SEMI E104-0303 © SEMI 2000, 2003 10
9.3.3.5 Fixed velocity at the LPPD: v
L
= 1 m/s
(independent of line diameter at the LPPD).
9.3.3.5.1 Some LPPDs, e.g. a scanning LPPD for the
use inside a process chamber, require a different
velocity than 1 m/s for calibration. In such cases the
used aerosol velocity at the location of the LPPD shall
be reported in the calibration report form.
9.3.4 Initial Setup
9.3.4.1 Connect the aerosol generator to a supply of
filtered, dried, and compressed zero gas (see Figure 1).
Connect a dryer, neutralizer, and size separator (if
necessary) to the aerosol outlet of the particle generator.
If a dilution of the aerosol before the injection in the
calibration line is necessary, install a dilution stage
behind the size separator.
9.3.4.2 Connect the aerosol line to the calibration line.
Mount the LPPD under test behind the injection and
mixing device. Install the sample acquisition device and
the velocity or flow meter into the calibration line.
9.3.4.3 Connect a PHA to the sensor calibration output
if an appropriate built-in PHA is not available.
9.3.4.4 Before starting the following procedures, make
sure all instruments are turned on and allow a certain
time for warm-up and temperature stabilization.
9.3.4.5 Purge the aerosol line and the calibration line
with filtered and dried air for a minimum of 30 minutes.
9.3.4.6 Check the quality of the zero gas and the
condition of the calibration line using the RPC. Collect
counts for three one-minute periods and determine the
average number of particles reported in counts per
minute. The volume flow of the zero gas shall be
enough for correct operation of the RPC. Record the
cumulative particle count reported for the zero gas for
particles equal to and larger than the LDL of the LPPD
under test.
9.3.4.7 Clean the aerosol generator liquid reservoir and
fill it with clean diluent (particle-free deionized or
destilled water or Isopropanol). Run the aerosol
generator and adjust the pressure and flow rates as
recommended by the aerosol generator manufacturer.
Check the condition of the aerosol line using the RPC.
Collect counts for three one-minute periods and
determine the average number of particles reported per
liter of air. If the number exceeds 1 particle per liter air,
the diluent shall be refiltered before further use. The
average number of particles (counts per minute) should
be reported in the calibration report form.
9.3.4.8 Adjust the volume flow through the calibration
line for the stipulated aerosol velocity at the LPPD
under test.
9.3.4.9 Allow a certain time for stabilization of the
adjusted calibration parameters before any test is
performed.
9.3.5 Sizing LPPD
NOTE 11: For more information and background see
Appendix 2.
9.3.5.1 Size Calibration
NOTE 12: To define the calibration settings either the mode
or median of the voltage pulse distribution could be used. The
use of the modal procedure is common. The calibration report
form should identify the used method.
9.3.5.1.1 By running different monodisperse aerosols
(with v
L
and V´
L
= V´
0
+
G
), size calibration is
performed. Begin with the largest particles for which a
calibration value is required. Select the particle size
range for which calibration is required. Be sure that the
particle concentration is low enough that predominantly
individual pulses are generated by the LPPD.
Accumulate enough data to avoid statistical errors. The
procedure is repeated for each particle size of interest.
Determine the average LPPD pulse voltage amplitude
with the PHA.
9.3.5.1.2 The mode of the voltage pulse distribution is
determined by the highest point in the distribution. If
the distribution is interfered with noise near the peak
and the peak is not well defined, the data should be
averaged and the peak of the averaged distribution
should be used defining the mode.
9.3.5.1.3 The median voltage is determined by defining
a “Region of Interest” (ROI) which includes the peak
representing the voltage pulses from single polymer
spheres. The PHA determines the total number of
pulses under this peak. The median voltage is that
voltage which devises the number of pulses in the
region such that half are greater and half are less than
the median.
NOTE 13: If the procedure is carried out for the smallest size
of which the LPPD is specified, record the noise level of the
LPPD. The average voltage for this size should be at least
10% higher than stated noise level voltage at which one noise
pulse occurs in one minute.
9.3.5.1.4 The particle size and the average voltage
should be reported in the calibration report form.
9.3.5.1.5 For calibration checks, a reference particle
counter (RPC) with a good sizing capability could be
used. The counting efficiency of the RPC is defined as
100% for the particle sizes used in test. The counts of
the RPC and the LPPD under test in the corresponding
channel should be normalized to a standard detection
area (mm
2
). The ratio of these counts (in %), the
employed particle size and the threshold settings should
be recorded in the calibration report form.
SEMI E104-0303 © SEMI 2000, 2003 11
9.3.5.2 Zero Counting
NOTE 14: The intent of this procedure is not to adjust the
thresholds for the zero count rate but to verify whether the
LPPD is within its specification.
9.3.5.2.1 By running zero gas (with v
L
and V´
L
= V´
0
)
through the system the zero count rate, specified by the
sensor manufacturer, is checked. The reference particle
counter is used to check the quality of the zero gas. The
minimum acceptable sample time is 10 minutes. The
sampling time shall be long enough to provide adequate
sampling statistics. The zero count rate (average counts
per minute) should be recorded in the calibration report
form.
9.3.5.3 Counting Efficiency
NOTE 15: The actual particle concentration is determined
with a reference particle counter with a 100% counting
efficiency for the employed particle size. It is important that
there are no differences between the concentration within the
detection area of the LPPD and the concentration monitored
by the RPC.
9.3.5.3.1 By running a monodisperse aerosol (with v
L
and V´
L
= V´
0
+
G
), the maximum of detected
particles is counted. The ratio of the counts of the
LPPD standardized per mm
2
to the counts of the
reference particle counter per mm
2
detection area (in %)
for each particle size should be recorded in the
calibration report form.
