semi合集-English.pdf - 第7468页

SEMI MF1617-0304 © SEMI 2003, 2004 4 Absorption Sp ectroscopy (VPD/AAS) or Vapo r Phase Decomposition/Inductiv ely-Coupled-Plasma Mass Spectrometry (VPD/ICP-MS) and shown by qualitative SIMS to be spatially uniform, sinc…

100%1 / 7923
SEMI MF1617-0304 © SEMI 2003, 2004 3
secondary ion count rate or ion count rates of known
relative intensity such as natural
28
Si/
30
Si) on each
detector.
6.7 The net integrated
23
Na,
27
Al,
39
K, and
54
Fe signals
are converted to quantitative areal densities using the
detector efficiency ratios (if multiple detectors are used)
and RSFs measured from reference materials.
34
7 Apparatus
7.1 SIMS Instrument — Equipped with a primary ion
beam, preferably O
2
+
, a mass spectrometer with some
method of discriminating molecular ion interferences
from elemental ions of interest, an electron multiplier
detector, or faraday cup detector, or similar detector
system capable of measuring secondary ion count rates,
or any combination thereof, and the ability to sputter
the surface at less than 0.015 nm/s. An molecular
oxygen jet may be used to stabilize the surface ion
yield; the local oxygen gas flux to the specimen surface
must be stable enough that the secondary ion yield does
not vary during the analysis. The stability of the
oxygen jet effect can be checked by monitoring a
silicon matrix signal during a profile; and if the matrix
signal is not monitored during a profile, the vacuum
chamber pressure can be monitored for fluctuations in
the oxygen gas pressure. The SIMS instrument should
be adequately prepared and maintained so as to provide
the lowest possible instrumental backgrounds.
7.2 Test Specimen Holder — Reserved for the SIMS
measurement. In some instruments the holder can
support multiple 5 by 5-mm samples that are held face
down against metal (tantalum) windows. In other
instruments the holder can support one or more 15 by
15-mm samples by spring clips on the edge of the
specimen or silver paste on the sample back surface.
Some instruments can hold full silicon wafers. In all
cases, the sample holder should be adequately prepared
and maintained so as to provide the lowest possible
contribution to instrumental backgrounds.
7.3 Stylus Profilometer — Or equivalent device (for
example, atomic force microscope) to measure SIMS
crater depths. This is required to calibrate depth scales
for concentration profiles of calibration standard
samples. For routine depth scale calibration of test
specimens, the instrument must be capable of crater
depth measurements of 10 nm with an accuracy and
precision of 10%. In the absence of such an instrument,
a secondary depth calibration standard (such as a silicon
wafer with a known impurity depth profile) must be
prepared and can be used to measure the SIMS sputter
rate during the measurement.
NOTE 1: The accurate measurement of sputter crater depths
on the order of and below 10 nm is still an area of research,
because the sputtered silicon is very reactive to oxygen in the
air when removed from the SIMS chamber for crater depth
measurement. For example, sputter craters expected to be
only 1 nm in depth can actually be thicker than the original
native oxide due to this rapid oxide growth on sputtered
silicon, that is, one observes a bump on the surface where the
erosion was, rather than a crater. If a reference sample is
similar to the unknown in type, that is, intentional surface
contamination rather than shallow ion implant, it may be
possible to avoid the need for a direct shallow crater depth
measurement on each sample.
8 Sampling
8.1 Since this procedure is destructive in nature, a
sampling procedure must be used to evaluate the
characteristics of a group of silicon wafers. No general
sampling procedure is included as part of this test
method, because the most suitable sampling plan will
vary considerably depending upon individual
conditions. For referee purposes, a sampling plan shall
be agreed upon before conducting the test. See ASTM
Practice E 122 for suggested choices of sampling plans.
9 Specimen Requirements
9.1 Sample specimen surfaces must be flat and smooth
(see Sections 3.5 and 3.6).
9.2 Sample specimens must be cleaved or diced to fit
within the sample specimen holder. Specimens need
not be all the same size, but typical specimen sizes are
approximately square with one side of dimension 5 to
15 mm.
9.3 Prepare a BLANK specimen with a level of surface
metal below that expected to be detected by this test
method, and which has surface organics of a level and
type expected for the unknown specimens.
NOTE 2: If measurements are to be made on specimens
taken soon after a cleaning process, the organics on the
surface from clean room air and cassette carriers are very low
and negligible. However, if the specimens are stored for one
day or more, the level of organics may be significant for a
molecular ion interference in the measurement
5
; this
interference must be removed by the detection scheme.
