semi合集-English.pdf - 第4727页

SEMI C3.55-02 00 © SEMI 2000 3 5.2.4 Calibration 5.2.4.1 Run titration bla nk consisting o f 1 50 mL deionized water. 5.2.4.2 Calcu la te the concentration of c h lorid e, using the formula below. The result may not exce…

100%1 / 7923
SEMI C3.55-0200 © SEMI 2000 2
5.1.4 Operating Procedures
5.1.4.1 Inject the calibration standard and sample as
described in Section 6.6 of SEMI C3. The order of
elution is nitrogen, argon, carbon monoxide, methane,
and carbon dioxide. (See NOTE 2.)
5.1.4.2 Compare the average peak area of the
calibration standard to that of the silane sample being
tested. Calculate the concentrations of the analytes as
described in Section 6.8 of SEMI C3. The result shall
be reported in the unit of concentration needed and to
the number of significant digits of the variable with the
least number of significant digits. This value may not
exceed that specified in Section 2.
5.2 Total Chlorides — This procedure is for the
determination of total chlorides in silane by titration of
a hydrolyzed sample of silane. (See NOTE 1.)
5.2.1 Detection Limit — 0.5 ppm (mole/mole)
5.2.2 Instrument Parameters — See Figure 1.
5.2.2.1 Equipment
5.2.2.1.1 Safety purge regulator with proper fittings.
5.2.2.1.2 Flow meter capable of measuring 0.25
standard liters per minute (0.5 SCF/hour) (Brooks or
equivalent), calibrated for silane.
5.2.2.1.3 One cylinder of nitrogen with regulator.
5.2.2.1.4 One ice bath.
5.2.2.1.5 Three magnetic stirring bars.
5.2.2.1.6 Two 2000 mL heavy-duty sidearm flasks
(Fisher Cat. No. 10-181G or equivalent).
5.2.2.1.7 One #9 one-hole rubber stopper.
5.2.2.1.8 One #9 two-hole rubber stopper.
5.2.2.1.9 Two Pyrex-brand tubes with fritted cylinders
(Fisher Cat. No. 11-138B or equivalent).
5.2.2.1.10 One gas washing bottle with fritted
cylinder, 125 mL cap. (Fisher Cat. No. 03-040A or
equivalent).
5.2.2.1.11 One buret, 50 mL capacity with Teflon
stopcock (Fisher Cat. No. 03-700-22C or equivalent).
5.2.2.1.12 One buret stand (Fisher Cat. No. 14-688 or
equivalent).
5.2.2.1.13 Miscellaneous clamps, support stands, and
rubber hose.
5.2.2.1.14 Two 250 mL Erlenmeyer flasks.
5.2.2.2 Reagents
5.2.2.2.1 Potassium hydroxide flakes, technical (Fisher
Cat. No. P-246 or equivalent).
5.2.2.2.2 Mercuric nitrate crystal (Fisher Cat. No. M-
168 or equivalent).
5.2.2.2.3 Sodium chloride crystal (Fisher Cat. No. S-
271 or equivalent).
5.2.2.2.4 Diphenyl carbazone (Fisher Cat. No. D-86
or equivalent).
5.2.2.2.5 Bromophenol blue (Fisher Cat. No. B-392 or
equivalent).
5.2.2.2.6 Nitric acid (Fisher Cat. No. AA-200 or
equivalent).
5.2.2.2.7 Ethanol, denatured (Fisher Cat. No. AA-407
or equivalent).
5.2.3 Operating Procedure
5.2.3.1 Assemble apparatus as shown in Figure 1 and
purge entire system with nitrogen for 30 minutes.
5.2.3.2 Fill each 2000 mL sidearm flask with about
1700 mL of 15% potassium hydroxide solution (15 g
KOH per each 100 mL water).
5.2.3.3 Fill the gas washing bottle with about 75 mL
of deionized water that has been degassed.
5.2.3.4 Pass exactly one cubic foot of silane through
the system at a rate of 0.5 cubic feet per hour.
5.2.3.5 Stop silane flow and purge system for 30
minutes.
5.2.3.6 Remove gas washing bottle fro m system and
transfer contents quantitatively and with the aid of three
25 mL deionized water washings, to a 250 mL
Erlenmeyer flask.
5.2.3.7 Add a few drops of indicator solution, 5 g
diphenyl carbazone plus 0.5 g bromophenol blue
dissolved in 750 mL ethanol, plus 250 mL deionized
water.
5.2.3.8 Add, in a drop-wise fashion, sufficient 0.2 N
HN0
3
(13 mL concentrated. HN0
3
, diluted to 1 L in
deionized water) in order to just turn the solution to
yellow from purple.
