semi合集-English.pdf - 第7573页

SEMI MF1982-1103 © SEMI 2003 4 7.1.2.3 Cleave the waf er sample into strips abo ut 5 to 7 mm wide by scri bing a line on the wafer where one wants it to brea k and then placing i t on a straight wire and breakin g it alo…

100%1 / 7923
SEMI MF1982-1103 © SEMI 2003 3
atomic emission detector (AED). The
nitrogen/phosphorus thermionic ionization detector
(NPD) also responds to nitrogen containing
compounds.
NOTE 1: A nitrogen/phosphorus thermionic ionization
detector (NPD) may also be used as a phosphorus selective
detector. This type of detector also responds to nitrogen
containing compounds. If an NPD is used, the total
organophosphorus reported should exclude any signals due to
nitrogen containing compounds. Often, identification from
the mass spectra can be used to determine whether compound
contains nitrogen or phosphorus, or both.
6.1.2 Sample Thermal Desorption Tubes — Stainless-
steel tubes packed with adsorbent medium, are used to
trap compounds of interest and release them onto a
thermal desorption unit.
NOTE 2: Note that stainless steel is catalytically active and
can corrode with time, affecting recovery for some
compounds. In this case, deactivated stainless steel, glass, or
quartz tubes also can be used. Several adsorbent materials
can be used for trapping organic compounds desorbed from
silicon wafer samples. Some examples are activated carbon,
graphitized carbon, and poly (2,6-diphenyl-p-phenylene
oxide).
6.1.3 Thermal Desorption Unit — Used to desorb
organics from sample thermal desorption tubes. The
thermal desorption unit is coupled to the GC instrument
via a heated transfer line, for example, heated to 225°C
or above.
6.1.4 Thermal Annealer — Or a high temperature
furnace used to obtain a blank wafer (see Section
7.1.1.2 ).
6.1.5 Temperature Controllable Wafer Desorption
Oven — Used to hold and heat a wafer desorption tube.
6.1.6 Wafer Desorption Tube — Stainless-steel tube
with an approximate dimension: 12.7 mm (0.5 in.) in
outer diameter, 9.53 mm (0.375 in.) in inner diameter
and 254 mm (10 in.) in length (see Figure 1), used to
desorb organics from cleaved wafers. Larger tubes may
be used for larger wafer sizes and to increase the
sensitivity of the test (see Note 2).
6.1.7 Analytical Balance
6.1.8 Quartz Plate and Carbide-Tipped Scribe — Used
to cleave wafer samples.
6.1.9 Quick Connect — Used to connect a nitrogen line
to the wafer desorption tube.
6.2 Method B
6.2.1 GC Instrument — Utilizes a capillary column to
separate a wide variety of organic compounds, coupled
to a mass spectrometer (MS) or a phosphorus selective
detector, or both. Examples of phosphorus selective
detectors are flame photometric detector (FPD), atomic
emission detector (AED), or nitrogen/phosphorus
thermionic ionization detector (NPD). The
nitrogen/phosphorus thermionic ionization detector
(NPD) also responds to nitrogen containing compounds
(see Note 1).
6.2.2 Quartz Chamber Unit — Used to desorb organic
contaminants from the wafer surface and transfer them
to a glass TD tube (see Figure 2).
6.2.3 Quartz Chamber — May be used also to purge
out any organic contaminants from a blank wafer.
6.2.4 Glass TD Tube — Packed with adsorbent
medium used to trap compounds of interest and release
them onto a cold trap. Several adsorbent materials can
be used for trapping organic compounds desorbed from
silicon wafer samples. Some examples are activated
carbon, graphitized carbon, and poly (2,6-diphenyl-p-
phenylene oxide).
6.2.5 Cold Trap — Used to concentrate organics
desorbed from glass TD tube and release the organics to
a GC instrument by rapid heating.
7 Procedure
7.1 Method A
7.1.1 Sample Handling and Preparation
7.1.1.1 At all times, avoid manual handling of samples
for analysis to prevent any secondary contamination of
samples. Use stainless steel tweezers for sample
preparation. Use a propane torch to flame stainless
steel tweezers, weigh boats, and other accessories that
come into direct contact with the sample before their
usage.
7.1.1.2 Thermally debsorb any organics from blank
wafers in a rapid thermal annealer or a high temperature
furnace. In order to desorb any organics from blank
wafers, the temperature of a rapid thermal annealer is
kept at 900°C for 15 s, or the temperature of a high
temperature furnace is kept at 700°C for 30 min.
Heating under air or oxygen will make a consistent
organic-free SiO
2
surface. Place the thermally treated
blank wafers directly into petri dishes and wrap the
dishes with organic-free aluminum foil.
7.1.2 Wafer Desorption
7.1.2.1 Clean and precondition sample thermal
desorption tubes in accordance with ASTM Practice
D 6196 prior to attaching to the wafer desorption tube.
7.1.2.2 Turn on the temperature controller for the wafer
desorption oven and wait for the temperature to reach
275°C.
