semi合集-English.pdf - 第7586页

SEMI MF2139-1103 © SEMI 2003 6 12.5 If a cryopanel is used, fill the liquid nitrogen or helium cold trap. 12.6 For t he calibration specim en use a cesium primary ion beam and detect the ne gative ion of 14 N 28 Si or 15…

100%1 / 7923
SEMI MF2139-1103 © SEMI 2003 5
7 Apparatus
7.1 SIMS Instrument — Equipped with a cesium
primary ion source, electron multiplier detector, and
Faraday cup detector, capable of measuring negative
secondary ions. The SIMS instrument should be
adequately prepared (that is, baked) so as to provide the
lowest possible instrumental background, since the
instrumental background can affect the detection
capability. High quality vacuum is desired. A liquid
nitrogen- or liquid helium-cooled cryopanel, which
surrounds the test specimen holder in the analysis
chamber, may be helpful.
7.2 Test Specimen Holder
7.3 Oven — For baking the test specimen holder.
7.4 Stylus Profilometer — Or equivalent device to
measure SIMS crater depths. This device is required to
calibrate depth scale for concentration profiles of
calibration standard samples.
8 Reagents and Materials
8.1 No reagents or materials are required for this test
method.
9 Safety Precautions
9.1 The preparation of silicon test specimens from
silicon substrates requires the crystal to be cleaved or
broken. This procedure may generate very sharp
silicon shards. Care must be taken to protect the body
and particularly the eyes from these shards.
10 Sampling, Test Specimens, and Test Units
10.1 Since this procedure is destructive in nature, a
sampling procedure must be used to evaluate the
characteristics of a group of silicon wafers. No general
sampling procedure is included as part of this test
method, because the most suitable sampling plan
depends upon individual conditions. For referee
purposes, a sampling plan shall be agreed upon before
conducting the test. See ASTM Practice E 122 for
suggested choices of sampling plans.
10.2 The calibration or reference material shall be
silicon crystal with boron concentration below 2 × 10
17
atoms/cm
3
and implanted with ions of
14
N or
15
N; an
ion implant of energy 200 keV and dose of about 1 ×
10
14
atoms/cm
2
is preferred.
10.3 Sample specimens must have a chemo-mechanical
polished surface on the side used for analysis.
11 Calibration and Standardization
11.1 Calculate the RSF for an element in a matrix
experimentally from the SIMS profile of a reference
material containing a known areal density of the
impurity of interest as follows:
)( nISId
tInD
RSF
bi
m
= (7)
where:
D = a real density of the impurity, atoms/cm
2
,
n = number of data cycles in the profile,
I
m
= matrix isotope secondary ion intensity, counts/s,
d = sputtered depth, cm,
SI
i
= sum of the impurity isotope secondary ion
(counts over the depth of the profile),
I
b
= constant background intensity of the impurity
isotope, counts/cycle and
t = analysis time for the species of interest, s/cycle.
11.2 Calculate the calibration of concentration in the
SIMS profile as follows:
EM
FC
RSF
I
I
C
m
i
i
××= (8)
where:
C
i
= impurity atomic concentration, atoms/cm
3
,
I
i
= isotope secondary ion intensity, counts/s,
I
m
= matrix isotope secondary ion-intensity,
counts/s,
RSF = relative sensitivity factor, atoms/cm
3
, and
EM
FC
= ratio of matrix intensities on the Faraday
cup (FC) to the electron multiplier (EM),
when two detectors are used.
11.3 Complete the conversion of data cycles or time
into depth by measuring the crater depth and the total
time (or data cycles) of the sputtering that formed the
crater.
NOTE 1: The depth scale is assumed to be linear with time or
data cycles.
12 Procedure
12.1 Cleave or break specimens from the silicon
substrates in order to fit within the SIMS sample holder.
Specimens must be handled under normal analytical
laboratory practices, but no special ambient is required.
12.2 Load the specimens into the SIMS sample holder,
checking to see that the specimens are flat against the
backs of the windows and cover the windows as much
as is possible. A specimen load includes one
calibration specimen, and one or more test specimens.
12.3 Bake the loaded sample holder at 100 ± 10° C for
a minimum of 1 h in air.
12.4 Turn on the instrument in accordance with the
manufacturer' s instructions.
SEMI MF2139-1103 © SEMI 2003 6
12.5 If a cryopanel is used, fill the liquid nitrogen or
helium cold trap.