9.3.5.4 Sensitivity
9.3.5.4.1 By running different monodisperse aerosols
(with v
L
and V´
L
= V´
0
+
G
) with particle sizes at the
expected sensitivity where the counting efficiency is
between 40% and 60%, the sensitivity is verified. The
counts of the LPPD per mm
2
detection area compared
to the counts of the reference particle counter per mm
2
detection area (in %) should be determined. The
determined particle size shall have a counting
efficiency between 40% and 60% and should be
recorded in the calibration report form.
9.3.5.5 Particle Size Resolution
NOTE 16: The particle size shall be at least twice the lower
counting limit to ensure measurement of the entire
distribution and accurate characterization of the calibration
response curve below the particle size of interest. The
resolution is specified by the coefficient of variation (in %)
obtained by dividing the portion of the standard deviation of
the distribution σ
LPPD
that is contributed by the LPPD by the
mean size of the distribution d
p
.
9.3.5.5.1 Record the size d
p
and the standard deviation
σ
d,p
of the used particles in the calibration report form.
By running a monodisperse aerosol (with v
L
and V´
L
=
0
+
G
), determine the standard deviation of the
observed particle distribution σ
p,LPPD
. To determine the
standard deviation of the pure sensor signal σ
V,LPPD
, a
pulse height analyzer can measure the voltage peaks at
the analog sensor calibration output. The portion of the
standard deviation of the distribution σ
LPPD
that is
contributed by the LPPD is calculated by the following
formula:
(
)()
2
,
2
, pdLPPDpLPPD
σσσ
=
9.3.5.5.2 The value of the standard deviation of the
observed distribution σ
p,LPPD
, the standard deviation of
the sensor signal σ
V,LPPD
, and the coefficient of
variation (in %) should be recorded in the calibration
report form.
9.3.6 Non-Sizing LPPD
NOTE 17: For more information and background see
Appendix 2.
9.3.6.1 Zero Counting
9.3.6.1.1 See Section 9.3.5.2.
9.3.6.2 Counting Efficiency
NOTE 18: The actual particle concentration is determined
with a reference particle counter with a 100% counting
efficiency for the employed particle size. It is important that
there are no differences between the concentration within the
detection area of the LPPD and the concentration monitored
by the RPC.
9.3.6.2.1 By running a monodisperse aerosol (using the
smallest particles size which can be counted with v
L
and V´
L
= V´
0
+
G
), the maximum of detected
particles is counted. The counts of the LPPD per mm
2
detection area compared to the counts of the reference
particle counter per mm
2
detection area (in %) and the
employed particle size should be recorded in the
calibration report form. These measurements should be
executed for the defined set of particle sizes including
one size no larger than 1.3 times the sensitivity limit as
specified by the manufacturer.
9.3.7 Interim Procedure
9.3.7.1 Between the settings of the separate thresholds
or the runs with different particle sizes, the system
should be purged with zero gas for a sufficient time.
The aerosol generator should be rinsed with clean
diluent for a sufficient time before adding new
suspension of particles with other size.
10 Calibration Report Form
10.1 The calibration report form shall contain the
following information:
SEMI E104-0303 © SEMI 2000, 2003 12
10.1.1 General part:
Date and time of calibration
Name of the operator
Manufacturer of the LPPD
Model and serial number of the LPPD to be
calibrated (if sensor and counter are separate,
model and serial number for each)
Size of the detection area (if detection area is
dependent on the particle size, all established
values)
Model, serial number, manufacturer, and date of
last calibration of the reference particle counter
Performance of the reference particle counter
(volume flow, sensitivity, zero count rate)
Environmental conditions: temperature T
A
, relative
humidity RH, pressure p
Quality of zero gas as measured with the reference
particle counter (counts per minute)
Specifications of the particles used for calibration
(manufacturer, certified size and tolerance,
standard deviation σ
d,p
, coefficient of variation,
refractive index, lot number)
Calibration parameters: particle concentration C
L
,
particle velocity v
L
at the location of the LPPD,
line diameter D
L,
opening area A
RC
of the probe
inlet for each particle size
Calibration procedure: sizing or non-sizing
10.1.2 For sizing LPPD:
Size calibration: value of the average voltage of the
different particle sizes, employed particle sizes,
mode or median method
Size calibration check: values of the thresholds,
employed particle sizes, ratio of the normalized
counts in the corresponding channel (in %)
Zero counting: value of the zero count rate
(average counts per minute), sampling time
Particle size resolution: value of coefficient of
variation, value of standard deviation of the
observed distribution, mean size of the distribution,
value of standard deviation of the sensor signal,
employed particle sizes
Counting efficiency: value of the counting
efficiency (in %), employed particle sizes
10.1.3 For non-sizing LPPD:
Zero counting: value of the zero count rate
(average counts per minute), sampling time
Counting efficiency: value of the counting
efficiency (in %), employed particle sizes
10.2 Report significant variations from data reported
from the previous calibration.
11 Related Documents
NOTE 19: These publications related to particle
measurement, vacuum ISPM, and calibration are just
informative to improve understanding of these standard.
Raasch, J.; Umhauer, H.: “Errors in Determination of
Particle Size Distributions Caused by Coincidence in
Optical Particle Counters” Particle Characterization 3,
1990, 424-427
Jaenicke, R.: “The Optical Particle Counter: Cross
Sensitivity and Coincidence” J. Aerosol Sci., 30(5),
1972, 95-111
Borden, P.: “Monitoring Vacuum Process Equipment:
In Situ Monitors – Design and Specification”
Microcontamination, 1991, 43-47
Raabe, O.G.: “The Generation of Aerosols of Fine
Particles” Fine Particles, Ed. Liu, B.; Academic Press,
New York, 1976, 57-110