semi合集-English.pdf - 第4202页
SEMI F73-1102 © SEMI 2002 4 collection effici ency. Any adhesives used to mount samples must not contam inate the surface to be analyzed. To remove loosely adhe red surface particles, the sample may be blown off immedia …

SEMI F73-1102 © SEMI 2002 3
Kβ peak with Fe Kα peak interferes with determination
of low levels of Manganese in Iron-Chromium matrix.
8 Apparatus
8.1 Instrumentation — Any SEM instrument used for
this method must be capable of a minimum point-to-
point resolution of 30 nm, as measured with a NIST
4
traceable standard. Any commercially available SEM
with image recording capabilities of at least 100 cm
2
image and an EDS analyzer capable of 170 eV or less
FWHM resolution for Mn Kα may be employed.
8.2 The SEM shall have a sample stage capable of
aligning the sample to provide multiple areas of view
with the orientation of the electron beam approximately
normal to the surface, and to optimize secondary
electron and X-ray signal collection efficiencies.
8.3 Grid Overlay — A transparent grid overlay of grid
size equal to 1.814 micrometers multiplied by the pre-
selected magnification in the range 3000 to 3600 (grid
size would be 6.35 mm [= ¼ inch] square for 3500×
magnification). The grid lines shall be as fine as
possible but clearly visible. The grid overlay shall be
designed to overlay the image of a photomicrograph
with 252 grids (18 by 14 grids for the 3.5 by 4.5 inch
image of a standard Polaroid photomicrograph) with
minimal interference from alphanumeric information
recorded with the photomicrograph, but without
reference to any features in the photomicrograph. The
rows and columns of the grid overlay may be indexed
with alphanumerics to identify and locate specific
features in the photomicrographs (see examples in
Appendices 1 and 2). It is suggested that a fixed
reference point for the edges of the photomicrograph be
established to be consistently used in all overlays.
9 Reagents and Materials
9.1 Sample Preparation Materials — Equipment
required to section the test specimen without damaging
or contaminating the surfaces to be analyzed. A clean,
dry hacksaw or dry bandsaw is recommended, using a
slow cutting speed to avoid excessive sample heating.
9.2 Sample Mounting Materials — Sample mounting
holders specific to the SEM instrument used.
Conductive paste or tape used to adhere the sample to
the holder must be vacuum stable and applied so that
the area of analytical interest is not contaminated.
10 Safety Precautions
4 National Institute of Standards and Technology, 100 Bureau Drive,
Stop 3460,Gaithersburg, MD 20899, (301) 975-NIST (6478) TTY (301)
975-8295, website: www.nist.gov.
10.1 This test method does not purport to address the
safety considerations associated with use of high
voltage, vacuum, electron and X-ray producing
equipment.
10.2 The method assumes a SEM and EDS analyst
with adequate skill level as well as knowledge of
instrumentation and associated safety precautions.
11 Test Specimen
11.1 Specimens are to be sectioned to appropriate size
for the particular SEM instrument. Any sample
preparation technique used shall avoid introducing
contamination onto the surface to be measured. In
addition, preparation must avoid heating of the sample
in excess of approximately 100°C to avoid oxide
growth or change in surface composition. Slow speed
dry cutting is recommended.
11.2 After preparation, samples shall be protected from
contamination by means such as sealing in
noncontaminating bags or wrapping in clean aluminum
foil. Appropriate cleaning of the analysis surface to
remove sample preparation residues is allowed but not
required. Any such cleaning procedures should be
stated in the report. After sectioning, samples should
be analyzed promptly.
11.3 The samples shall be analyzed in the uncoated
condition; ie, without sputtering with a conductive
coating.
12 Preparation of Apparatus
12.1 The SEM shall be in good condition to assure
proper performance in accordance with the
manufacturer’s specifications.
13 Calibration and Standardization
13.1 Instrument calibration for magnification and EDS
performance shall be performed in accordance with
ASTM E 766 and ANSI-IEEE 759 and/or the
instrument manufacturer’s recommendations. The
magnification should be calibrated at the working
distance selected. Calibration frequency shall be per
instrument manufacturer’s recommendations.
14 Procedure
14.1 The sample is to be mounted in accordance with
manufacturer' s recommendations and in a manner
consistent with high vacuum analytical procedures. The
sample shall be oriented such that the areas of interest
for SEM imaging can be viewed at 70 to 90° incidence
angle of the electron beam to the surface with optimum
secondary electron collection efficiency, and the ability
to re-orient the sample for optimization of X-ray

