semi合集-English.pdf - 第258页

SEMI E43-0301 © SEMI 1995 , 2001 6 10.3 Demonstrating Ability to Meas u r e Examp le Item(s) Acceptably — T he certifier w ill have charged/uncharged example(s ) of items/objects for m eas urement of static field(s) by c…

100%1 / 7923
SEMI E43-0301 © SEMI 1995, 20015
9.2.3 Measurement Limitations — Measurements
made to this guide are only valid for surfaces that are
flat to a radius of 1.5 times the measurement distance
from a point directly below the sensor head. For
surfaces that are not flat, measurements should be made
by moving the sensor over the surface such that the
specified measurement distance is maintained as closely
as possible. These measurements may only be stated as
a range, with rounding as applicable to the meter's
measurement range according to Section 9.2.2. See
Figure 4, Example of a Survey of a Carrier of
Semiconductor Wafers. See Related Information 2 for
notes on test methods environment and measurements.
9.3 Electrostatic Voltmeter Measurements — Refer to
Figure 4.
9.3.1 Verifying the Voltmeter Verify the
performance of the voltmeter as in Section 7 above.
Check/reset the zero periodically per manufacturer's
instructions. Assure that the voltmeter and operator are
grounded.
9.3.2 Selecting the Voltmeter Select an electrostatic
voltmeter with a measurement range consistent with the
anticipated levels of charge on the objects to be
measured. The selection of too high a measurement
range will sacrifice voltage resolution, while selection
of too low a range will cause out-of-range operation
(saturation).
9.3.2.1 To measure moving objects, select an
electrostatic voltmeter with a response speed fast
enough to detect the objects when they are moving past
the electrostatic voltmeter probe at the highest
anticipated velocity.
9.3.2.2 Select a side- or end-viewing probe for the
electrostatic voltmeter as is best suited to view the
target object or surface when the probe is installed in an
apparatus.
9.3.3 Measurements Position the probe in front of
the surface to be measured. Best results are obtained
when the probe is placed less than two (probe) aperture
diameters from the object or surface to be measured.
At these closer spacings, the effects of extraneous fields
are minimized.
9.3.3.1 To resolve a small surface area, the distance
between the probe and the surface-under-measurement
must be less than 1/5
th
of the diameter of the surface
area to be measured. At wider spacings the surface area
resolved by the probe will exceed the surface area of
interest, and measurement accuracy may be reduced
together with the possibility of introducing effects of
extraneous fields to the measurement.
9.3.4 Measurement Limitations Voltage levels on
isolated conductors can be measured. Insulators do not
have a uniform surface charge distribution. Therefore,
it is considered that voltage levels measured on
insulators indicate an electrostatic field strength in a
particular area.
9.4 Electrometer MeasurementsMeasuring with an
Electrometer is very similar to measuring with any
other voltmeter or multimeter.
9.4.1 Connect the “common” terminal of the
electrometer through a test lead to the reference plane
or ground. Connect the “hot” or signal lead to the object
or test point of interest. Some electrometer
measurements will use a separate wire or shield
connected to the electrical ground or a guard ring.
Connect this as recommended by the manufacturer of
the equipment.
9.4.2 The major difference between the ordinary
voltmeter and the electrometer is the orders of
magnitude higher input impedance of the electrometer.
An electrometer will therefore pick up voltage signals
produced by stray electric fields, potentials associated
with noise currents, and artifacts caused by intentional
or unintentional ionization of the ambient air when
measuring high voltages. An electrometer can, for
example, be used to measure the triboelectric and the
piezoelectric properties of a piece of coaxial cable:
connect the cable under test to the electrometer, and
flex it or tap on it with a finger. The voltages induced
on the center conductor can be measured by the
electrometer.
10 Certification
10.1 Certification to survey areas to this guide is for
the person doing the certification (certifier) to assure
that the person being certified (certifyee) can calibrate
the meter and make acceptable measurement of known
static field(s) per Section 7 and applicable example(s)
per Section 9. The certifier shall be someone qualified
by education and/or training to calibrate and make
measurements with the equipment called out in this
guide or someone previously certified. The ESD
Association conducts such training programs and the
National Association of Radio and Television
Engineers (NARTE) administers an ESD Engineer and
ESD technician certification program.
10.2 Demonstrating Ability to Verify the Performance
of the Meter Against Known Source The certifyee
shall charge the test plate, zero the meter, and perform
the measurement a minimum of two times per Section
7; record values per Section 11. Readings obtained shall
be within 5% of expected values.
SEMI E43-0301 © SEMI 1995, 2001 6
10.3 Demonstrating Ability to Measure Example Item(s) AcceptablyThe certifier will have charged/uncharged
example(s) of items/objects for measurement of static field(s) by certifyee(s) a minimum of two times per Section 9;
record values per Section 11.
10.4 Certifying Readings obtained per Section 10.2 shall be within 12% of expected values and methodology of
obtaining readings shall be acceptable per certifier observation(s). A permanent record of certification is realized by
certifyee when certifier signs record sheet(s) for file in personnel records and/or when certifier issues a certificate.
Refer to Section R1-3.
11 Documentation
11.1 Meter calibration check(s), benchmark or laboratory measurements, items/areas surveys, and/or any other
electrostatic field measurements should be recorded in permanent records. Recorded are initial reading,
second/validation reading, and any subsequent readings taken to acceptable range/bound levels observed. Record
sheet(s) should show meter used, area (name), location, date, and items listed opposite readings, name of person
who took readings, and space for comments.
Charging
Source
STEP 1
Charge
Capacitor
Charging
Source
STEP 2
Disconnect
Charging
Source from
Capacitor
Coulombmeter
STEP 3
Connect
Coulombmeter
to Capacitor
Figure 1
Verifying Performance of the Coulombmeter
SEMI E43-0301 © SEMI 1995, 20017
A) Connect meter directly to charged plate ground reference.
B) Meter approximately 2.54 cm and parallel to charge plate.
C) Meter reading should be within 5% of voltage applied to plate.
D) Assure that both the meter and the operator are properly grounded.
Meter
Field emanating from charge plate
Plate
Ground Plane
Insulators
Charged Plate monitor
or power supply
Figure 2
Fieldmeter and Voltmeter Verification Check
Coulombmeter
Place object into
Faraday
enclosure.
Faraday Enclosure
Figure 3
Measurement With a Coulombmeter and Faraday Enclosure