10 Calibration
10.1 The calibration is by RSFs which are determined
empirically from reference materials.
10.1.1 Reference materials may be either:
10.1.1.1 Spin Coat Contaminated Samples
Calibrated by Vapor Phase Decomposition/Atomic
5 Muller, A. J., Sinclair, J. D., Psota-Kelty, L. A., and Krauttler, H.
W., “Measurement of Airbourne Concentrations and Surface Arrival
Rates of Organic Contaminants in Clean Rooms,” 1993 Proceedings,
Vol 1 (Institute of Environmental Sciences, Mount Prospect, IL,
1993) pp. 373–379.
SEMI MF1617-0304 © SEMI 2003, 2004 4
Absorption Spectroscopy (VPD/AAS) or Vapor Phase
Decomposition/Inductively-Coupled-Plasma Mass
Spectrometry (VPD/ICP-MS) and shown by qualitative
SIMS to be spatially uniform, since spatial non-
uniformity can introduce variability or bias to the SIMS
quantitative measurement. With a spin coat
contaminated reference sample, it may be possible to
avoid the need for the crater depth measurement and its
bias/variability. For this reference sample, the areal
density D to be used in Equation 1 (see Section 12.1) is
the elemental areal density determined by the
VPD/AAs or VPD/ICP-MS, corrected for isotopic
abundance since the SIMS measurement of SI
i
is
isotopic.
10.1.1.2 Silicon Wafers — Dipped in an intentionally
contaminated SC-1 (NH
4
OH:H
2
O
2
:H
2
O) bath where
the calibration of the metal deposition is by VPD/AAS
or VPD/ICP-MS and shown by qualitative SIMS, or by
TXRF for iron, to be spatially uniform, since spatial
non-uniformity can introduce variability or bias to the
SIMS quantitative measurement. With an SC-1 dipped
reference sample, it may be possible to avoid the need
for the crater depth measurement and its
bias/variability. For this reference sample the areal
density D to be used in Equation 1 (see Section 12.1) is
the elemental areal density determined by the
VPD/AAS or VPD/ICP-MS, corrected for isotopic
abundance since the SIMS measurement of Si
i
is
isotopic.
10.1.1.3 Ion Implants — Such as
23
Na,
27
Al,
39
K, and
56
Fe, (or
54
Fe) implanted into silicon and shown by
qualitative SIMS, or by TXRF for iron to be spatially
uniform, since spatial non-uniformity can introduce
variability or bias to the SIMS quantitative
measurement. It is necessary to demonstrate that RSFs
obtained from these reference materials and the
instrumentation and operation used for the
measurement are consistent with RSFs from spin coated
reference samples.
5
For this reference sample, the areal
density D to be used in Equation 1 (see Section 12.1) is
the implanted isotopic dose.
11 Procedure
11.1 Specimen Loading and SIMS Instrument Set-up
11.1.1 Cleave or dice individual specimens (unknowns,
reference, BLANK) to fit within the sample holder.
The reference sample must contain
23
Na,
27
Al,
39
K, and
56
Fe (or
54
Fe) or there must be multiple reference
samples, each with one or more of these elements. Do
this preparation in such a way as to minimize
contamination of the specimens with surface metals.
11.1.2 Load the specimen(s) into the SIMS sample
holder.
11.1.3 Transfer the loaded holder into the SIMS
instrument sample chamber.
11.1.4 Turn on the instrument in accordance with the
manufacturer' s instructions.
11.1.5 Set the appropriate analytical conditions
including the method for molecular ion mass
interference discrimination.
11.1.5.1 Select a primary ion current, primary beam
raster size, and secondary spectrometer transmission,
which will result in an appropriate sputter rate (less
than 0.015 nm/s).
11.1.5.2 Select the secondary spectrometer conditions
so that the maximum secondary ion-count rates result in
dead time losses below 10%.
6
11.1.6 Confirm that analytical conditions are
appropriate to the measurement goals by measurement
of known reference standard(s) and BLANK sample.
11.1.6.1 Confirm that the analytical sputter rate is such
that each element monitored in a single profile is
sampled at a frequency greater than or equal to once
every 0.2 nm of sputtered depth.
11.1.6.2 If an oxygen jet is used, confirm that 0-leak
pressure is suitable by measuring a matrix ion depth
profile and demonstrating that matrix secondary ion
yield is constant (within 20%) during the analysis of the
first 10 nm of a specimen. Perform this confirmation
measurement on a typical specimen at the same sputter
rate to be used for the measurement of the surface metal
impurities. If the ion yield shows significant variation,
increase the 0-leak pressure by a factor of 2× and
reconfirm stability.
