SEMI M46-1101 E © SEMI 2001 1 SEMI M 46-1101 E TEST ME THOD FOR MEASURING CA RRIER CONCENTRATIONS IN EPITA XI A L LA YER STRUCTURES BY ECV PROFI LING This te st m ethod was tec hnically approv ed by the G lobal Compound …

SEMI M45-0703 © SEMI 2001, 2003 5
APPENDIX 1
NOTICE: This appendix was approved as an official part of SEMI M45 by full letter ballot procedure, but the recommendations
in this appendix are optional and are not required to conform to this standard.
A1-1.1 The 300 mm Wafer shipping box must be re-
usable in view of the wafer manufacturing cost and
environmental concerns.
A1-1.2 When the shipping box is re-used, the customer
who received wafers must keep the shipping box in
good condition after opening the package and return to
the wafer supplier.
A1-1.3 The box cleaning process of the supplier can
not remove ink, adhesive, and excessive metal
contamination of the returned shipping box.
A1-1.4 The customers must avoid contaminating the
shipping boxes in this way. The customers must keep
the shipping box separate from those contaminated
conditions. Contaminated shipping boxes must not be
returned to the supplier.
NOTICE: SEMI makes no warranties or
representations as to the suitability of the standards set
forth herein for any particular application. The
determination of the suitability of the standard is solely
the responsibility of the user. Users are cautioned to
refer to manufacture' s instructions, product labels,
product data sheets, and other relevant literature,
respecting any materials or equipment mentioned
herein. These standards are subject to change without
notice.
By publications of this standard, Semiconductor
Equipment and Materials International (SEMI) takes no
position respecting the validity of any patent rights or
copyrights asserted in connection with any items
mentioned in this standard. Users of this standard are
expressly advised that determination of any such patent
rights or copyrights, and the risk of infringement of
such rights are entirely their own responsibility.
Copyright by SEMI® (Semiconductor Equipment and Materials
International), 3081 Zanker Road, San Jose, CA 95134. Reproduction o
f
the contents in whole or in part is forbidden without express written
consent of SEMI.

SEMI M46-1101
E
© SEMI 2001 1
SEMI M46-1101
E
TEST METHOD FOR MEASURING CARRIER CONCENTRATIONS IN
EPITAXIAL LAYER STRUCTURES BY ECV PROFILING
This test method was technically approved by the Global Compound Semiconductor Committee and is the
direct responsibility of the European Compound Semiconductor Materials Committee. Current edition
approved by the European Regional Standards Committee on June 29, 2001. Initially available at
www.semi.org December 2001; to be published March 2002.
E
This document was editorially modified in November 2001 to correct a typographical error. A change was
made to Section 7.3.3.
1 Purpose
1.1 The purpose of this document is to specify a
method to measure the carrier concentration and carrier
concentration vs. depth profile of epitaxial layers by
Electrochemical Capacitance Voltage ECV profiling.
2 Scope
2.1 This test method covers a procedure for measuring
the carrier concentration of epitaxial layers by ECV
profiling. This method focuses on improving the
accuracy and repeatability of the measurement by
standardizing the test conditions and reporting and by
routine calibration of the measurement.
2.2 This test method is intended to cover the majority
of routine samples measured. However, because of the
number of different materials encountered it cannot
cover every contingency.
2.3 This standard may involve hazardous materials.
This standard does not purport to address safety issues,
if any, associated with its use. It is the responsibility of
the users of this standard to establish appropriate safety
and health practices and determine the applicability of
regulatory limitations prior to use.
3 Referenced Standards
3.1 SEMI Standards
SEMI C1 — Specifications for Reagents
3.2 ASTM Standards
D1125-95 (1999) — Standard Test Methods for
Electrical Conductivity and Resistivity of Water
4 Terminology
4.1 bias — the potential applied to the sample with
respect to a reference electrode.
4.2 capacitance voltage CV measurements — electrical
measurements where the capacitance of a rectifying
barrier is measured as a function of applied bias and is a
measure of the net fixed ionized charge per unit
volume.
4.3 carrier concentration — the net fixed ionized
charge per unit volume. Equal to the free carrier
concentration if the dopant is fully ionized and the
material is free of traps.
4.4 current voltage IV measurements — electrical
measurements where the current through the rectifying
barrier is measured as a function of applied bias.
4.5 dissipation factor — the ratio of the real part to the
imaginary part of the complex admittance. It is a
measure of the non-ideality of the barrier.
4.6 electrochemical — chemical reaction in which
charge transfer takes place via an external circuit.
4.7 epitaxial layer — a layer of single crystal
semiconductor material grown on a host substrate
which determines its orientation.
4.8 excess area — the difference between the wetted
and illuminated areas.
4.9 flatband potential — the intercept on the voltage
axis of the 1/C
2
vs V plot. A measure of the built in
field or barrier height.
4.10 illuminated area — the area of the sample which
can be illuminated during electrochemical etching.
4.11 rectifying barrier — a potential gradient formed
at the junction between two materials which permits the
flow of charge in one direction only.
4.12 reference electrode — half cell which has a
constant electrode potential, such as a saturated calomel
electrode, SCE.
4.13 rest potential — the open circuit potential of the
sample with respect to the reference electrode.
4.14 wetted area — the area of contact between the
electrolyte and the sample.
5 Summary of Method
5.1 In this method, the carrier concentration of the
epitaxial layer is measured by the CV method.
5.1.1 Ohmic contacts are formed on the sample.

