semi合集-English.pdf - 第4698页

SEMI C3.39-0304 © SEMI 1989, 2004 4 Table 1 Calibration C oncentrations Added Volume of Working Standard (ml) Total Volume (ml) Resulting F - Concentration (M) 0.2 100.2 1.0 × 10 -6 0.2 100.4 2.0 × 10 -6 0.4 100.8 4.0 × …

100%1 / 7923
SEMI C3.39-0304 © SEMI 1989, 2004 3
5.4.4 Operating Procedure
5.4.4.1 Inject the calibration standard into the column
using a gas sampling valve. Record the retention times
and peak areas. The order of elution is oxygen + argon
(not separated) and nitrogen.
5.4.4.2 Inject the sample to be tested in the same
manner as the calibration standard. Record the retention
times and peak areas.
5.4.4.3 Repeat Section 5.4.4.1.
5.4.4.4 Compare the average peak area of the
calibration standard to that of the nitrogen trifluoride
sample being tested. Calculate the concentrations of
oxygen + argon and nitrogen, using the formula below.
The results may not exceed the specifications in Section
3 of this standard.
Sample Peak Area
Standard Peak Area
×
C
oncentrat
i
on
of Standard
=
C
oncentrat
i
on
of Sample
5.5 Water — This procedure is for the determination of
trace moisture (water) in nitrogen trifluoride using a
continuous flowing piezoelectric hygrometer. (See
Notes 3 and 4.)
5.5.1 Detection Limit — 0.04 ppm (vol/vol) or -95° C (-
139° F).
5.5.2 Flow Requirements — Set the sample pressure
and flow rate set in accordance with the instrument
manufacturer’s instructions.
5.5.3 Calibration Standards — Construct a calibration
curve which contains at least three points covering the
range of interest. Verify the standards employed by
another analytical method.
5.5.4 Operating Procedure
5.5.4.1 Obtain a continuous flow of sample gas from
the source using a clean and passivated 316 stainless
steel line which has been purged dry after exposure to
ambient moisture.
5.5.4.2 After prepurging with a dry gas, allow the
sample as to flow through the sampling system and the
piezoelectric hygrometer until a stable reading is
obtained. The result may not exceed the specification in
Section 2 of this standard.
5.6 Hydrolyzable Fluorides as HF This procedure is
for the determination of hydrolyzable fluorides in
nitrogen trifluoride using fluoride ion selective
electrode.
5.6.1 Detection Limit Detection limits should be
checked for any new implementation of a method.
Detection limits below 0.1 ppm have been determined
with this method. Detection limits can be improved by
increasing the volume of gas sampled.
5.6.2 Equipment
5.6.2.1 mV meter (0.1 mV scale)
5.6.2.2 Reference electrode (single junction type)
5.6.2.3 Fluoride ion selective electrode
5.6.2.4 Magnetic stir bars (PTFE coated)
5.6.2.5 Magnetic stirrer
5.6.2.6 Plastic beakers (See NOTE 5.)
5.6.2.7 PTFE bubblers
5.6.2.8 1000 ml and 100 ml plastic volumetric flasks
(see NOTE 5).
5.6.2.9 0.2 ml and 1 ml plastic volumetric pipettes (see
NOTE 5).
5.6.2.10 Flow controller or flowmeter (0–1000 sccm
NF
3
)
5.6.3 Reagents
5.6.3.1 Distilled or deionized water
5.6.3.2 5 N sodium hydroxide
5.6.3.3 0.2 N sodium hydroxide
5.6.3.4 Glacial acetic acid
5.6.3.5 Buffer Solution To 500 ml distilled or
deionized water in a 1000 ml volumetric flask, add 57
ml glacial acetic acid and 58 g of sodium chloride.
Adjust the pH to between 5.0–5.5 with 5 M sodium
hydroxide. Cool to room temperature. Dilute to one
liter with distilled or deionized water.
5.6.4 Calibration Standard — Sodium fluoride
standard (10
-3
M F
-
in water, freshly prepared)
5.6.5 Operating Procedure
5.6.5.1 Prepare working standard by adding 100 ml 10
-
3
M F
-
to 100 ml buffer solution.