11.1.6.3 Depending upon the instrumentation used and
if needed, measure the detector efficiency ratio(s) for
any detectors used during the measurement (for
example, the electron multiplier and the faraday cup
detectors). Make this measurement by intercomparing
standard secondary ion signals of the appropriate
intensity (to minimize dead-time losses). The
secondary ion count rates used here may be different
for this measurement than for the analysis, and may be
generated under different sputter rate conditions than
the analysis.
11.2 Analysis of Specimen
11.2.1 Center the primary ion beam, initialize the
SIMS instrument control software, and begin a SIMS
profile. For sodium, aluminum, potassium, and iron,
the secondary ion signals generated by sputtering these
6 Liu, Y. A., and Fleming, R. H., “Reduced Electron Multiplier Dead
Time in Ion Counting Mass Spectrometry,” Rev. Sci. Instr, 64, 1661
(1993).
SEMI MF1617-0304 © SEMI 2003, 2004 5
contaminants from the free surface or native oxide layer
on the specimen should ordinarily reach a maximum
within the first 1 nm of the depth profile, and decrease
monotonically thereafter. If this general profile shape
is not obtained, repeat the measurement on a fresh
surface of the specimen.
11.2.2 When the impurity signal has either: (1)
decreased to at least <1% of the maximum, (2) or
reached a constant background count rate, then measure
and record the silicon matrix ion count rate on the
appropriate detector.
11.2.3 Measure the sputter rate under the analysis
conditions. This can be done by measuring the SIMS
crater depth with a stylus profilometer or equivalent
instrument if available, and combining the depth
measurement with the recorded time elapsed during the
sputter profile. If the analytical craters are too shallow
to be measured by available instrumentation, the sputter
rate must be monitored during a measurement series by
repeated analysis of a previously characterized standard
sample.
11.2.4 Measure the BLANK under the same analytical
conditions as the unknown specimens.
12 Calculations
12.1 The RSF for an element in a matrix can be
determined experimentally from the SIMS profile of a
reference material containing a known areal density of
the impurity of interest as follows:
()
NISId
tDNI
bi
m
=RSF (1)
where:
D = the areal density of the impurity in
atoms/cm
2
,
N = the number of data cycles in the profile,
I
m
= the matrix isotope secondary ion intensity,
(counts/s),
d = the sputtered depth, cm,
SI
i
= the sum of the impurity isotope secondary
ion-counts over the depth of the profile,
I
b
= a constant background intensity of the
impurity isotope, and
t = the analysis time for the species of interest,
s/cycle.
12.2 The calibration of concentration in a SIMS profile
is calculated as follows:
=
EM
FC
I
I
C
m
i
i
RSF (2)
where:
C
i
= the impurity atomic concentration,
atoms/cm
3
,
I
i
= the isotope secondary in intensity, counts/s,
I
m
= the matrix isotope secondary ion-intensity,
counts/s,
RSF = relative sensitivity factor, atoms/cm
3
, and
FC
EM
= the ratio of matrix intensities on the faraday
cup (FC) to the electron multiplier (EM),
when two detectors are used.
12.3 The conversion of data cycles or time into depth is
completed by measuring the crater depth and the total
time (or data cycles) of the sputtering that formed the
crater. The depth scale is assumed to be linear with
time or data cycles.
12.4 Calculate the RSFs and impurity concentration
versus depth from Equations 1 and 2.
12.5 Calculate the surface areal density by integrating
the concentration versus depth profile after any
subtraction of a constant background count rate that
may be appropriate.
13 Report
13.1 Report the following information:
13.1.1 The instrument used, the operator, and the date
of the measurements,
13.1.2 Identification of test, BLANK, and reference
specimens,
13.1.3 Calibration procedure used,
13.1.4 Method used to discriminate molecular ion
interferences,
13.1.5 Surface areal density of the impurity, and
13.1.6 Surface areal density of the impurity for the
BLANK specimen.
14 Precision and Bias
7
14.1 Precision — Tables 1, 2, and 3 summarize the
statistics from a spoke-wheel round robin including 4
test samples and 10 laboratories. Each laboratory was
provided with a spin coated reference standard of
surface sodium, aluminum, and potassium. The 95%
repeatability, r, and reproducibility, R, limits are
7 Support data are available from SEMI Headquarters. Request
SEMI MF1617 round robin report (formerly ASTM RR: F01-1010).