semi合集-English.pdf - 第4071页
SEMI F56-0600 © SEMI 2000 1 SEMI F56-0600 TEST METHOD FOR DETERMINING STEA DY-ST ATE SUPPLY VOLTAGE EFFECTS FOR M A SS FLOW CONTROLLE RS This test method was tec hnically approved by the G lobal Facilitie s Committee and…

SEMI F55-0600 © SEMI 2000 8
MFC Corrision Test
0.71
0.72
0.73
0.74
0.75
0.76
0.77
0.78
0.79
0.8
0.81
0.82
0.83
0.84
0.85
0.86
0.87
0.88
0.89
0 100 200 300 400 500 600 700 800 900 1000 1100 1200 1300 1400 1500 1600
Test Cycles Completed
Full Scale Percent
DUT Indicated Flow
DUT Setpoint
Actual Flow
Figure A1-1
Test Results Graph
NOTICE: SEMI makes no warranties or representations as to the suitability of the standard set forth herein for any
particular application. The determination of the suitability of the standard is solely the responsibility of the user.
Users are cautioned to refer to manufacturer’s instructions, product labels, product data sheets, and other relevant
literature respecting any materials mentioned herein. These standards are subject to change without notice.
The user’s attention is called to the possibility that compliance with this standard may require use of copyrighted
material or of an invention covered by patent rights. By publication of this standard, SEMI takes no position
respecting the validity of any patent rights or copyrights asserted in connection with any item mentioned in this
standard. Users of this standard are expressly advised that determination of any such patent rights or copyrights, and
the risk of infringement of such rights, are entirely their own responsibility.
Copyright by SEMI® (Semiconductor Equipment and Materials
International), 3081 Zanker Road, San Jose, CA 95134. Reproduction o
f
the contents in whole or in part is forbidden without express written
consent of SEMI.

SEMI F56-0600 © SEMI 20001
SEMI F56-0600
TEST METHOD FOR DETERMINING STEADY-STATE SUPPLY
VOLTAGE EFFECTS FOR MASS FLOW CONTROLLERS
This test method was technically approved by the Global Facilities Committee and is the direct responsibility
of the North American Facilities Committee. Current edition approved by the North American Regional
Standards Committee on April 10, 2000. Initially available on www.semi.org May 2000; to be published
June 2000.
1 Purpose
1.1 The purpose of this document is to define a
method for characterizing mass flow controllers
(MFCs) being considered for installation into a high-
purity gas distribution system. This method will
quantify the steady-state supply voltage effects on the
MFC's ability to accurately deliver set point flow
values.
2 Scope
2.1 This procedure applies to thermal mass flow
controllers. It is intended to measure the delivered
mass flow rate variation as a function of deviation from
the reference steady-state supply voltage. The test
method is designed for DC-powered MFCs. The
supply voltage effects include voltage depression and
over-voltage variations in the DC supply.
2.2 This standard does not purport to address safety
issues, if any, associated with its use. It is the
responsibility of the users of this standard to establish
appropriate safety and health practices and determine
the applicability of regulatory limitations prior to use.
3 Limitations
3.1 This test method is not designed for AC-powered
MFCs. This test method addresses steady-state effects
and does not address any effects caused by transient
power supply behavior.
4 Referenced Standards
4.1 None.
5 Terminology
5.1 Acronyms and Abbreviations
5.1.1 DUT — device under test
5.1.2 MFC — mass flow controller
5.1.3 NIST — National Institute of Standards and
Technology
5.1.4 psia — pounds per square inch absolute
5.1.5 psig — pounds per square inch gauge
5.2 Definitions
5.2.1 actual flow — flow value measured by the flow
standard.
5.2.2 indicated flow — flow value derived from the
MFC.
5.2.3 Reference voltage(s) — manufacturer's recom-
mended power supply voltage(s).
5.2.4 stable — the state a signal level obtains when its
magnitude varies by less than or equal to ±2.0% of full
scale over a one-minute period.
6 Summary of Method
6.1 This test method consists of varying the steady-
state supply voltage to an MFC in ±1% (of rated
supply) increments from the reference voltage span.
Both the positive and negative supply voltages will be
varied independently and together in 1% increments.
The MFC flow output is monitored at 0% and 100% of
its operating range (see Table 1). At each incremental
change as well as at the reference supply voltage, the
MFC flow and flow standard outputs are recorded at
0% and 100% of its operating range. The test shall end
when a ±3% of full-scale output flow change is noted
between the flow at the reference voltage and at an
incremental change point. A flow chart outlines the
procedure (see Figure 1).
7 Interference’s
7.1 Because of fluctuations in ambient temperature or
a changing load on the power supply, the reference
steady-state supply voltages may change. These
changes in supply voltages may adversely effect the
MFC's ability to deliver setpoint flow. The magnitude
of this effect can be measured by this test method.
7.2 The user of an MFC can use the data generated by
this method to evaluate the impact of steady-state
supply voltage variations on the MFC's ability to
deliver setpoint flow values. Knowing the magnitude
of this effect and the power level variations, allows the
user to decide on what measures to take to reduce this
effect, if necessary.

