semi合集-English.pdf - 第6965页
SEMI S9-1101 © SE MI 1995, 2001 2 4.2 IEC Standard 1 IEC 61010- 1 — Safety Requirements for Electrical Equipment for Measurement, Control, an d Laboratory Use, Part 1: General Requirements NOTE 1: Unless otherwise indica…

SEMI S9-1101 © SEMI 1995, 20011
SEMI S9-1101
SAFETY GUIDELINE FOR ELECTRICAL DESIGN VERIFICATION
TESTS FOR SEMICONDUCTOR MANUFACTURING EQUIPMENT
This guideline was technically approved by the Global Environmental Health and Safety Committee and is
the direct responsibility of the North American Environmental Health and Safety Committee. Current edition
approved by the North American Regional Standards Committee on August 27, 2001. Initially available at
www.semi.org September 2001; to be published November 2001. Originally published in 1995.
NOTICE: This document as balloted is intended to
replace SEMI S9-95 in its entirety.
NOTICE: Paragraphs entitled “NOTE” are not an
official part of this document and are not intended to
modify or supercede the official guideline. The task
force has supplied them to clarify and to enhance usage
of the guideline by equipment designers.
1 Purpose
1.1 The purpose of this document is to provide
electrical design verification tests, test methods, and
acceptance criteria for semiconductor manufacturing
equipment. Some of these tests are used as part of the
electrical safety evaluation in SEMI S2.
2 Scope
2.1 This safety guideline should be applied to one or
more representative samples of the equipment (or parts
of the equipment) used for the manufacturing,
measurement, assembly, and testing of semiconductor
products.
2.2 The following tests are discussed in this document:
• Leakage Current Test (Section 9.1)
• Grounding Continuity Test (Section 9.2)
• Starting Current Test (Section 9.3)
• Input Test (Section 9.4)
• Dielectric Test (Section 9.5)
• Strain Relief Test (Section 9.6)
• Transformer Output Short Circuit Test (Section
9.7)
• Power Supply Output Short Circuit Test (Section
9.8)
• Safety Circuit Function Test (Section 9.9)
• Safety Circuits Conductor Disconnection Test
(Section 9.10)
• Capacitor Stored Energy Discharge Test (Section
9.11)
• Temperature Test (Section 9.12)
2.3 This safety guideline does not purport to address all
of the safety issues associated with its use. It is the
responsibility of the user of this safety guideline to
establish appropriate safety and health practices and
determine the applicability of regulatory limitations
prior to use.
3 Limitations
3.1 This document is not intended to be a
comprehensive compilation of electrical tests specified
in product safety standards. This document is also not
intended to replace any test methods described in any
appropriate national or international product safety
standards.
3.2 It is not the intent of this guideline to require
repetition of tests where the equipment has been
certified or tested to other relevant electrical product
safety standards. Any applicable test from this
guideline that has been previously completed under an
applicable product standard and which satisfies the
intent of that test should be accepted, even if there are
differences in test methods.
3.3 Engineering analysis, when based on sound
engineering principles, may serve as an alternative to
conducting a test.
3.4 If it is evident from the design and construction of
the equipment that a particular test is not safety-
relevant, the test need not be performed.
3.5 This document is not intended to address
production line or routine testing.
4 Referenced Standards
4.1 SEMI Standards
SEMI S2 — Environmental, Health, and Safety
Guideline for Semiconductor Manufacturing Equipment

SEMI S9-1101 © SEMI 1995, 2001 2
4.2 IEC Standard
1
IEC 61010-1 — Safety Requirements for Electrical
Equipment for Measurement, Control, and Laboratory
Use, Part 1: General Requirements
NOTE 1: Unless otherwise indicated, all documents cited
shall be the latest published versions.
5 Terminology
5.1 1500 ohm impedance network — a network
consisting of a 1500 ohm resistor in parallel with a 0.15
µF capacitor.
5.2 accessible — capable of being contacted by an IEC
accessibility probe (IEC standard test finger, as
described in IEC 61010).
NOTE 2: The test finger is also known, in various IEC
documents, as the “jointed test finger” and “jointed standard
test finger”.
5.3 full load current — current when the equipment is
operated at the maximum manufacturer's specified
operating conditions including all motors and heaters
designed to operate simultaneously.
5.4 hazardous energy — energy of 20 J (Joules) or
more, or an available power level of 240 V*A or more.
5.5 least favorable condition — the condition which is
most likely to result in a test failure.
5.6 permanently connected equipment — Equipment
that is intended to be electrically connected to a supply
by means of connection which can be detached only by
the use of tools.
5.7 primary circuit — an internal circuit which is
directly connected to the external supply mains or other
equivalent source (such as a motor-generator set) which
supplies the electric power. It includes the primary
windings of transformers, motors, other load devices
and the means of connection to the supply mains.
5.8 safe condition a condition in which all relevant
hazardous energy sources are removed or suitably
contained and all relevant hazardous production
materials are removed or contained, unless this results
in additional hazardous conditions.
5.9 PE Terminal — a terminal which is bonded to
conductive parts of a piece of equipment for safety
purposes and is intended to be connected to an external
protective earthing system.
1 International Electrotechnical Commission, 1, rue de Varembé,
Case Postale 131, CH-1211 Geneva 20, Switzerland. Website:
www.iec.ch
5.10 standby condition — condition in which
equipment is energized and in its idle state.
5.11 tool — an external device used to aid a person in
performing a mechanical function. As used in this
document, tool includes devices such as keys.
5.12 V*A — Volt-ampere.
6 Safety Precautions
6.1 The tests outlined in this document are to be
performed by trained and qualified personnel who have
knowledge of the techniques and the test apparatus
described herein.
7 Calibration and Standardization
7.1 All test equipment should be calibrated and
traceable to a calibration standards organization (e.g.,
National Institute of Standards and Technology (NIST)
in the United States or the National Metrology Institute
in Japan).
7.2 The calibration interval for test equipment should
not exceed one year.
8 Test Conditions
8.1 Except where noted otherwise, the equipment
should be tested under the least favorable conditions
within the manufacturer's operating specifications.
These conditions include:
• supply potential
• supply frequency
• position of movable parts
• operating mode (e.g. full temperature conditions,
motors in operation)
• adjustment of thermostats, regulating devices, or
similar controls in operator-accessible areas
8.2 Test Supply Potential — To determine the least
favorable supply potential for a test, consider:
• multiple-nominal rated potentials (e.g., 120/240 V)
• extremes of nominal rated potential ranges (e.g.,
208–240 V)
8.2.1 Consideration of the tolerance on a nominal rated
potential (e.g., 120 ± 5%) is not necessary.
NOTE 3: Some standards (e.g., IEC 61010-1 and IEC 60950)
may specify 90% and 110% of any rated supply voltage.
8.3 Test Supply Frequency — To determine the least
favorable supply frequency for a test, consider the
nominal frequencies as specified (e.g., 50 Hz, 60 Hz, or
50/60 Hz).

