semi合集-English.pdf - 第770页
SEMI E104-0303 © SEMI 2000, 2003 6 8.3 Humidity 8.3.1 The LPPD sh ould wor k correctly at an ambient humidity range RH as specified in ISO 14644-5. If the use of an LPPD at an exte nded ambient hum idity range is require…

SEMI E104-0303 © SEMI 2000, 2003 5
5.2 Space, Distances, and Installation
5.2.1 The use of an LPPD should not have a negative
impact on the process performance. The sensor location
in the equipment shall ensure an optimum use of the
sensor sensitivity in order to allow a measurement
which is representative of the main particle flow. The
influence of the process and the process equipment on
the LPPD shall be minimized (thermal background
radiation, vibrations etc.)
5.2.2 The standard does not specify where an LPPD
should be located in original equipment, or where space
should be allowed for retro-fitting. It is recognized that
the location is dependent on factors including transport
behavior of particles, thermal background radiation etc.
The designers of semiconductor process equipment
should resort to the experience of application engineers
of the LPPD suppliers and the end users. However, in
either case designers shall take into account access
requirements for maintenance and re-calibration. In the
case of retro-fitting, the design should facilitate sensor
installation rather than hinder it.
6 Electrical Interfaces
6.1 Sensor/Controller Communication
6.1.1 Depending on the choice of the LPPD supplier
and the semiconductor equipment manufacturer, one of
the following interfaces for data exchange between the
LPPD and process equipment controller should be used.
6.1.2 Sensor/Actuator Network (SAN)
6.1.2.1 The electrical interfaces of the Sensor/ Actuator
Network are described in the suite of SEMI E54
standards.
6.1.3 Serial Communication
6.1.3.1 RS 232 C (V.24/V.28)
6.1.3.2 RS 485
6.1.4 Others
6.1.4.1 IEEE 488 or IEC 60625
6.2 Others
6.2.1 Sensor calibration output (direct analog output)
6.2.1.1 The analog output for access to the sensor
signal used for calibration purpose should be supported.
7 Communication Interfaces
7.1 Sensor Bus Communication
7.1.1 It is recommended to use a Sensor/Actuator
Network (SAN) for the intra-tool communication
between the LPPD controller and the controller of the
process equipment. This communication is based on a
suite of SEMI standards including a network
communication standard, several common device
models (DeviceNet, SDS, Lonworks, ProfiBus etc.),
and the specific device model for the sensor. The
Network is described by the following suite of SEMI
standards: SEMI E54. The specific device model for
ISPM is described by SEMI E54.10.
7.2 SEMI Equipment Communication (SECS)
7.2.1 If LPPD controllers provide the collected particle
data to the process equipment controller, factory
automation system, or SPC system, the SECS tool-to-
host communication could be used. SECS is described
in the following standards:
• SEMI E5
• SEMI E4
• SEMI E37
7.3 Attribute Definitions
7.3.1 SEMI E54.10 addresses the minimum attributes,
services and behavior an ISPM-device shall support. If
any attributes and services are used by communication
of the ISPM and the equipment controller via IEEE 488
(IEC 60625) or serial communication, they should be
concurring.
8 Operating Conditions
8.1 To specify the operating conditions of an LPPD
(temperature, humidity, electromagnetic compatibility,
vibrations etc.) refer to the following standards:
• IEC 60654-1
• IEC 60654-2
• IEC 60654-3
• IEC 60654-4
• IEC 60801-1
8.2 Temperature
8.2.1 The LPPD should work correctly at an ambient
temperature range T
A
as specified in ISO 14644-5. If
the use of an LPPD at an extended ambient temperature
range is required, the specific LPPD should comply
with these conditions.
8.2.2 The temperature T
G
inside the pump line or
inside the process equipment depends on the actual
application. The application engineers shall check the
use of LPPDs under these specific conditions. The
temperature T
G
is measured at the flange which the
LPPD is mounted on.

SEMI E104-0303 © SEMI 2000, 2003 6
8.3 Humidity
8.3.1 The LPPD should work correctly at an ambient
humidity range RH as specified in ISO 14644-5. If the
use of an LPPD at an extended ambient humidity range
is required, the specific LPPD should comply with
these conditions.