SEMI F56-0600 © SEMI 2000 2
8 Apparatus
8.1 The equipment and instrumentation required to
complete this test method are shown in Figure 2.
8.2 DC Power Supply, to provide the required voltage
to the MFC under test, with the capability to vary its
output over the range of ±30 VDC at 500 mA, and with
ripple less than 0.1% rms.
8.3 DC Power Supply Monitor/Recorder, placed in
parallel with the DC power supply, and capable of
measuring the DC voltage to within ±5 mV over the
entire operating range of the supply.
8.4 Flow Output Monitor, connected to the MFC
output and signal common/ground points. The
monitor/recorder shall be capable of measuring over a
range of 0–10 VDC to within ±5 mV.
8.5 Flow Standard, installed downstream and in series
with the flow through the MFC. The flow standard
shall be capable of measuring flow changes within
±0.3% of full scale.
8.6 Current Meter, connected in series in the power
common line of the DC power supply, capable of
measuring over a range of 0–200 mA to within ±1 mA.
9 Reagents and Materials
9.1 Test Gas, nitrogen
10 Safety Precautions
10.1 The user must know the respective
instrumentation, practice proper handling of test
components, and understand good laboratory practices.
10.2 The user should not exceed the ratings (such as
pressure, temperature, flow, and voltage) of the
components.
11 Preparation of Apparatus
11.1 Setup and Schematic — See Figures 2 and 3.
11.2 The test gas source and delivery system must be
capable of satisfying the test volume flow rate at a
constant pressure, +0.1 psia.
11.3 The test gas source and delivery system must be
capable of delivering a gas at ambient temperature
±2°C for the duration of each analysis. The ambient
temperature shall be held to 22 ±1°C.
12 Calibration and Standardization
12.1 All instrumentation shall be calibrated with NIST
traceable standards and shall be under current
calibration.
13 Procedure
13.1 Install the MFC into the test setup per
manufacturer's recommendations.
13.2 Apply power to all devices sho wn in Figure 2 per
manufacturer's specifications. Allow the devices to
warm up for the duration specified by the equipment
manufacturer.
13.3 Set the DC power supply to the manufacturers
recommended reference voltage. Verify the voltage
magnitude using the power supply monitor.
13.4 Purge the system with nitrogen for a length of
time equal to ten times the amount of time it takes to
replace the system volume with the test MFC at its full-
scale rated flow rate.
13.5 Close the inlet shut-off valve and then the outlet
shut-off valve located adjacent to the MFC (see Figure
3). Adjust the MFC setpoint to zero flow. Follow the
manufacturer's recommendations for adjusting the MFC
zero. If the MFC has an auto-zero function, leave it
active and note this fact on the data sheet in Table 1.
Wait for the signals to become stable. Record on the
data sheet (see Table 1) three separate readings of the
MFC flow output, the power supply voltages, and the
current of the power supply common.
13.6 Vary the positive DC power supply voltage in
+1% steps from the reference value while holding the
negative supply at its reference value.
NOTE 1: If reference voltage is +15 V, the first step is a
change in voltage of +0.15 V.
13.7 After the monitored signals become stable, record
the MFC zero value, the current value, and the power
supply voltage magnitudes. Record the measurements
at each step-change of power level on the data sheet.
Continue to increase the positive supply voltage span
until the MFC zero point changes by ±3% of full scale
flow, the MFC zero output point remains unstable for
five minutes, or the voltage level changes to +115% of
original value.
13.8 Return the power supply level to the original
reference value and repeat Section 13.5.
13.9 Repeat the procedure described in Section 13.6,
with the following exception: Decrease the power
supply span in 1% decrements until either the MFC
zero changes by ±3% of full scale, the zero output point
remains unstable for five minutes, or 85% of the
original value is reached.
13.10 Return the power supply level to the original
reference value, repeat Section 13.5, and go to Section
13.10.