semi合集-English.pdf - 第4704页

SEMI C3.58-0303 © SEMI 2000, 2003 4 6.3.3 Operation Check — The instrument sho uld be checked p eriodically for correct operation. The instr ument should be zero ed by flowing refere nce gas, whic h is dry oc tafluorocyc…

100%1 / 7923
SEMI C3.58-0303 © SEMI 2000, 2003 3
5.1.4.6.1 Calculate the sample’s free acidity using the
following equation:
ppb HCl = (C2-C1) × F
W
Where: ppb HCl = free acidity
C1 = blank conductivity, µmho/cm
C2 = sample conductivity, µmho/cm
F = calibration factor, ng HCl/µmho
W = sample weight used in analysis,
grams
6 Analytical Procedures - Instrumental
Analysis
6.1 Air (N
2
,O
2
, CO, CO
2
) and OxygenThis
procedure is for the determination of air components in
octafluorocyclobutane using a gas chromatograph with
a discharge ionization detector. Column switching is
used to separate and elute the components and to
backflush octafluorocyclobutane. One method/column
is used for N
2
, O
2
, and CO and a second
method/column is used for CO
2
. The sample should be
taken from the vapor in the container directly into the
instrument sampling system. The container being
analyzed should have equilibrated at approximately
30°C before analysis is begun.
6.1.1 Detection Limits — Nitrogen, 10 ppb vol;
Oxygen, 50 ppb vol; Carbon monoxide, 10 ppb vol;
Carbon dioxide, 10 ppb vol.
6.1.2 Instrument Parameters
6.1.2.1 Columns: Precolumn: Haysep DB on
Carbopack B, 80/100 mesh, 2 ft × 1/8 in stainless steel.
N
2
, O
2
, CO column: Molecular Sieve 5a, 80/100 mesh,
8 ft × 1/8 in stainless steel. CO
2
column: Haysep DB
on Carbopack B, 80/100 mesh, 8 ft × 1/8 in stainless
steel.
6.1.2.2 Gas Flows: Helium carrier gas 40 mL/min for
each column.
6.1.2.3 Sample volume: 1.0 mL
6.1.2.4 Temperatures: Detector setpoint = 25°C
(actual is higher ~60°C due to proximity to the oven),
Oven 77°C isothermal.
6.1.2.5 Calibration Standard: 1-30 ppm vol oxygen,
nitrogen, carbon monoxide, and carbon dioxide in
helium.
6.1.2.6 Chromatograms — See Figures 1,2.
6.2 Organic Impurities — This procedure is for the
determination of organic impurities in
octafluorocyclobutane using a gas chromatograph with
a packed column and flame ionization detector.
6.2.1 Detection Limits
PFC-1216 (hexafluoropropene),
PFC-31-10mc (n-decafluorobutane),
PFC-1318my (octafluorobutene, cis and trans)
0.1 vol, ppm
PFC-31-10my (iso-decafluorobutane),
CFC-114 (1,2-dichloro-1,1,2,2-tetrafluoroethane),
CFC-114a(1,1-dichloro-1,2,2,2-tetrafluoroethane),
THF(tetrahydrofuran)
0.2 vol, ppm
6.2.2 Instrument Parameters
6.2.2.1 Column — 1% SP-1000, on Carbopack B,
60/80 mesh, 24 ft × 1/8 in stainless steel or equivalent.
6.2.2.2 Gas Flows — Helium carrier gas, 20
mL/minute.
6.2.2.3 Sample volume — 1.0 mL vapor.
6.2.2.4 Temperatures — Injector = 150°C, Detector =
250°C, Oven: 75°C for 10 minute, increase 8°C/min to
200°C and hold for 10 minutes.
6.2.2.5 Calibration Standard: 1-50 ppm vol, PFC-
1216, PFC-31-10my, PFC-31-10mc, PFC-1318my (cis
and trans), CFC-114/114a, and THF in
octafluorocyclobutane.
6.2.2.6 Chromatogram — See Figure 3.
6.3 Water — This procedure is for the determination of
trace moisture (water) in Octafluorocyclobutane, using
oscillating crystal technology. The instrument monitors
the change in vibrational frequency of a
hygroscopically sensitized quartz crystal when it is
exposed alternately to wet and dry sample gas. The
change in vibrational frequency is a function of the
amount of moisture sorbed from the wet sample gas.
The sample should be taken from the vapor in the
container directly into the instrument sampling system.
The container being analyzed should have equilibrated
at approximately 30°C before analysis is begun.
6.3.1 Detection Limit — 0.04 ppm vol (40 ppb vol).
6.3.2 Sample Pressure and Flow — These parameters
should be set in accordance with instrument
manufacturer’s instructions.
SEMI C3.58-0303 © SEMI 2000, 2003 4
6.3.3 Operation Check — The instrument should be checked periodically for correct operation. The instrument
should be zeroed by flowing reference gas, which is dry octafluorocyclobutane (dried by use of an ultra dryer system
provided by the instrument manufacturer). Then the instrument should be calibrated by use of the internal moisture
generator (permeation tube) which has been certified by the manufacturer at some known moisture level. When not
in use the instrument should be purged with dry gas (i.e., helium or nitrogen).
6.3.4 Operation Procedures
6.3.4.1 Initiate flow of the reference gas to the instrument and allow 30 minutes for stabilization.
Figure 1
Nitrogen, Oxygen, and Carbon Monoxide Determination in Octafluorocyclobutane
Figure 2
Carbon Dioxide Determination in Octafluorocyclobutane
SEMI C3.58-0303 © SEMI 2000, 2003 5
Figure 3
Determination of Organic Impurities in Octafluorocyclobutane
6.3.4.2 Obtain a continuous flow sample of the Octafluorocyclobutane source. The sample system by which the
octafluorocyclobutane is passed to the instrument should consist of electroplated stainless steel tubing, zero dead-
volume fittings, and be heated and purged with dry gas while not in use.
6.3.4.3 Allow the sample gas to flow through the sampling system and instrument until a stable reading is obtained.
NOTICE: SEMI makes no warranties or representations as to the suitability of the standard set forth herein for any
particular application. The determination of the suitability of the standard is solely the responsibility of the user.
Users are cautioned to refer to manufacturer’s instructions, product labels, product data sheets, and other relevant
literature respecting any materials or equipment mentioned herein. These standards are subject to change without
notice.
By publication of this standard, SEMI takes no position respecting the validity of any patent rights or copyrights
asserted in connection with any item mentioned in this standard. Users of this standard are expressly advised that
determination of any such patent rights or copyrights, and the risk of infringement of such rights, are entirely their
own responsibility.
Copyright by SEMI® (Semiconductor Equipment and Materials
International), 3081 Zanker Road, San Jose, CA 95134. Reproduction o
the contents in whole or in part is forbidden without express written
consent of SEMI.