SEMI MF1982-1103 © SEMI 2003 4
7.1.2.3 Cleave the wafer sample into strips about 5 to 7
mm wide by scribing a line on the wafer where one
wants it to break and then placing it on a straight wire
and breaking it along the scribed line. The portion of
the wafer sample needed for the analysis changes with
the size of the wafer. For a 100-mm wafer, about half
of its surface area is needed. Larger samples (having
larger surface area) may be used to increase the
sensitivity of the test.
7.1.2.4 Place the wafer strips of wafer into the wafer
desorption tube. Attach a sample thermal desorption
tube to the side of the wafer desorption tube. Connect
the other end of the wafer desorption tube to a nitrogen
line with a flow of 15 to 100 mL/min.
7.1.2.5 Place the wafer desorption tube in the oven and
heat it for 30 min at 275°C. This allows the volatile
organics to be thermally desorbed from the sample
wafer and transferred to the sample thermal desorption
tube.
7.1.2.6 Seal both ends of the sample thermal desorption
tube into the thermal desorption unit or by using caps if
the thermal desorption-GC system is automated. Tubes
must not be left open and exposed to the laboratory
atmosphere. The end caps should be made of a material
that has low outgassing and low reactivity properties,
such as poly tetrafluoroethylene (PTFE), or stainless
steel.
7.1.2.7 Analyze a blank wafer at least daily. Use a
blank wafer treated with the same desorption procedure
just described. Report blank data with the sample data.
7.1.3 GC Analysis
7.1.3.1 Place the sample thermal desorption tube,
which contains organic contaminants desorbed from the
wafer, onto the thermal desorption unit. Enter the
sample information into the computer, and start the GC
analysis.
7.1.3.2 Flow the helium carrier gas through the sample
thermal desorption tube during primary desorption.
Heat the sample thermal desorption tube and sweep out
the target organics contained in the sample to a cold
trap where they are preconcentrated. Two
recommended conditions for heating the sample
desorption tube are as follows. When graphitized
carbon is used as adsorbent material, heat the tube to
400°C and hold at 400°C for 15 min. When poly (2,6-
diphenyl-p-phenylene oxide) is used as adsorbent
material, heat the tube to 270°C and hold at 270°C for
15 min. Cold trap parameters, that is, sorbent, sold
temperature, etc., should be selected such that all target
analytes are retained quantitatively throughout the tube
desorption process, for example, using a cold trap
temperature of 30°C. Refer to ASTM Practice D 6196
for a method for testing desorption efficiency and
analyte recovery. At the end of this period, heat the
cold trap rapidly to 300°C to release the target organics
to the GC column head.
NOTE 3: For higher sensitivity analysis, larger sample sizes
can be used, but clogging by water is possible. In this case, a
subambient cold trap with an adsorbent may be used and
maintained higher than 0°C during trapping to prevent freeze-
up of trap.
7.1.3.3 Separate volatile organics desorbed from the
sample thermal desorption tube by an appropriate
column temperature program. Two recommended
temperature programs are as follows. For high
resolution analysis, using polydimethylsiloxane-coated
column (60 m by 0.25 mm by 0.25–µm film thickness),
heat the column from 40 to 280°C at a rate of 10°C/min
and hold the temperature at 280 °C for 16 min. For
rapid analysis, using polydimethylsiloxane-coated
column (25 m by 0.32 mm by 0.52–µm film thickness),
heat the column from 30 to 265°C at a rate of
12.5°C/min and hold the temperature at 265°C for 16
min.
7.1.3.4 Use n-Hexadecane, n-C
16
H
34
(C
16
), as a standard
for total organic analysis 4
. Use tris (2-chloroethyl)
phosphate, (ClCH
2
CH
2
O)
3
PO (TCEP) or tributyl
phosphate, (C
4
H
9
O)
3
PO (TBP), as a standard for
organophosphorus analysis (see Section 6.1.1 ). These
standards are introduced into the GC system from a
precleaned sample thermal desorption tube. Two-point
calibration method is recommended for checking the
instrument performance. Recommended standards are
two different concentrations of n-C
16
H
34
(C
16
) for total
organic analysis; and, two different concentrations of
tris (2-chloroethyl) phosphate (TCEP) for total
organophosphorus analysis. Recommended frequency
of these measurements is at least once a week.
NOTE 4: AED and FID are recommended for accurate total
carbon quantification.
7.1.4 Quantification
7.1.4.1 Obtain each area below for the calculation of
total organic contaminants. Obtain total organic
contaminants of the sample wafer (A
s
) by summing up
all the peak areas integrated from GC-AED, FID, or
MS chromatogram of the sample wafer. Obtain total
area of the blank wafer (A
b
) by summing up all the peak
areas integrated from GC-AED, FID, or MS
chromatogram of the blank wafer. Obtain the area of
the standard peak (A
c
) integrated from GC-AED, FID,
or MS chromatogram of the standard sample.
