SEMI G75.1-069 8 © SEMI 1998 1 SEMI G75.1-0698 TEST ME THOD FOR MEASUREME NT OF IONIC IM PURITIES I N LEADFRAME TAPE 1 Summary of Method 1.1 Ion ic impurities are extracted f r o m sample tape by water at 100°C or 121°C …

SEMI G75-0698 © SEMI 19983
NOTICE: These standards do not purport to address
safety issues, if any, associated with their use. It is the
responsibility of the user of these standards to establish
appropriate safety and health practices and determine
the applicability of regulatory limitations prior to use.
SEMI makes no warranties or representations as to the
suitability of the standards set forth herein for any
particular application. The determination of the
suitability of the standard is solely the responsibility of
the user. Users are cautioned to refer to manufacturer’s
instructions, product labels, product data sheets, and
other relevant literature respecting any materials
mentioned herein. These standards are subject to
change without notice.
The user’ attention is called to the possibility that
compliance with this standard may require use of
copyrighted material or of an invention covered by
patent rights. By publication of this standard, SEMI
takes no position respecting the validity of any patent
rights or copyrights asserted in connection with any
item mentioned in this standard. Users of this standard
are expressly advised that determination of any such
patent rights or copyrights, and the risk of infringement
of such rights, are entirely their own responsibility.
Copyright by SEMI® (Semiconductor Equipment and Materials
International), 3081 Zanker Road, San Jose, CA 95134. Reproduction o
f
the contents in whole or in part is forbidden without express written
consent of SEMI.

SEMI G75.1-0698 © SEMI 19981
SEMI G75.1-0698
TEST METHOD FOR MEASUREMENT OF IONIC IMPURITIES IN
LEADFRAME TAPE
1 Summary of Method
1.1 Ionic impurities are extracted from sample tape by
water at 100°C or 121°C and quantitatively analyzed by
ion chromatography and flame photometry or atomic
absorption spectrometry. The pH and electrical
conductivity of the extracted solutions are also
measured.
1.2 If sample tape needs the cure before the extraction,
vender should inform the customer of the necessity of
curing and report the impurity data after cure.
1.3 Ion chromatography is used to determine the
presence and concentration of the following ionic
species: Na
+
, NH4
+
, K
+
, NO
3
-
, Cl
-
, Br
-
, SO
4
2-
, PO
4
3-
.
1.4 Flame photometry or atomic absorption
spectrometry is used to determine the concentration of
Na
+
and K
+
ions.
1.5 pH and conductivity measurements are also de-
scribed.
2 Referenced Documents
NOTE: As listed or revised, all documents cited shall be the
latest publications of adopted standards.
2.1 SEMI Standards
SEMI G29 — Test Method for Trace Contaminants in
Molding Compounds
SEMI G59 — Test Method for Measurement of Ionic
Contamination on Leadframe Interleafing and the
Contamination Transferred from the Interleafing to the
Leadframes
2.2 ASTM Standards
1
ASTM D 1125 — Standard Test Methods for Electrical
Conductivity and Resistivity of Water
ASTM D 1193 — Standard Specification for Reagent
Water
ASTM D 4327 — Standard Test Method for Anions in
Water by Ion Chromatography
ASTM E 70 — Standard Test Method for pH of
Aqueous Solutions with the Glass Electrode
2.3 JIS Specifications
2
1 American Society for Testing Materials, 100 Barr Harbor Drive,
West Conshohoken, PA 19428-2959
JIS K 0121 — General Rules for Atomic Absorption
Spectrochemical Analyses
JIS Z 8802 — Method for Determination of pH of
Aqueous Solutions
3 Interferences
3.1 All the sample preparations and measurements
must be carried out in clean containers which have been
washed in deionized water in order to reduce any
spurious readings.
3.2 Electrodes and syringes must be similarly cleaned
before each test.
3.3 Contamination from the sampling process may
also affect the results.
4 Equipment
4.1 Ion Extraction Vessels-Parr bomb with Teflon
liner.
4.2 Ion chromatograph for anion and cation analysis
(see SEMI G59).
4.3 Flame photometer or atomic absorption
spectrometer.
4.4 Standard solution preparation vessels, balance and
volumetric dispensers (see SEMI G29 and SEMI G59).
4.5 pH meter.
4.6 Microconductivity cell and meter.
5 Reagents and Materials
5.1 Deionized water, conductivity of less than 1.0 µ
S/cm at 25°C per ASTM D 1193.
5.2 Eluents and regenerants for specific
chromatograph columns prepared per chromatograph
equipment manufacturer’s recommendations so that the
water peak can be separated from the ionic peaks.
5.3 Compounds for Standard Solutions
5.3.1 Cations — NaCl, NH
4
Cl, KCl
5.3.2 Anions — NaCl, Na2HPO
4
•12H2O, NaBr,
NaNO
3
, K
2
SO
4
NOTE: All compounds to be reagent grade.
2 Japanese Standards Association, 1-24, Akasaka 4 Chome, Minato-
ku, Tokyo, Japan