8.4 Electromagnetic Compatibility
8.4.1 The equipment should comply with SEMI E33.
8.4.2 Sensors will be incorporated into equipment
either as original equipment or retro-fitted. In either
case, the sensors or the equipment of which they are a
part should comply with the current regulations
covering EMC in the country or region where the
equipment or sensor is used.
8.5 Vibrations
8.5.1 Process equipment designers are advised to
consider the impact of vibration on the performance of
the sensor while it is collecting data. Therefore, they
should minimize vibrations. The designers of the LPPD
sensors are also advised to consider the impact of
vibrations on the equipment at a time the sensor is not
collecting any data. It might be possible that these
vibrations are stronger than those occurring while the
sensor is collecting data.
9 Reference Calibration Procedure
9.1 The response of real contamination particles,
typically with refractive indices and shapes different
from calibration particles, will differ slightly from the
results obtained by the procedures in this document. It
is known that LPPDs with different optical design may
not produce the same data from identical aerosol
samples. This may happen even with similar LPPDs if
calibration differences have occurred. Therefore, before
the first use the sensor should be calibrated by the
manufacturer. This calibration should be compliant
with or should be reviewed with the following reference
calibration equipment and procedure. This reference
calibration allows the characterization of the
performance of the LPPD under test. The LPPD should
be recalibrated at regular intervals and also in case of
unusual measurement readings to ensure correct results.
9.1.1 The parameters calibrated for LPPDs with sizing
capability are:
• Sizing calibration,
• Resolution,
• Zero counting,
• Counting efficiency, and
• Sensitivity.
9.1.2 The parameters calibrated for LPPDs with non-
sizing capability are:
• Zero counting, and
• Counting efficiency.
9.1.3 Field calibration may not necessarily require the
calibration of all parameters performed by calibration at
the sensor manufacturers site.
9.1.4 Due to the fact that different LPPDs might be
working with different detection areas A
L
, the number
of counts should be printed out in counts per mm
2
detection area. Therefore, a comparison of the
measurement results of different LPPDs is possible.
The size of the detection area A
L
shall be reported in
the calibration report form. Any changes of calibration
parameters or of the calibration setup shall be reported
in the calibration report form. A copy of the calibration
report form shall be delivered with the sensor.
9.2 Apparatus
9.2.1 Materials
9.2.1.1 Particles
9.2.1.1.1 Calibration particles are polymer spheres
composed of polystyrene or a similar polymer, having a
refractive index of 1.58-1.61+ 0i (absorption coefficient
α = 0), a sizing accuracy of at least 95%, and a size
distribution in which the coefficient of variation is 5%
or less. They should be traceable to a nationally or
internationally recognized standard (e.g. NIST
9
).
9.2.1.1.2 The calibration particles are normally
supplied in concentrations too high to be used directly
in aerosol generators. The particles should be dispersed
and diluted in either deionized, distilled water in
accordance with ASTM D1193, Type 1, or Isopropanol.
The diluent should be cleaned using a filter with a pore
size no more than 10% of the size of the particles being
used. The solution should be stored in a clean container.
For generation and dilution of the suspension see
Appendix 3. After generation the particles should be
neutralized to avoid surface charge. For all tests
described in this document, the concentration should be
no more than 25% of the maximum recommended
concentration limit specified by the manufacturer. For
calibration a suitable set of particle sizes shall be used.
This set should contain at least 5 sizes that cover the
LDL size to at least 80% of the specified maximum size
measurement capability of the LPPD.
NOTE 4: Most of the ISPM sensors are based on light
scattering. The intensity of the scattered light detected by a
photodetector depends on intensity, polarization state, and
9 National Institute of Standards and Technology, 100 Bureau Drive,
Gaithersburg, MD 20899-0001, USA

SEMI E104-0303 © SEMI 2000, 2003 7
wavelength of the incident light beam, diameter, shape, and
refractive index of the particle and the suspension fluid, as
well as on the geometrical layout of the collection optics and
detector. In the particle size range near the wavelength (0.1λ
< d
P
< 10λ), large oscillations can be seen in the intensity
curve of the light scattered by spherical particles as a function
of all these parameters. This phenomenon should be taken
into account when selecting a suitable set of particle sizes for
calibration. The used calibration particles should be within a
monotonic response range of the LPPD response curve.