7.1.4.2 Obtain each area below for the calculation of
total organophosphorus (TP). Obtain total area of
phosphorus compounds of the sample wafer (A
s
) by
SEMI MF1982-1103 © SEMI 2003 5
summing up all the peak areas integrated from GC-
(phosphorus selective detector) chromatogram of the
sample wafer. Obtain total area of the blank wafer (A
b
)
by summing up all the peak areas integrated from GC-
(phosphorus selective detector) chromatogram of the
blank wafer. Obtain the area of the standard peak (A
p
)
integrated from GC-(phosphorus selective detector)
chromatogram of the standard sample.
7.1.4.3 Before starting analysis, weigh and get the total
weight of each sample wafer to be tested, and also the
weight of unused part of each wafer. These weights are
used for the calculation of total organic contaminants
and total organophosphorus (TP).
7.2 Method B
7.2.1 Sample Handling and Preparation
7.2.1.1 At all times, avoid manual handling of samples
for analysis to prevent any secondary contamination of
samples. Use stainless steel tweezers for sample
preparation. Use a propane torch to flame stainless
steel tweezers and other accessories that come into
direct contact with the sample before their usage.
7.2.1.2 Thermally desorb any organics from blank
wafers in the quartz chamber unit. Recommended
conditions for preparing blank wafers are to heat them
for 30 min in the quartz chamber kept at 700°C.
Usually, helium is used as purge gas. Heating under air
or oxygen will make a consistent organic-free SiO
2
surface.
7.2.2 Wafer Desorption
7.2.2.1 Enter the analytical information into the quarts
chamber unit, cold trap and GC. Wait for the system to
be stabilized. Place the sample wafer into the quartz
chamber unit and start it. Supply helium gas into the
quartz chamber and heat the chamber from initial
temperature (40°C or lower) to 400°C at a rate of 10 to
30°C/min, and hold the temperature for 15 min. Sweep
out the volatiles contained in the sample to a glass TD
tube where they are adsorbed.
7.2.2.2 Analyze a blank wafer at least daily. Report
blank data with the sample data.
7.2.3 GC Analysis
7.2.3.1 Heat the glass TD tube and sweep out the target
organics adsorbed in the tube to a cold trap where they
are preconcentrated. Two recommended conditions for
heating the glass TD tube are as follows:
When graphitized carbon is used as adsorbent
material, heat the tube to 400°C and hold at 400°C
for 15 min.
When poly (2,6-diphenyl-p-phenylene oxide) is
used as adsorbent material, heat the tube to 270°C
and hold for 15 min.
7.2.3.2 Select cold trap parameters, that is, sorbent,
cold temperature, etc., such that all target analytes are
quantitatively retained throughout the tube desorption
process, for example using a cold trap temperature of
–130°C. Refer to ASTM Practice D 6196 for a method
for testing desorption efficiency and analyte recovery.
At the end of this period, heat the cold trap rapidly to
300°C to release the target organics to the GC column
head.
7.2.3.3 Separate volatile organics desorbed from the
glass TD tube by an appropriate column temperature
program. Two recommended temperature programs are
as follows. For high resolution analysis, use
polydimethylsiloxane coated column (60 m by 0.25 mm
by 0.25-µm film thickness), heat the column from 40 to
280°C at a rate of 10°C/min and hold the temperature at
280°C for 16 min. For rapid analysis, use
polydimethylsiloxane coated column (25 m by 0.32 mm
by 0.52-µm film thickness), heat the column from 30 to
265°C at a rate of 12.5°C/min and hold the temperature
at 265°C for 16 min.
7.2.3.4 Use n-Hexadecane, n-C
16
H
34
(C
16
), as a standard
for total organic analysis (see Note 4). Use tris (2–
chloroethyl) phosphate, (ClCH
2
CH
2
O)
3
PO (TCEP) or
tributyl phosphate, (C
4
H
9
O)
3
PO (TBP), as a standard
for organophosphorus analysis (see Section 6.1.1).
These standards are introduced into the GC system
from a precleaned wafer. A two-point calibration
method is recommended for checking the instrument
performance. Recommended standards are two
different concentrations of n-C
16
H
34
(C
16
) for total
organic analysis; and two different concentrations of
tris (2-chloroethyl) phosphate (TCEP) for total
organophosphorus analysis. Recommended frequency
of these measurements is at least once a week.
7.2.4 Quantification
7.2.4.1 Obtain each area below for the calculation of
total organic contaminants. Obtain total area of organic
contaminants of the sample wafer (A
s
) by summing up
all the peak areas integrated from GC-AED, FID, or
MS chromatogram of the sample wafer. Obtain total
area of the blank wafer (A
b
) by summing up all the peak
areas integrated from GC-AED, FID, or MS
chromatogram of the blank wafer. Obtain the area of
the standard peak (A
c
) integrated from GC-AED, FID,
or MS chromatogram of the standard sample.
7.2.4.2 Obtain each area below for the calculation of
total organophosphorus (TP). Obtain total area of
phosphorus compounds of the sample wafer (A
s
) by
summing up all the peak areas integrated from GC-