semi合集-English.pdf - 第535页

SEMI E77-1104 © SEMI 1998, 2004 3 12 Preparation of Apparatu s 12.1 Locate the D.U.T. in the test environment to stabilize temperature for 24 ho urs prior to warm up. 12.2 The reference operating conditions shall b e as …

100%1 / 7923
SEMI E77-1104 © SEMI 1998, 2004 2
5.2.8 nameplate gas — the gas intended to be
controlled by the MFC in operation.
5.2.9 range — the algebraic difference between the
maximum and minimum values.
5.2.10 setpoint — the input signal provided to achieve
a desired flow, reported as sccm, slm, or percent full
scale.
5.2.11 span — the full scale range of the D.U.T.
5.2.12 surrogate gas — the gas substituted for the
nameplate gas during the calibration process.
5.2.13 zero drift — the undesired change in electrical
output (i.e., indicated flow), at a no-flow condition,
over a specified time period, reported in sccm or slm.
5.2.14 zero offset — the deviation from zero at a “no-
flow” condition reported in sccm, slm, or mV.
6 Summary of Test Method
6.1 Gas flow and setpoint data are collected for two
gases. This data is reduced to quantify the relationship
between the flow measurement by the MFC on one gas
to another gas.
6.2 This method allows the user to determine the
conversion factor between two gases for the MFC and
to determine the onset of errors in the MFC calibration
due to the conversion factor/function effects.
7 Interferences
7.1 The accuracy rating of the measuring equipment
shall be superior to that of the D.U.T. Preferably the
measuring equipment will have an accuracy that is four
times better than the D.U.T. Calibration equipment
must have a valid calibration certificate.
7.2 Take care when using test instruments with a
specified accuracy expressed in percent of full scale.
7.3 Installation effects on the flow should be
minimized. Monitor pressure upstream of the D.U.T.
to ensure that flow variations due to pressure are
minimized.
7.4 Verify electrical signals directly at the D.U.T.
connector to ensure that the signals at the D.U.T. and
standard agree with the signals at the data recording
equipment.
7.5 Certain gases will contaminate the D.U.T. This
test should be considered a destructive test in such
cases.
7.6 All electrical measurements should be read on
devices with at least 4.5 digits of resolution. These
devices must have valid calibration certifications.
7.7 The device mounting position must be in
accordance with the manufacturer’s specifications.
8 Apparatus (See Figure 1.)
8.1 Flow Standard — A device or system that
accurately measures the flow and reports the actual
flow.
8.2 Data Acquisition System — The system that
measures the electrical signals from the device under
test. The data acquisition system may also read the
signals from the flow standard, record test data and
control the test sequence.
8.3 Temperature Probe — A device to measure the
temperature of the flowing gas.
8.4 Three-Way Valve — A special valve to switch the
system from one gas source to another.
8.5 Manual Isolation Valves — Valves that will
positively shut off the gas line.
8.6 Pressure RegulatorA device that regulates gas
pressure to a set value.
8.7 Pressure TransducerAn instrument to measure
the gas pressure and report it as an electrical signal.
9 Materials
9.1 Clean, Dry N
2
— With 99.999% minimum purity,
to be used for purging.
9.2 Test Gas “A”
9.3 Test Gas “B”
10 Safety Precautions
10.1 This test method may involve hazardous
materials, operations, and equipment. This test method
does not purport to address the safety considerations
associated with its use. It is the responsibility of the
user to establish appropriate safety and health practices
and to determine the applicability of regulatory
limitations before using this method.
10.2 Follow the manufacturer’s specifications and
instructions for installation and operation whenever
possible. Note any exceptions in the test report.
11 Test Specimen
11.1 Allow all components in the test apparatus to
warm up following the manufacturer’s specification.
11.2 Take necessary steps when switching gases to
ensure that only the desired gas is in the D.U.T. and
flow standard at the time the test is performed.
SEMI E77-1104 © SEMI 1998, 2004 3
12 Preparation of Apparatus
12.1 Locate the D.U.T. in the test environment to
stabilize temperature for 24 hours prior to warm up.
12.2 The reference operating conditions shall be as
follows:
12.2.1 Ambient Temperature — 23 ± 2°C
12.2.2 Gas Temperature — Same as ambient. In the
case of a condensible gas, the gas temperature should
be maintained as a gradient, with highest temperature at
the outlet end of the test set up, and slightly reduced
temperatures back to the source. This will ensure that
condensing vapors do not accumulate in the test set up.
Gradient should simulate actual process parameters
where the MFC will be installed.
12.2.3 Ambient Pressure — 101.3 kPa (+ 4.7 or – 15.3
kPa)
12.2.4 Gas Pressure, Inlet — 172 ± 34 kPa unless the
gas is not capable of delivering this pressure, then
normal fab operating conditions should be observed.
12.2.5 Gas Pressure, Outlet — < 80 kPa
12.2.6 Relative Humidity — 40% ± 5%, non-
condensing (suggestion: record if outside this range)
12.2.7 Magnetic Field 50 µT
12.2.8 Electromagnetic Field 100 µV/m
12.2.9 Vibration 0.5 m/s at 50 to 200 Hz
12.3 Following the conditioning period (See Section
12.1), warm up the device according to manufacturer’s
specifications.
12.4 Perform an adequate nitrogen purge to ensure all
previous gases and moisture have been removed from
the system. Prior to corrosive gas testing, a cyclic
pump and purge operation is recommended, alternately
backfilling with nitrogen, and evacuating the test
manifold.
12.5 Leak check the manifold, using available
methodologies to verify the test system leak integrity.
Introduce the test gas at a sufficient rate and time to
ensure the test apparatus is completely filled with the
test gas and only the test gas.
12.6 Record the zero offset with line pressure inside
the MFC to best simulate normal operating conditions.
Line pressure during testing is to be 170.30 kPa (24.7
psia) unless safety practices for the gas under test
dictate that a lower line pressure be used.
NOTE 1: In addition, if this test is performed on hazardous
gases, bleeding off gas pressure to obtain atmospheric
pressure inside the MFC may not be easy to do.
12.7 Command 100%, and establish flow, then close
the downstream isolation valve; then close the upstream
isolation valve (see Figure 1).
12.8 With both isolation valves closed, and a 100%
setpoint wait until the pressure drop across the MFC is
dissipated, ensuring a “no flow” condition through the
MFC. Dissipation of the pressure across the MFC is
indicated when the indicated flow drops to a steady
state value near zero.
12.9 After the electrical output signal has stabilized for
at least three minutes, record the MFC zero offset in
Table 1.
Figure 1
Mass Flow Controller Test Fixture
SEMI E77-1104 © SEMI 1998, 2004 4
Table 1 Indicated and Actual Flow vs. Setpoint
MFC Mfg/Model/Serial # ____________________________________
Name Plate Gas/Range ____________________________________
Test Gas ____________________________________
Temp ___________ Bar ___________ Date ___________
Factory Calibration Gas ____________________________________
Column A
D.U.T. Setpoint
Indication in % F.S.
Column B
D.U.T. Flow
Indication in % F.S.
Column C
Flow Standard
Raw Data
0%
#1
0
50
60
70
80
90
100
90
80
70
60
50
40
30
10
0% __________________ 0
10
20
30
40 0
50
60
70
80
90
100
90
80
70
60
50
40
30
10
0% __________________ 0
10
20
30
40