SEMI G75.1-0698 © SEMI 1998 2
5.4 pH standard solutions.
6 Sampling
6.1 Personnel handling the tape must wear double-
layered cotton gloves and polyethylene outer gloves in
order to avoid additional sample contamination.
6.2 The sampling is one sample per one lot. The
vendor must report the definition of lot, if customer
requires it.
7 Preparation of Standard Solution and
Equipment Calibration
7.1 Preparation of Standard Solutions and
Chromatograph Calibration
7.1.1 Refer to SEMI G59, Section 9 and ASTM D
4327.
7.2 Preparation of Standard Solutions and Flame
Photometer or Atomic Absorption Spectrometer
Calibration
7.2.1 Refer to SEMI G29 and JIS K 0121.
7.3 Preparation of Standard Solutions and pH Meter
Calibration
7.3.1 Refer to ASTM E 70 or JIS Z 8802.
7.3.2 Using the standard pH solutions, calibrate the pH
meter according to the manufacturer's instructions.
7.4 Preparation of Standard Solutions and
Conductivity Cell Calibration
7.4.1 Refer to ASTM D 1125.
8 Measurement
8.1 Procedure for Impurity Extrac tion
8.1.1 Remove the protective film from the sample of
tape. Cure the tape per tape manufacturer's
recommendation, if necessary.
8.1.2 Cut the sample of tape so that it becomes smaller
than 2 cm × 2 cm.
8.1.3 Place 5.0 ± 0.1 grams of the tape sample into the
Parr bomb, and add 50 ± 0.5 grams of deionized water.
8.1.3.1 In a second bomb, add deionized water only so
that a blank solution may be obtained for measurement
of the ionic contributed by the bomb.
NOTE: It may be necessary to perform the blank test on both
bombs in order to leach out excessive extractable ionic
species.
8.1.4 Seal the caps and heat the bombs in an oven set
at 100°C ± 2°C, for 20 ± 0.25 hours.
NOTE: The oven temperature may be set at 121°C ± 2°C for
20 hours, if specified.
8.1.5 Allow the bombs to reach room temperature.
8.1.6 Open the bombs just before the measurements
are to be made. Measurements must be made within
eight (8) hours of the completion of the sample
preparation.
8.2 Quantitative Analysis by Ion Chromatography
8.2.1 Prepare the chromatograph for operation by
regenerating the columns according to the
manufacturer’s recommendations.
8.2.2 Run the eluent through the chromatograph until a
stable baseline calibration is established.
8.2.3 Inject the recommended sample size of test
solution into the chromatograph and record the ion
chromatogram.
8.2.4 Repeat for all the samples and also run the
background sample.
8.2.5 Determine the ionic type and concentration from
the calibration curves.
8.2.6 Results
8.2.6.1 The concentration of ionic species in the
leadframe tape is given by:
A=
(B − C) × 50
5
where :
A is the ionic content of the leadframe
tape (ppm)
B is the ionic content on the extracted
solution, determined from the calibration
curves (ppm)
C is the ionic concentration of the back-
ground sample (ppm)
8.3 Quantitative Analysis by Flame Photometer or
Atomic Absorption Spectrometer
8.3.1 Set a specific operational condition for an
apparatus.
8.3.2 Introduce the recommended sample size of test
solution into a flame.
8.3.3 Repeat for all the samples and also introduce the
background sample.
8.3.4 Determine the concentration from calibration
curves.
8.3.5 Results