coulombmeter
Refer to Figure 1.
7.2.1.1 Zero the coulombmeter prior to each
measurement.
7.2.1.2 Maintain a reference calibration capacitor. It
should be a polystyrene or polypropylene 10 nF
capacitor (Mallory SX-110 or equivalent). Measure the
value of the capacitor to better than 1%. It is important
to handle the reference calibration capacitor very
carefully. Do not hold the capacitor by its body or
discharge it by touching both leads with the fingers.
Hold the capacitor by one lead only. Use a clip lead
connected between ground and this lead of the capacitor
to maneuver the other lead of the capacitor between the
“hot” side of the charging source and the input terminal
of the coulombmeter.
SEMI E43-0301 © SEMI 1995, 20013
7.2.1.3 Charge the reference calibration capacitor to 1
volt with a charging source (power supply). Calculate
the amount of charge on the capacitor by multiplying
the voltage by the value of the capacitor. Example: 1V
x 10 nF = 10 nC of charge.
7.2.1.4 Disconnect the charging source from the
capacitor.
7.2.1.5 Connect the coulombmeter input probe to the
capacitor and discharge the capacitor into the
coulombmeter. The coulombmeter should indicate the
calculated value.
7.2.2 Performance Verification of Fieldmeters and
Voltmeters — Refer to Figure 2.
7.2.2.1 Choosing Test Voltage(s) Choose one or
more test voltage(s) from Table 1, based upon the
electrostatic field level of concern:
Table 1 Test Voltages
Field of Concern Test Voltage
Under 4,000 volts/meter or
100 volts/2.5 cm
100 volts
Under 40,000 volts/meter or
1000 volts/2.5 cm
1,000 volts
Over 200,000 volts/meter or
5,000 volts/2.5 cm (See NOTE 1.)
5,000 volts
NOTE 1: If fieldmeter or voltmeter performance verification is
needed above 5,000 volts, it is left to the user to select values using
the table as guide.
7.2.2.2 Instrument Performance Verification
Charge a conductive test plate to the desired
verification voltage. Use of a suitable power supply or a
charged plate monitor for test purposes is
recommended.
7.2.2.3 Assuring Meters and Operator Are Grounded
Assure that the fieldmeter, voltmeter and operator
are grounded. Turn on the meter and zero it as required
according to manufacturer's instructions.
7.2.2.4 Directing or Pointing the Sense Head — Direct
or point the sense head of the fieldmeter or voltmeter at
the center and parallel to the surface of the plate at a
distance at least twice the manufacturer’s recommended
measurement. Slowly move the sense head toward the
center of the charged plate until a reading equal to the
voltage applied to the plate in Section 7.2.2.1 above is
displayed by the meter. Measure and record the
distance from the sense head to the surface to the plate.
Using the plate voltage from Section 7.2.2.1 above and
the recorded distance, compute the field strength for the
fieldmeter. See Figure 2, Fieldmeter and Voltmeter
Verification Check.
7.2.2.5 Alternative to Section 7.2.2.4 — Take
measurements at a specified/fixed distance per
manufacturer's instructions. Locate the sense head of
the fieldmeter or voltmeter as in Section 7.2.2.4, but, at
specified distance; reading displayed (on meter) should
be within 5% of applied voltage to plate.
NOTE 1: Section 7.2.2.4 or 7.2.2.5 should be applicable to
most meters. However, in every case, the electrostatic
fieldmeter or voltmeter manufacturer's instructions should be
read, understood, and followed.
7.2.2.6 Other Desired Test Voltages — Repeat
Sections 7.2.2.4 and 7.2.2.5 for any other desired test
voltages.
7.2.3 Performance Verification of an Electrometer
It is good practice to occasionally check the
performance of the electrometer by connecting it to a
known voltage source, and comparing its readings with
readings taken by another reference voltmeter.
7.2.4 Meter Stability All measure ment devices
should be turned on and pre-conditioned for as long a
warm-up period as recommended by the manufacturer
7.2.4.1 Reset (zero) the coulombmeter prior to each
measurement.
7.2.4.2 Check the zero on the fieldmeter or voltmeter
as specified by the manufacturer. Usually this is done
while the probe is positioned to view a grounded
surface. If the zero of the meter has drifted by more
than 5% of the test voltage for any range contained in
Table 1, the meter is not suitable for use for
measurements over that range. It may be suitable for
use over other ranges contained in Table 1, using other
test voltages. Reverify the meter’s calibration at the
selected test voltage.