5.2.3.9 Titrate with a solution of mercuric nitrate
Hg(NO
3
)
2
, (diluted to 1000 mL in deionized water; then
50 mL diluted to 500 mL in deionized water) which has
been previously standardized to 5.0 mL of a sodium
chloride solution (approximately 165 ng NaCl,
accurately weighed, dissolved in 100 mL deionized
water).
SEMI C3.55-0200 © SEMI 20003
5.2.4 Calibration
5.2.4.1 Run titration blank consisting of 150 mL
deionized water.
5.2.4.2 Calculate the concentration of chloride, using
the formula below. The result may not exceed the
specification in Section 2 of this standard.
mL mL
mL mL
wt.NaCl(g)
100.0mL
60.66%Cl
100%NaCl
100mL
1L
0.005l
35.453g / mole
0.082atmL/ moleK
28.316L
sample blank
standard
blank
××
×
××
mL mL
mL mL
wt.NaCl(g)
100.0mL
60.66%Cl
100%NaCl
100mL
1L
0.005l
35.453g / mole
0.082atmL/ moleK
28.316L
sample blank
standard
blank
××
×
××
×
1 × 10
6
ppm = ppm Cl-
5.3 Hydrocarbons C
2
– C
4
— This procedure is for the
determination of hydrocarbons (methane, ethane, and
propane) in silane using a gas chromatograph with a
flame ionization detector.
5.3.1 Detection Limit — 50 ppb (mole/mole)
5.3.2 Instrument Parameters
5.3.2.1 Columns:
Column 1: VZ-10, 60/80, 2 m (6.6 ft) by 3.2 mm (1/8
in) OD by 2.2 mm (0.085 in) ID stainless
steel;
Column 2: VZ-10, 60/80, 1 m (3.3 ft) by 3.2 mm (1/8
in) OD by 2.2 mm (0.085 in) ID stainless
steel;
or equivalent.
5.3.2.2 Carrier Flow — 30 mL/min nitrogen
5.3.2.3 Sample Volume — 10 mL
5.3.2.4 Temperatures:
Detector 150°C
Column Oven 45°C
5.3.3 Calibration Standards — 1 ppm (mole/mole) for
all components to be tested, balance helium.
5.3.4 Operating Procedures (see SEMI C3, Section
6, Gas Chromatography for standard procedures).
5.3.4.1 Inject the calibration standard and sample as
described in Section 6.6 of SEMI C3. The order of
elution is methane, ethane, propane and butane (see
NOTE 2).
5.3.4.2 Compare the average peak area of the methane
peak in the calibration standard to that of the
hydrocarbons found in the silane sample being tested.
Calculate the concentrations of the analytes as
described in Section 6.8 of SEMI C3 based on the
methane response. The result shall be reported in the
unit of concentration needed and to the number of
significant digits of the variable with the least number
of significant digits. This value may not exceed that
specified in Section 2.
5.4
Hydrogen This procedure is for the
determination of hydrogen in silane using a gas
chromatograph with a thermal conductivity detector.
5.4.1 Detection Limit — 15 ppm (mole/mole)
5.4.2 Instrument Parameters
5.4.2.1 Columns:
Molecular sieve 5A, 2.4 m (8 ft) by 6.4 mm (1/4 in) OD
by 4.7 mm (0.185 in) ID, or equivalent.
5.4.2.2
Carrier Flow — 30 mL/min argon
5.4.2.3 Sample Volume — 1.0 mL
5.4.2.4 Temperatures:
Detector 40°C
Column Oven 40°C
5.4.3 Calibration Standards — 20–100 ppm
(mole/mole), balance argon.
5.4.4 Operating Procedures (see SEMI C3, Section
6, Gas Chromatography for standard procedures)
5.4.4.1 Inject the calibration standard and sample as
described in Section 6.6 of SEMI C3. (See NOTE 2.)
5.4.4.2 Compare the average peak area of the
hydrogen peak in the calibration standard to that of the
hydrogen found in the silane sample being tested.
Calculate the concentrations of the analytes as
described in Section 6.8 of SEMI C3 based on the
methane response. The result shall be reported in the
unit of concentration needed and to the number of
significant digits of the variable with the least number
of significant digits. This value may not exceed that
specified in Section 2.
5.5
Disiloxane, Disilane and Methyl Silane — This
procedure is for the determination of disiloxane,
disilane and methyl silane in silane using a gas
chromatograph with flame ionization detector.
