semi合集-English.pdf - 第3890页
SEMI F27-0997 © SEMI 1997, 2003 5 Regulators should be tested in the fully-open cond ition (i.e., with the regulator adjus ted for minim u m pressure drop). The quantity o f published data on regulator interaction with m…

SEMI F27-0997 © SEMI 1997, 2003 4
7.1.1 Initial Dry Down — Start the experiment with the
blank in place of the test specimen and a flow of dry
gas through the APIMS. The APIMS output should be
at equilibrium with the lowest moisture level of interest,
and in any case no higher than 500 ppt. The flow rate
through the test blank should be set according to the
Flow Table (see Table 1).
7.1.1.1 Switch the gas flow to pass primarily through
the bypass loop while maintaining a small flow through
the test blank. Remove the test blank completely from
the system. If a glove box or other such enclosure is
used, do not remove the test blank from the glove box.
Immediately reconnect the test blank to V
4
, leaving it
disconnected from V
5
. Allow dry nitrogen to flow
through the test blank for five minutes to purge the air
from inside before reconnecting to V
5
. Switch the gas to
flow only through the blank and not through the bypass
loop. The APIMS will show an increase in moisture
concentration. Record the APIMS output until it
reaches 500 ppt or for 24 hours, whichever is less.
7.1.1.2 Repeat the above test twice for a total of three
data sets.
7.1.2 Moisture Input Test — After the APIMS has
returned to equilibrium with its initial moisture level,
switch the input gas to 50 ppb moisture (by switching
valves V
1
, V
2
, and V
3
simultaneously, if the
arrangement of Figure 1 is used) while recording the
APIMS response. Maintain this moisture input for 20
minutes. Switch the input to the test blank back to dry
nitrogen and record the decease in moisture level until
the initial background is again reached.
7.1.2.1 Repeat the above test twice for a total of three
data sets.
7.2 Tests on Actual Specimen
7.2.1 Initial Dry Down — This portion of the test is
designed to determine the quantity of removable
moisture on the wetted surfaces of a component in the
condition in which it is typically supplied. Thus, the
results of this test will reflect, by design, any
precautions which the supplier has taken to remove
moisture and maintain its dryness during shipping.
7.2.1.1 Start the test as in Section 7.1.1. The test
specimen should not be unpacked until after the APIMS
has equilibrated with the background moisture level. If
a glove box or other such enclosure is used, unpack the
component in the glove box. Switch the dry nitrogen
flow to pass through the bypass loop while maintaining
a small component through the test blank. Undo the
final layer of packing and any shipping caps or plugs on
the component at this point. Remove the sample blank
and connect the test specimen to valve V
4
as quickly as
possible. Allow the dry nitrogen flow to purge out any
ambient air in the specimen for five minutes, then
connect the specimen to V
5
. Switch the gas to flow only
through the test specimen and not through the bypass
loop. The APIMS will show an increase in moisture
concentration. Record the APIMS output until it
reaches 500 ppt or for 24 hours, whichever is less.
7.2.1.2 After 24 hours, the dry-down part of the test
can be terminated. If necessary, the moisture level can
be reduced below 500 ppt by baking the component at
200°C or the maximum temperature allowed by the
manufacturer for six hours and/or purging at the highest
flow of dry gas available. If this is not sufficient to
reduce the moisture concentration below 500 ppt, the
test may be abandoned.
7.2.2 Moisture Input Test — After the APIMS has
returned to equilibrium with its initial moisture level,
switch the input gas to 50 ppb moisture (by switching
valves V
1
, V
2
, and V
3
simultaneously, if the
arrangement of Figure 1 is used) while recording the
APIMS response. Maintain this moisture input for 20
minutes. Switch the input to the test blank back to dry
nitrogen and record the decease in moisture level until
the initial background is again reached.
7.2.2.1 Repeat the above test twice for a total of three
data sets.
7.2.2.2 In case the moisture level recorded by the
APIMS does not reach 50 ppb within 20 minutes, an
additional test should be performed in which the
moisture input should be continued until equilibrium is
reached. Then switch the input to the test specimen
back to dry nitrogen and record the decease in moisture
level until the initial background is again reached.
This test should be repeated once (two data sets total).
7.2.3 Bake-Out Test — Heat component to maximum
bake temperature, according to manufacturer’s
specifications, for three hours. Allow it to return to
room temperature. Continue by repeating the moisture
input test, after baking, as in Section 7.2.2.
7.3 Temperature, Flow Rate, and Pressure
Specification — The recommended test pressure is 700
kPa (7 bar). However, lower test pressures are
appropriate for some models of APIMS. Also, the
maximum pressure rating of the test components should
not be exceeded. The component temperature shall be
maintained at 35 ± 1°C during the test.
7.3.1 In the case of tubing, the test shall be carried out
at a single flow determined by the tubing diameter
according to the following table. In the case of valves,
regulators, mass flow controllers (MFC’s), and passive
components (gauges, flow meters, and fittings), the test
will be carried out at a single flow determined by the
table according to the size of the connecting tube stubs.