5.6.5.2 Prepare a blank containing 50.0 ml 0.2 N
NaOH and 50.0 ml buffer solution in a plastic beaker.
5.6.5.3 While stirring blank gently, record mV reading
from the blank once reading is stable.
5.6.5.4 Successively add increments of working
standard to the blank to generate a calibration curve.
Record stable mV reading after each addition. Table 1
shows recommended increments and resultant
concentrations.
SEMI C3.39-0304 © SEMI 1989, 2004 4
Table 1 Calibration Concentrations
Added Volume of
Working Standard
(ml)
Total Volume
(ml)
Resulting F
-
Concentration (M)
0.2 100.2 1.0 × 10
-6
0.2 100.4 2.0 × 10
-6
0.4 100.8 4.0 × 10
-6
0.4 101.2 5.9 × 10
-6
0.8 102.0 9.8 × 10
-6
1.0 103.0 1.5 × 10
-5
2.0 105.0 2.4 × 10
-5
5.6.5.5 Put 50 ml 0.2 N NaOH into each of two
bubblers connected in series.
5.6.5.6 Establish a flow of < 1000 sccm of NF
3
through
the bubblers using a suitable flow controller or
flowmeter.
5.6.5.7 Sample approximately 15 liters of NF
3
. Record
flowrate and time of sampling to determine total
volume sampled (flowrate × time). A wet test meter
can also be used to measure total volume. The amount
of gas sample must be the volume at STP. If the
flowmeter or wet test meter is not reference to 0º and
760 torr, use the formula below to correct sample
volume.
K 273
273T
P
Torr 760
Liters Measured STPat Liters
+
××=
P: Pressure of sampled gas (mm Hg)
T: Temperature of sampled gas or reference
temperature of the flow controller or meter in ºC.
5.6.5.8 Transfer contents of each bubbler to individual
100 ml volumetric flasks and add 50 ml Buffer solution
to each. Then, if necessary, add deionized or distilled
water to bring the volume up to 100 ml.
5.6.5.9 Transfer contents to a plastic beaker.
5.6.5.10 While stirring, measure and record mV
readings for each sample.
5.6.5.11 Determine F
-
concentration in solution using
calibration curve generated in Section 5.6.5.4.
5.6.5.12 Calculate gas phase hydrolyzable fluoride
concentration using the equation below. Note that the
equation assumes the hydrolyzable fluoride is hydrogen
fluoride.
6
v
10
Vs
l/mole 22.4
0.11 C )(ppm HF Phase Gas ×××=
C: Measured hydrolyzable fluoride concentration (M)
determined in Section 5.6.5.11.
Vs: Volume of NF
3
sampled (liters at STP)
5.6.5.13 The HF concentration of the second bubbler
should be insignificant compared to the first bubbler. If
significant HF levels are found in the second bubbler,
resample with a lower flowrate.
5.7 Notes
NOTE 1: Pass the carrier gas through a trap (3 m (10 ft.) by
6.4 mm (1/4 in.) OD) containing 50% by volume molecular
sieve 5A and 50% by volume molecular sieve 13X submerged
in liquid nitrogen.
NOTE 2: The methanizer catalyst can be destroyed by
nitrogen trifluoride passing through to at high temperature.
After the carbon monoxide peak is detected, the bulk nitrogen
trifluoride is bypassed to vent by means of a four-port
switching valve installed downstream of the separating
column and before the methanizer (see Figure 1).
NOTE 3: The sampling system and hygrometer must be
designed to operate under the sample pressure, or the sample
pressure must be reduced, by a regulator with a diaphragm of
stainless steel or other suitable material, to accommodate the
pressure restrictions of the analytical hygrometer.
NOTE 4: A passivation procedure is described in the Metals
Handbook, 8th ed., Vol. 2, American Society for Metals,
Metals Park, OH.
NOTE 5: Polytetrafluoroethlyene, polymethylpentene and
polypropylene are suitable plastic materials.
SEMI C3.39-0304 © SEMI 1989, 2004 5
Figure 1
Configuration for the Analysis of CF
4
, CO
2
, N
2
O, SF
6
, and CO in NF
3
Figure 2
Configuration for the Analysis of O
2
/Ar and N
2
in NF
3