SEMI S9-1101 © SEMI 1995, 20013
NOTE 4: Consideration of the tolerance on a nominal rated
frequency (e.g., 50 ± 0.5 Hz) is not usually necessary.
8.4 As an alternative to carrying out tests on the
complete equipment, tests may be conducted on
circuits, components and sub-assemblies independent of
the equipment, provided that the results of the tests
would be representative of those performed as part of
the assembled equipment.
EXCEPTION: The leakage current and grounding
continuity tests identified in Sections 9.1 and 9.2 should
be completed only on fully assembled equipment.
9 Electrical Tests
9.1 Leakage Current Test for Cord-and-Plug
Equipment
9.1.1 Test Equipment A 1500 ohm impedance
network and a true RMS voltmeter with an accuracy of
1.0%. The impedance network can be a separate
assembly or incorporated within a leakage current
measuring instrument.
9.1.2 Procedure For equipment connected to the
facility branch circuit with a cord-and-plug (plug/socket
combination), ensure that the equipment is isolated
(e.g., by placing the equipment on a wooden or other
isolating surface). Connect the equipment to its rated
source of supply with the equipment grounding (PE)
conductor disconnected and operate it at the least
favorable conditions specified by the manufacturer.
Connect the 1500 ohm impedance network between
each accessible metal part and the supply equipment
grounding (PE) conductor. In determining accessibility
of live parts, remove all doors, panels, etc. that are to be
removed by the operator during normal operation.
Using a true RMS voltmeter, measure the voltage drop
across the impedance network. Calculate the leakage
current using the formula:
I
leaka
g
e
=
Voltage
measured
1500 ohms
9.1.3 Acceptable Results The maximum calculated
leakage current does not exceed 3.5 mA.
9.2 Grounding Continuity Test
9.2.1 Test Equipment Low range ohmmeter with a
range to measure 0.10 ohm with an accuracy of 1.0%.
9.2.2 Procedure
Disconnect the equipment from the
supply. For equipment installed with fixed wiring
methods, disconnect the supply equipment grounding
conductor (protective earthing conductor) from the
main equipment grounding terminal (PE Terminal).
Measure the resistance between the power supply
equipment grounding terminal (PE Terminal) and each
accessible metal part (handle, monitor, doors, etc.) on
the equipment using a low-range ohm-meter. Upon test
completion, reconnect the supply equipment grounding
conductor (protective earthing conductor).
EXCEPTION: Grounding Continuity Test is not
required to be measured where accessible metal
surfaces are not likely to become energized in a single
fault condition.
NOTE 5: Some standards (e.g., IEC 60204-1, IEC 61010-1)
may specify this test to be performed using a current injection
method.
9.2.3 Acceptable Results The resistance between the
grounding conductor terminal and each accessible part
shall not exceed 0.1 ohm.
9.3 Starting Current Test
9.3.1 Test Equipment None.
9.3.2 Procedure — Start the equipment in accordance
with manufacturer's instructions three times from a
completely stopped condition. Ensure that the time
interval between successive starts is sufficient to allow
the equipment return to ambient conditions.
9.3.3 Acceptable Results None of the equipment's
overload or overcurrent protections activates during this
test.
NOTE 6: It is recommended that the peak inrush starting
current be measured using an appropriate current measuring
device and recorded in the test report.
9.4 Input Test
9.4.1 Test Equipment True RMS current measuring
equipment, with accuracy of 3.0%.
9.4.2 Procedure — Measure the input current to the
equipment under the maximum normal operating load
conditions (i.e., with all motors, heaters, etc. running at
manufacturer’s specified maximum loading conditions).
9.4.3 Acceptable Results The measured current does
not exceed 110% of the rated full load current value
specified on the equipment nameplate.
9.5 Dielectric Test
9.5.1 Test Equipment Timer with accuracy of ± 5
seconds. Dielectric Withstand Tester with means of
indicating test potential, as well as an audible or visual
indicator of electrical breakdown, or an automatic-
reject feature for any unacceptable unit. In an
alternating current test, the test equipment should
include a transformer having sinusoidal output. This
transformer should have a rating of 500 VA or greater
unless it is provided with a voltmeter that directly
measures the applied output potential.