5.5.1 Detection Limit — 0.1 ppm (mole/mole)
5.5.2 Instrument Parameters
5.5.2.1 Column:
28% DC-200 on Chromosorb PAW, 45/60 mesh, 14 m
(46 ft) SS, or equivalent.
5.5.2.2 Carrier Flow — 30 mL/min helium
5.5.2.3 Sample Volume — 2 mL
5.5.2.4 Temperatures — Column 60°C
SEMI C3.55-0200 © SEMI 2000 4
5.5.2.5 Calibration Standards — 1–5 ppm (mole/
mole)
5.5.3 Operating Procedure — (see SEMI C3, Section
6, Gas Chromatography for standard procedures).
5.5.3.1 Inject the calibration standard and sample as
described in Section 6.6 of SEMI C3. The order of
elution is silane matrix, methyl silane, siloxane (for
information only) and disilane. (See NOTE 2.)
5.5.3.2 Compare the average peak area of the
components being measured in the calibration standard
to that of the same analytes found in the silane sample
being tested. Calculate the concentrations of the
analytes as described in Section 6.8 of SEMI C3. The
result shall be reported in the unit of concentration
needed and to the number of significant digits of the
variable with the least number of significant digits.
This value may not exceed that specified in Section 2.
5.6
Water — This procedure is for the determination
of trace moisture (water) in silane using a vibrating
quartz hygrometer.
5.6.1 Detection Limit — 0.1 ppm (vol/vol)
5.6.2 Instrument Parameters
5.6.2.1 Flow Requirements — Set the sample pressure
and flow rate in accordance with the instrument
manufacturer’s instructions.
5.6.3 Calibration Standards — Check the hygrometer
calibration with silane using an external permeation
device. Construct a calibration curve which contains at
least three points covering the range of interest. The
standards employed will have been verified
independently on a condensation dewpoint/frostpoint
hygrometer. (See Figure 2.)
5.6.4
Operating Procedure — At the first start-up of
the hygrometer, calibrate it with moisture standards in
silane. Then check the calibration once a month and
modify the hygrometer if necessary. Direct standard to
unit with stainless lines which have been purged. Note
that silane is spontaneously flammable. Thus the
experimental set-up should be leak-free and thoroughly
purged before the introduction of silane. Obtain
representative sample of the gas to be analyzed and
direct to the unit. Wait until the equilibrium and read
the moisture content. The result may not exceed the
specification.
5.6.4.1
Construct a calibration curve with previously
verified standards. Direct standard to unit with
stainless steel lines which have been purged.
5.6.4.2 Obtain representative sample of gas to be
analyzed and direct to unit as with the standards. (See
NOTE 1.)
5.6.4.3
Construct a calibration curve and determine the
ppm moisture content in sample gas. The result may
not exceed the specification in Section 2 of this
standard.
5.7 Elemental Analysis in Deposited Layer
Procedures for the analysis of elemental impurities in
deposited layers of polysilicon should be designated
and should be referenced from a recognized industry
standard (including but not limited to ISO, ASTM,
etc.).
5.7.1 Typical Detection Limits
Compound Detection Limit
Carbon (C) 0.2 ppma
Oxygen (O) 0.2 ppma
Boron (B) 0.02 ppba
Aluminum (Al) 0.02 ppba
Gallium (Ga) 0.02 ppba
Phosphorus (P) 0.02 ppba
Arsenic (As) 0.02 ppba
Antimony (Sb) 0.02 ppba
Iron (Fe) 0.3 ppba
Chromium (Cr) 0.01 ppba
Nickel (Ni) 0.2 ppba
Copper (Cu) 0.02 ppba
Zinc (Zn) 0.03 ppba
5.8 Resistivity — Procedures for the measurement of
resistivity in deposited layers of polysilicon should be
designated and should be referenced from a recognized
industry standard (including but not limited to ISO,
ASTM, etc.).
NOTE 1: Before performing this or any other analytical
operation it is necessary to perform the following:
Familiarize yourself with the products involved and the
potential hazards by consulting the MSDS's for all
hazardous substances involved and by consulting the
CGAs “Handbook of Compressed Gases” for all gases
involved.
Know how to safely handle and store each hazardous gas
by consulting the applicable CGA pamphlets.
Use all necessary and applicable personal protective
equipment.
Follow all instructions relative to the safe operation of
the equipment by consulting the LAC Safety Manual and
applicable equipment manufacturers manual(s).
It must be noted that silane is a pyrophoric gas and all sample
plumbing must be purged of air completely before introducing
silane. Extra special attention must be paid to leak checking
all connections and checking the vent plumbing to be certain
that effluents are vented properly. If there are any