SEMI S3-91 © SEMI 199 1 1 SEMI S3-91 SAFETY GUIDELINES FOR HEATED CHEMICAL BATHS NOTICE: These gu idelines do not purport to addres s all of the safety issues associated with their use. It is the responsibilit y of the u…

100%1 / 7923
SEMI S2-0703a
E
© SEMI 1991, 2004 102
the frequency and complexity of operating, servicing and maintenance tasks; and
safety critical parts.
NOTE #: The term mishap was replaced with the results of harm in the SEMI S10 1103 revision.
D2-4 Revisions to Section 11.6 — OPTIONAL Before Effective Date
D2-4.1 Revision of the following in Section 11.6:
NOTE 26: Where a safety interlock is provided to safeguard personnel from a
severe or catastrophic mishap harm as categorized
by SEMI S10, consideration of positive-opening type of switches is recommended.
NOTICE: Paragraphs entitled “NOTE” are not an official part of this safety guideline and are not intended to
modify or supersede the official safety guideline. These have been supplied by the committee to enhance the usage
of the safety guideline.
SEMI makes no warranties or representations as to the suitability of the guideline set forth herein for any particular
application. The determination of the suitability of the guideline is solely the responsibility of the user. Users are
cautioned to refer to manufacturer’s instructions, product labels, product data sheets, and other relevant literature
respecting any materials mentioned herein. This guideline is subject to change without notice.
The user’s attention is called to the possibility that compliance with this guideline may require use of copyrighted
material or of an invention covered by patent rights. By publication of this guideline, SEMI takes no position
respecting the validity of any patent rights or copyrights asserted in connection with any item mentioned in this
guideline. Users of this guideline are expressly advised that determination of any such patent rights or copyrights,
and the risk of infringement of such rights, are entirely their own responsibility.
Copyright by SEMI® (Semiconductor Equipment and Materials
International), 3081 Zanker Road, San Jose, CA 95134. Reproduction of
the contents in whole or in part is forbidden without express written
consent of SEMI.
SEMI S3-91 © SEMI 19911
SEMI S3-91
SAFETY GUIDELINES FOR HEATED CHEMICAL BATHS
NOTICE: These guidelines do not purport to address
all of the safety issues associated with their use. It is the
responsibility of the users of these guidelines to
establish appropriate safety and health practices and
determine the applicability of regulatory limitations
prior to use.
NOTE: As listed or revised, all documents cited shall be the
latest publications of adopted standards.
1 Purpose/Scope
This safety guideline pertains to open top heated
chemical baths, or “wet stations”, used in the
manufacturing of semiconductors. The purpose of this
guideline is to provide general safety information for
the construction, operation, and maintenance of
chemical baths heated by electric immersion, externally
bonded or heat exchange type heating devices. This
guideline is not intended to provided detailed design
information for the individual components of the
heating systems, but rather to supply the safety criteria
for the design of these systems. (See Figure 1.)
2 Introduction
2.1 Failure to follow the manufacturer’s instructions as
to the application or installation of heating systems used
for chemical baths can result in accidents and fires in
which such systems may be implicated.
2.2 The human element plays a major role in how well
the hardware is used. The user of heated chemical bath
systems needs to understand the available technology to
minimize the risk involved.
2.3 The issue of bath operation and maintenance has
been included in this guideline to support the safe use
and maintenance of heated baths over their normal
service life.
3 Heated Bath Methodology
3.1 Electric Immersion Heaters Electric immersion
bath heaters are resistive element heaters. The element
itself is typically protected by an outer sheathing which
acts as a corrosion and electrical barrier. The element is
mounted to the inside of the bath.
3.2 External Tank Heaters — External type heaters
are normally bonded by a mastic to the outside of a
chemical bath. The most common types are
manufactured from a thin metal foil material that results
in a very low profile configuration.
3.3 Heat Exchangers — The most common type of
heat exchanger employs a separate heating source that
may be remote from the station. This heating source can
be electrically powered to generate steam or some other
heat transfer fluid. The heat transfer fluid is then
delivered by chemically compatible plumbing to the
bath and recirculated to the heater.
4 Minimum Heated Bath Saf ety Requirements
To ensure the safe operation and maintenance of
heating systems, the following features are required.
overcurrent protection
power interrupt
manual reset
automatic temperature controller
liquid level sensor
overtemperature protection
proper grounding and ground fault protection
compatible construction materials
NOTE: Interlocks should operate as outlined in SEMI S2.
These features should be totally compatible with each other.
4.1 Overcurrent ProtectionA fused disconnect
switch or circuit breaker sized for the amperage of the
heater.