SEMI F73-1102 © SEMI 2002 4
collection efficiency. Any adhesives used to mount
samples must not contaminate the surface to be
analyzed. To remove loosely adhered surface particles,
the sample may be blown off immediately before
introduction into the SEM with clean, dry, 0.1
micrometer filtered gas.
14.2 Place the sample in the SEM chamber for
pumpdown. Activate the electron beam when vacuum
conditions meet manufacturer’s recommendations. The
instrument accelerating voltage should be a consistently
selected value within the range 15 to 25 Kev; 20 KeV is
suggested. The working distance should be within the
range recommended by the instrument manufacturer.
Sample position (tilt angle and orientation) may be
adjusted to optimize EDS detector collection efficiency
for EDS analysis.
14.3 Increase the magnification to ≥ 400× for initial
focus; adjust instrument parameters for astigmatism and
other anomalies. Decrease magnification to 200× and
move the sample while viewing until an area judged as
representative of the whole is in view. The area to be
analyzed should be as free of particles and defect
features produced by sample preparation as possible.
14.4 Refocus and record images of this area at 200×,
1000×, and a pre-selected, consistently used
magnification within the range 3000 to 3600×. Move
to another representative area and record an image at
the same 3000–3600× magnification. Move to a third
representative area and record another image at the
3000–3600× magnification.
14.5 Each area selected for recording should include a
defect or metallurgical feature to demonstrate proper
focusing, contrast adjustment and resolution capability.
Contrast adjustment should be performed per the
technique described in Section 4.7.2.1 of Scanning
Electron Microscopy and X-Ray Microanalysis.
5
All
defects and/or features for which the change in signal
∆S due to the contrast exceeds the noise N by a factor
of 5 minimum must be clearly distinguishable in the
image recorded (see Section 4.5 of reference 5).
14.6 If any inclusions or contamination are noted in the
3000–3600× images, EDS analysis of representative
defect(s) may be performed if requested by the
company for which the test method is performed.
Collect X-ray signals for a sufficient length of time to
obtain an integrated count of ≥ 250,000 within the
range of 0 to 10 keV.
5 Scanning Electron Microscopy and X-Ray Microanalysis: A Text
for Boilogists, Materials Scientists, and Geologists; Joseph I.
Goldstein, Dale E. Newbury, Patrick Echlin, David C. Joy, A. D.
Romig, Jr., Charles E. Lyman, Charles Fiori, Eric Lifshin; Second
Edition, Plenum Press, New York (1992).
15 Calculations or Interpretation of Results
15.1 The 3000–3600× recorded images shall be placed
under the transparent overlay grid. The number of grid
squares that contain all or part of a surface defect shall
be counted and summed for each photomicrograph.
Each such grid square shall only be counted once; it is
not the intent of this Test Method to count individual
defects. If a defect appears in more than one adjacent
grid square, each grid square occupied shall be counted.
NOTE 2: Examples of scanning electron micrographs
overlaid with an alphanumeric indexed grid showing specific
types of defects with comments on their locations and
whether they should be counted may be viewed at
http://www.semi.org/web/wstandards.nsf/url/SurfaceAnalysis
Examples.
15.2 Surface film residues resulting from improper
cleaning are generally diffuse and difficult to quantify.
These residues, if present, are counted as one per
image. Particles that loosely adhere to the surface must
be presumed to be artifacts of sample preparation or
exposure, and therefore shall be ignored.
15.3 Peaks appearing in the EDS spectra shall be
identified and the spectra labeled to indicate whether it
is a representative area, or identified as a defect from a
photomicrograph.
16 Reporting Results
16.1 A tabular summary of defects counted per 3000–
3600× photomicrograph shall be presented.
Additionally, the table should indicate an average of
defect counts and the maximum from the three 3000–
3600× images. An example of a table reporting defect
counting results and the photomicrographs from which
the defect counts were taken are shown in Appendix 1.
16.2 All photomicrographs are to be permanently
labeled with sample identification, magnification, a
magnification scale bar, date, and analyst identification.
The photomicrographs should be available for
inspection upon request.
16.3 EDS spectra are to be labeled with the same
information, plus peak identifications. If any
instrument parameters (e.g., tilt angle) were changed for
EDS analysis, these must be noted. If peak height
differences are such that adequate representation of all
peaks cannot be made from the same plot, two plots
having different scaling shall be included.
Identification of each defect analyzed, either by arrow
on photomicrograph or by notation of grid position,
shall be made.

SEMI F73-1102 © SEMI 2002 5
17 Related Documents
17.1 ASTM Standards
ASTM F1372-93 — Standard Test Method for Scanning Electron Microscope (SEM) Analysis of Metallic Surface
Condition for Gas Distribution System Components
ASTM F1375-92 — Standard Test Method for Energy Dispersive X-Ray Spectrometer (EDX) Analysis of Metallic
Surface Condition for Gas Distribution System Components