12.6 For the calibration specimen use a cesium primary
ion beam and detect the negative ion of
14
N
28
Si or
15
N
28
Si, depending upon the nitrogen isotope used in
the ion implantation; detect during or at the end of the
profile a matrix negative ion (
28
Si,
29
Si, or
30
Si). After
the entire sample load is analyzed, measure the SIMS
crater depth in this sample using a profilometer.
12.7 Position the specimen holder to sputter a crater in
a test specimen near the center of the window. Without
detecting secondary ion intensities, begin a SIMS
profile.
12.8 Repeat Section 12.7 for each test specimen.
12.9 If the boron level in the test specimen is below 1
× 10
17
/cm
3
, detect the negative ions of
12
C
28
Si,
14
N
28
Si,
and a matrix signal (
28
Si,
29
Si,
30
Si, or
30
Si
2
). If the
boron level is at or above 1 × 10
17
/cm
3
, include the
detection of the negative ion of
11
B
28
Si.
12.10 For the analysis of the test specimens, select two
raster conditions, one to maximize the sputter rate, and
one that provides a slower sputter rate. Choose
apertures or other means to maintain the analysis area
under raster change and to keep the negative ion count
rates on the electron multiplier detector below 1 × 10
5
counts/s for the test specimen.
12.11 Position the specimen holder to sputter a new
crater near the crater created in Section 12.7 or Section
12.8 . Do not make the second crater in the same crater
created in Section 12.7 or Section 12.8 .
12.12 Center the primary ion beam and begin a SIMS
profile with the raster area that produces the lower
sputter rate. Continue the profile until all signals
provide relatively constant intensity for at least 20 data
points.
12.13 Change the raster area to maximize the signal
intensity. Continue the profile under this raster
condition for at least 20 data points.
12.14 Reduce the raster area back to the original
condition.
12.15 If the boron content is less than 1 × 10
17
/cm
3
,
subtract 0.0341 times the measured mass 40 signal
(counts/s) from the measured mass 42 signal (counts/s)
to obtain the corrected mass 42 signal corresponding to
real nitrogen (
14
N
28
Si).
NOTE 2: If mass 40 carbon (
12
C
28
Si) is defined = 1, mass 42
(
12
C
30
Si +
13
C
29
Si) = 0.0335 + 0.0006 = 0.0341.
NOTE 3: A hydride correction should be negligible for H on
the order of 10
19
atoms/cm
3
or less.
12.16 If the boron concentration is at or greater than 1
× 10
17
/cm
3
, subtract 0.0583 times the mass 39 signal
from the carbon correction to correct for the boron
interference.
NOTE 4: The mass 39 boron (
11
B
28
Si +
10
B
29
Si) is defined =
1, and mass 40 (
11
B
29
Si +
10
B
30
Si) = 0.0583.
12.17 Calculate the bulk nitrogen concentration [N]
and the background nitrogen concentration [N
b
] for
each test specimen using Equations 5 and 6.
12.18 Record the specimen identification.
13 Report
13.1 Report the following information:
13.2 The instrument used, the operator, and the date of
the measurements,
13.3 Identification of the test and calibration
specimens,
13.4 The bulk nitrogen concentration values for the test
specimens,
13.5 The nitrogen instrumental background concentra-
tion during the faster sputter rate, and
13.6 Whether anomalous nitrogen spikes were present,
and if so, whether the averaging included or excluded
these spikes.
14 Precision and Bias
14.1 The precision of the nitrogen determination has
not yet been determined either by pilot study or by
interlaboratory test.
14.2 There is no accepted reference material for
determining the bias for the procedure in this test
method. Therefore, no statement on bias can be made.
15 Keywords
15.1 nitrogen concentration; secondary ion mass
spectrometry; silicon; SIMS
SEMI MF2139-1103 © SEMI 2003 7
NOTICE: SEMI makes no warranties or
representations as to the suitability of the standards set
forth herein for any particular application. The
determination of the suitability of the standard is solely
the responsibility of the user. Users are cautioned to
refer to manufacturer' s instructions, product labels,
product data sheets, and other relevant literature,
respecting any materials or equipment mentioned
herein. These standards are subject to change without
notice.
By publication of this standard, Semiconductor
Equipment and Materials International (SEMI) takes no
position respecting the validity of any patent rights or
copyrights asserted in connection with any items
mentioned in this standard. Users of this standard are
expressly advised that determination of any such patent
rights or copyrights, and the risk of infringement of
such rights are entirely their own responsibility.
Copyright by SEMI® (Semiconductor Equipment and Materials
International), 3081 Zanker Road, San Jose, CA 95134. Reproduction o
f
the contents in whole or in part is forbidden without express written
consent of SEMI.