4.2 Power Interrupt — In the event of an
overtemperature, low liquid level, or ground fault
condition, power to the heater(s) is interrupted upon
receiving the respective signal. The interrupt should be
separate from any relay incorporated into the automatic
temperature controller. The interrupt should be wired in
such a manner that any signal from the liquid level
sensor, overtemperature sensor, or ground fault
condition should shut off power to the heaters and place
the system in a “fail safe” condition.
4.3 Manual ResetThis reset should be incorporated
into the power interrupt device so that when the system
shuts off power to the heating element, a manual reset
is required to re-energize the system.
4.4 Automatic Temperature ControllerAn
automatic temperature controller maintains the liquid at
a set temperature by turning the heating element on and
off. The controller is activated by signals from a sensor
that is located at the bath. To preserve the integrity of
this system, quarterly service/testing and calibration in
SEMI S3-91 © SEMI 1991 2
accordance with the manufacturer’s recommendations
should be performed.
4.5 Liquid Level Sensor — This sensor will act to shut
down the heating element if the liquid level falls below
a safe operating level. When employing either an
immersion or externally mounted type heater, it is
recommended that this level be no less than five
centimeters (two inches) above the heating element’s
“hot zone” as established by the manufacturer. This
should prevent the element from losing the heat sink
created by the liquid.
4.5.1 The selection of a suitable sensor should take
into account:
4.5.1.1 The physical properties of the liquids being
heated and the environmental conditions to which the
sensor(s) will be subjected.
4.5.1.2 The mode of failure, if the sensor fails.
4.5.2 To preserve the integrity of these components,
they should be tested on a quarterly basis and serviced
following the manufacturer’s recommendations.
4.6 Overtemperature Protection
4.6.1 Liquid Overtemperature — A dedicated liquid
overtemperature sensor should typically be set to trip at
a temperature below the autoignition temperature of the
chemical being used in the bath, and in no case higher
than 10°C (18°F) above the normal operating
temperature.
4.6.1.1 This temperature setting should compensate
for the initial over-shoot often encountered when
bringing the liquid from ambient to operating
temperature.
4.6.1.2 The sensor should be located in the liquid at an
elevation approximating the top of the heating elements
“hot zone” and no more than seven centimeters (three
inches) away from the heater.
4.6.2 Heating Element OvertemperatureWhen
using an electrical immersion type heater, a sensor
should be used to shut down the heating element if the
surface temperature of the element exceeds a preset
value. The sensor is typically placed between the
resistance coil and the protective sheath covering.
4.7 Proper Grounding & Ground Fault Protection
The amount of insulation surrounding a heater element
depends upon the type of heater. This insulation not
only provides protection from the corrosive properties
of the solution, but also acts as an electrical barrier. If
the barrier breaks down, it is possible for the solution to
become energized.
4.7.1 The electric heater elements should be equipped
with a ground, the construction of which should be
approved by a nationally recognized testing laboratory
(NRTL).
4.7.2 A ground fault circuit interrupter (GFCI) should
be used to preclude worker exposures to potential
electrical shock.
4.7.3 The station or equipment in which the heated
bath will be used needs to be reviewed to ensure that all
National Electrical Code requirements have been
addressed.
4.8 Compatible Construction Materials — All
portions of the station that could come in contact with
process chemicals should be constructed from materials
compatible with these chemicals. It is recommended
that metal materials be used in the station’s construction
when using combustible or flammable solutions.
5 Operation and Maintenanc e
5.1 Before heated baths are put into service, a detailed
review of the wet station operation and maintenance
should be conducted for all employees operating or
servicing the station. New employees should receive the
same information prior to work assignment.
5.1.1 All safety features included with the station’s
heating system should be user tested, and the frequency
of testing should be noted in the user’s periodic
maintenance procedures.
NOTE: Recommended minimum frequency of testing is
quarterly and after any station maintenance.
5.1.2 Vendor-supplied information on the bath
operation and limitations that may apply should be
reviewed and appropriate employee training conducted.
5.1.3 When the user plans to install the baths and to
perform maintenance, the suppliers should provide
guidance to ensure that all aspects of the system, such
as the controller and the heated bath safety systems, are
integrated properly.
5.1.4 Review the process requirements to ensure that
operating temperatures are within the range and
capability of the safety controls. For flammable and
combustible liquids, operating temperatures should be
below the ignition temperature.
5.1.5 Limit the interchanging of parts that were not
included in the original design unless so advised by the
supplier. Alternate parts can lead to failures resulting in
a fire or physical injury.
6 Related Documents
6.1 UFC 1988 edition.
6.2 OSHA Safety & Health Standards (29 CFR 1910).