9.2.1.2 Zero Gas
9.2.1.2.1 Clean air or nitrogen filtered with a ULPA
filter.
NOTE 5: The calibration will be executed under atmospheric
pressure and zero gas. In semiconductor manufacturing,
pressure and process gases will differ from the calibration
conditions. This will affect the refractive index ratio of the
particles and the process gas and consequently the scattering
from the particles.
9.2.1.3 Surfaces
9.2.1.3.1 The materials of pump lines and other
components should be conductive to minimize
electrostatic interaction with the particles.
9.2.2 Instrumentation (see Figure 1) — Some LPPDs
may require specialized equipment not generally
available. Please contact the LPPD manufacturer.
9.2.2.1 Fan System
9.2.2.1.1 The fan or pump system should be adjustable
to transport the aerosol and the zero gas within the
stipulated velocity range v
L, min
... v
L,max
, specified by
the manufacturer of the LPPD.
9.2.2.2 Filter System
9.2.2.2.1 The ULPA filter system is used for gener-
ation of zero gas. The filter system should be capable of
removing particles at the minimum size detectable by
the LPPD or the reference particle counter.
9.2.2.3 Aerosol Generator
9.2.2.3.1 An atomizer converts the monodisperse
particle suspension to an aerosol by using compressed
zero gas for generation and transportation of the
particles. The aerosol generator should generate
monodisperse particles as defined in Section 9.2.1.1 in
constant and reproducible concentration C
G
under
constant and reproducible volume flow rate V´
G
. The
variation in particle concentration shall be no more than
10% as measured by the reference particle counter over
a time period of 10 times or more of the sample
measurement time. The generation should comply with
the German VDI-Richtlinie 3491, or an equivalent
standard in other countries.
9.2.2.4 Aerosol dryer
9.2.2.4.1 The monodisperse polymer spheres in the test
aerosol shall be thoroughly dry to avoid that the
particles have a water layer which would increase their
size. A diffusion dryer, another appropriate instrument,
or adequately dry dilution air should be used to dry the
particles. The diffusion dryer uses silica gel desiccant to
remove the moisture. The desiccant shall either be new
or freshly regenerated. The design flow rate of the
aerosol dryer shall at least match the output flow rate of
the aerosol generator.
NOTE 6: Some diffusion dryers may precipitate polymer
spheres and add other particles when the aerosol gets in direct
contact with the silica gel.
9.2.2.5 Neutralizer
9.2.2.5.1 An aerosol neutralizer should be connected in
line with the dryer to reduce electrostatic charges on the
dry polymer spheres and to avoid electrostatic
interaction with each other or the line wall. The design
flow rate of the aerosol neutralizer shall at least match
the output flow rate of the aerosol generator.
NOTE 7: Some electrostatic neutralizers may produce a large
number of ultra-fine particles which will combine with the
calibration aerosol.
9.2.2.6 Aerosol Size Separator
9.2.2.6.1 In the case of calibration with ultra-fine
particles, a system should be used to separate single
polymer spheres from the residual particles resulting
from vaporization of solutions and aggregate particles
consisting of several spheres. The size separation could
be achieved with an electrostatic classifier or a
diffusion battery.
9.2.2.7 Aerosol Dilution
9.2.2.7.1 If the particle concentration behind the
aerosol generator is too high, the particle flow shall be
diluted to achieve the required concentration and to
avoid coincidence errors. The dilution should comply
with the German VDI-Richtlinie 3491, or an equivalent
standard in other countries.
9.2.2.8 Aerosol Line System
9.2.2.8.1 The system consisting of
• the aerosol generator,
• the aerosol dryer,
• the neutralizer,
• the particle size separator,
• the dilution stage, and
• tubing connecting the devices with each other and
the filtered, dried and compressed zero gas.