SEMI M46-1101
E
© SEMI 2001 2
5.1.2 A rectifying barrier, of known contact area, is
formed by placing the sample in contact with an
electrolyte.
5.1.3 The quality of the barrier is assessed using I-V
and C-V measurements.
5.1.4 From the measured C-V data the carrier
concentration of the layer is determined.
5.1.5 The sample is anodically etched to produce a new
barrier deeper in the material. The etch depth is
determined using Faraday’s law of electrolysis.
5.1.6 Sections 5.1.4 and 5.1.5 are repeated to generate
a carrier concentration vs. depth profile.
6 Interferences
6.1 Thin Layers — Diffusion of carriers into adjacent
regions set a fundamental resolution limit to the
measurement. In addition if the layer has a high sheet
resistivity this can also affect the value obtained.
6.2 Double Layer Capacitance — A limitation
imposed by the physical nature of the
electrolyte/semiconductor barrier which sets an upper
limit to the measured carrier concentration, for accurate
measurements, of 1 × 10
19
cm
-3
. The technique can be
used beyond this value, but with progressive
degradation.
6.3 Etch Well Uniformity
6.3.1 Etch Well Flatness — Non-uniformity increases
the effective area and may result in the measurement
not being wholly made in the layer of interest. The
main causes of non-uniformity are gas bubbles,
inadequate electrolyte circulation and uneven
illumination.
6.3.2 Etch Well Roughness — indicates poor etching
and has a deleterious effect on the measurements.
Generally overcome by selecting a more appropriate
electrolyte and/or etching conditions.
6.4 Non-Ideal Electrical Characteristics
6.4.1 Series Resistance — From the contacts, sample
and electrolyte affect the accuracy of the measurements.
This effect is minimized by using a highly conductive
electrolyte, making large area ohmic contacts and by
employing a low measurement frequency.
6.4.2 Leakage Currents — Parallel conduction can
affect the accuracy of the measurement. Its influence is
reduced by employing a high measurement frequency.
6.4.3 Large Excess Area — Can give rise to errors on
n-type material particularly when profiling Hi-Lo
structures. The excess area is electrolyte dependent but
is mainly affected by the quality of the seal used to
define the area. Large excess areas indicate a poor seal
and are corrected by replacing the seal.
6.5 Hi-Lo Structures — Measurement is subject to
inaccuracy due to carrier diffusion and excess area
errors. For accurate measurements, chemically etch to
the layer of interest before measurement.
6.6 Contact Quality — Non-Ohmic or high resistance
contacts have a deleterious effect on the measurement.
Contacts should be checked by measuring the resistance
between two similar contacts and then reversing the
current direction and repeating the measurement.
6.7 Surface and Deep States — Cause the measured
capacitance to be frequency dependent and/or
dependent on the rate at which the bias is changed.
Usually resolved by using a high measurement
frequency. This should not be confused with the
frequency dependence which is more usually attributed
to high series resistance.
6.8 Surface Films — Have a deleterious effect on the
capacitance measurement. Film formation should be
avoided by using an appropriate electrolyte, control of
the etching bias and time and by electrolyte circulation.
6.9 Surface Roughness — The surface of the sample
should be smooth and free of features that would affect
the action of the seal.
6.10 Light — Light can affect the capacitance
measurement, so the sample must be placed in a dark
environment during the measurement.
6.11 Sample Loading — The accuracy of the
measurement depends on the repeatability of the area. If
a compliant seal is involved, such as a plastic ring, the
sample should be held in contact with the seal by
applying a controlled and constant force. No relative
lateral movement should take place during the loading
process.
6.12 Ring Variability — The condition of the ring has a
large effect on the measurement. Good seals will have a
small to zero excess area and exhibit a well defined and
reproducible contact area.
7 Apparatus
7.1 Measurement of Etch Well and Ring Areas —
Measuring microscope with either an XY stage, with a
linear resolution of 0.01 mm or better, or a camera and
image processing system capable of measuring an area
of 0.1 cm
2
to better than 0.5%.
7.2 Measurement of Etch Well Flatness — (Optional)
A means such as a stylus profiler for checking the
flatness and roughness of the etch well. In most cases
visual inspection of the etch well, for a mirror like
appearance, is sufficient indication of low roughness.