7.2.4.3 Zeroing an Electrometer — Except on some
older analog models, there are usually no provisions to
zero an electrometer. Some electrometers with analog
or digital read-outs do allow offsetting of a reading, as
well as relative (delta) measurements. However, the
electronic zero of the electrometer is usually set by the
manufacturer, and should be part of the normal
calibration. It is good practice to occasionally check the
zero by shorting the input terminals together and
verifying that the zero reading is within the
manufacturer’s specifications.
7.2.4.4 See Related Information 1 for notes on
equipment accuracy and limitations.
8 Sampling
8.1 Sampling methods for this guide should be
determined by the requirements of the user's appli-
cation. Electrostatic surveys can be repeated at different
times to make them more representative of actual static
charge conditions in the surveyed area. The results will
vary due to environment (e.g., humidity) and
SEMI E43-0301 © SEMI 1995, 2001 4
workstation setup/conditions. However, any measure-
ment that is in excess of a (user) defined maximum or
that is a benchmark value, should be repeated more than
once, after performing a zero check of the measuring
equipment. This is to validate previous reading(s)
and/or establish a range/bounds in the case of varying-
moving fields on previous reading(s).
9 Test Methods & Measurem ents
9.1 Coulombmeter Measurements
9.1.1 Verifying the Coulombmeter Verify the
performance of the coulombmeter as in Section 7
above. Check/reset the zero before each measurement
and/or per manufacturer's instructions. Assure that the
coulombmeter and operator are grounded.
9.1.2 Equipment Selection Use a coulombmeter for
direct measurement of charge. A feedback-type
coulombmeter is recommended for charge
measurements for the most complete transfer of charge.
Shunt-type coulombmeters do not completely transfer
charge and are not as straightforward to use as
feedback-type coulombmeters. When using a Faraday
enclosure, the Faraday enclosure must be large enough
to hold the objects to be measured. The Faraday
enclosure is used to measure charge on insulating
materials as well as on conductors.
9.1.3 Measurements Best results are achieved when
all surfaces surrounding the measurement area are
grounded (to minimize the effects of stray fields on the
measurement) and when a consistent, systematic
handling method is used during the measurement
process. The operator should be grounded using a
grounded wrist strap.
9.1.3.1 Isolated Conductors — To measure the charge
on an isolated conductor, touch the lead from the
coulombmeter to the isolated conductor.
9.1.3.2 Faraday Enclosure Measurements — Refer to
Figure 3. To measure the charge on an object, carefully
pick up the object with an insulated tool and place the
charged object into the Faraday enclosure. Special
handling considerations: Be careful not to add or
subtract any charge in the process of moving the
charged object into the Faraday enclosure. Don’t let the
charged object rub or slide against any other surface, as
this may add or subtract charge from the object.
9.1.4 Limitations Do not attempt to measure
charges of magnitudes that are below the drift rate of
the coulombmeter.
9.2 Electrostatic Fieldmeter Measurements
9.2.1 Verifying the Fieldmeter Verify the
performance of the fieldmeter as in Section 7 above.
Check/reset the zero periodically and/or per
manufacturer's instructions. Assure that the fieldmeter
and operator are grounded.
9.2.2 Measurements Measurements made to this
guide should be taken/reported in units that conform to
the customer specifications. Most common fieldmeters
manufactured to date have operating instructions that
reflect the user doing calibration and taking
measurements in English units of volts/inch or volts at a
fixed distance in inch(es) and in these cases, raw data
are reported/listed directly. The international
community specifies that units shall be in SI (Standard
International) Metric units and the SI conversion factor
in Section 7.1.1.1 will apply. However, by definition,
electric field is expressed in volts per meter, and thus
would be expressed according to Table 2.
9.2.2.1 For instance, when using a meter calibrated
only at 100 volts, measurements under 4,000 volts/m
would be expressed to the nearest 400 volts/m.
Measurements over 4,000 volts/m would be expressed
as > 4,000 volts/m. For a meter calibrated to all three
voltages, measurements under 4,000 volts/m would be
expressed to the nearest 400 volts/m, measurements
between 4,000 and 40,000 volts/m would be expressed
to the nearest 4,000 volts/m, and measurements over
40,000 volts/m would be expressed to the nearest
40,000 volts/m.
Table 2 Measurement Units
Test Voltage For Readings of Express in Multiples of For Readings of Express as
100 V < 4,000 V/m 400 V/m > 4,000 V/m > 4,000 V/m
1,000 V < 40,000 V/m 4,000 V/m > 40,000 V/m > 40,000 V/m
5,000 V < 200,000 V/m 20,000 V/m > 200,000 V/m multiples of 200,000 V/m
Note: Measurements above 1000 volts/2.54 cm may be made based on verification of the meter at 1000 volts where less precision is acceptable
due to safety concerns with verification equipment/setup or availability of such equipment.