SEMI F27-0997 © SEMI 1997, 2003 5
Regulators should be tested in the fully-open condition
(i.e., with the regulator adjusted for minimum pressure
drop). The quantity of published data on regulator
interaction with moisture is limited, and experiments to
investigate the effect of varying the pressure drop
across the regulator are encouraged but are not part of
this test. MFC’s should be tested with a flow control
device upstream and the MFC not actively controlling
the flow through it (usually referred to as the “purge”
setting). It may not be possible to test a given MFC at
the specified flow if the flow rating of the MFC is much
less than the specified flow. As MFC’s are not expected
to vary greatly in performance according to their flow
rating, MFC’s to be tested should generally be chosen
to be compatible with the table. If this is not possible,
then the MFC should be tested at its rated flow (lower
than the flow in the table). This can then be considered
a conservative test, as the MFC operated at higher flow
would be expected to dry down more quickly.
7.3.2 Particle filters with the same size tubing
connectors often have somewhat different sizes and
very different flow ratings. In order to be able to make
reasonable comparisons between the performance of
different filters, it is essential to test filters of similiar
size at the same flow. However, the flow dependence of
the moisture response of particle filters is more
complex than that of tubing, and they must be tested at
more than one flow in order to model their behavior. In
order to meet this requirement and provide data which
can be readily compared with the conditions under
which the filters can be expected to be used in practice,
filters shall be tested at two flows: The first, according
to the size of the connecting tube stubs, the second, to
be 25% of the rated flow of the filter or 50 slm,
whichever is less.
7.3.3 For systems of components, the system shall be
tested at the lowest flow of those determined for each
component in the system considered separately. If the
system includes a filter, it shall be tested at that flow
and additionally, at 25% of the rated flow of the filter or
50 slm, whichever is less.
Table 1 Flow Rate
Tubing Outer Diameter
(nominal)
Flow (all
components)
Flow for Second
Test on Filters
≥ 1/8", < 1/2"
(≥ 3.2 mm, < 12.7 mm)
6 slm
≥ 1/2"
(≥ 12.7 mm)
13 slm
25% of rated flow
or 50 slm
whichever is less
8 Reporting Results
8.1 Complete moisture response curves for all test
specimens and relevant test blanks should be included.
A summary sheet may compare components in terms of
“induction time,” peak height, and/or decay time.
Temperature, pressure, and flow measurements, and as
complete a record as possible of all experimental
variables should be noted.
NOTICE: SEMI makes no warranties or
representations as to the suitability of the standards set
forth herein for any particular application. The
determination of the suitability of the standard is solely
the responsibility of the user. Users are cautioned to
refer to manufacturer’s instructions, product labels,
product data sheets, and other relevant literature
respecting any materials mentioned herein. These
standards are subject to change without notice.
The user’s attention is called to the possibility that
compliance with this standard may require use of
copyrighted material or of an invention covered by
patent rights. By publication of this standard, SEMI
takes no position respecting the validity of any patent
rights or copyrights asserted in connection with any
item mentioned in this standard. Users of this standard
are expressly advised that determination of any such
patent rights or copyrights, and the risk of infringement
of such rights, are entirely their own responsibility.
Copyright by SEMI® (Semiconductor Equipment and Materials
International), 3081 Zanker Road, San Jose, CA 95134. Reproduction o
f
the contents in whole or in part is forbidden without express written
consent of SEMI.

SEMI F28-1103 © SEMI 1997, 2003 1
SEMI F28-1103
TEST METHOD FOR MEASURING PARTICLE GENERATION FROM
PROCESS PANELS
This test method was technically approved by the Global GasesCommittee and is the direct responsibility of
the North American Gases Committee. Current edition approved by the North American Regional Standards
Committee on September 25, 2003. Initially available at www.semi.org October 2003; to be published
November 2003. Originally published September 1997.
1 Purpose
1.1 The purpose of this document is to define a method
for testing process panels intended for installation in
high-purity gas distribution systems. Application of this
test method is expected to yield comparable data among
process panels tested for the purposes of qualification
for this installation.
1.2 This document describes a test method designed to
draw comparisons of particulate generation
performance of process panels. This test method
evaluates the cleanliness of process panels in the “as
received” condition as well as under normal operating
conditions. The “as received” test is intended to enable
the user to evaluate the fabrication, cleaning, and
packaging techniques of the manufacturer of the
process panel. The test under actual operating
conditions is intended to allow the user to evaluate the
manufacturer’s component selection as well as the
quality of the panel design. The specific flow rates
described in both test methods are representative of
relatively high flow conditions for a typical process
panel.
2 Scope
2.1 This test method addresses total particle counts
greater than the minimum detection limit (MDL) of the
particle counter and does not consider classifying data
into various size ranges.
2.2 This procedure utilizes a particle counter applied to
process panels typically used in semiconductor
applications. It applies to process gas supply systems
(e.g., gas cabinets) which include a process panel, an
inert purge panel, and a system vent. Both automatic
and manual process panels are within the scope of this
test procedure. Panels, as defined in this test method,
are considered to consist of 6.35 mm O. D. × 0.89 mm
wall (1/4" O.D. × 0.035" wall) tubing and components.
NOTICE: This standard does not purport to address
safety issues, if any, associated with its use. It is the
responsibility of the users of this standard to establish
appropriate safety and health practices and determine
the applicability of regulatory or other limitations prior
to use.
3 Limitations
3.1 This test method specifies flow and mechanical
stress conditions considered typical of conditions which
would be expected under moderately aggressive
operating conditions. These conditions should not
exceed those recommended by the manufacturer.
Actual performance under operating conditions at lower
flow rates or less aggressive conditions may differ. This
test method does not address particle generation under
vibrating conditions.
3.2 The test medium is limited to nitrogen, argon, or
clean dry air (CDA). Performance with other gases may
differ.
3.3 The accuracy of the data generated by this method
is limited to the accuracy of the particle measuring
instruments utilized.
3.4 This method is written with the assumption that the
operator understands the use of the apparatus at a level
equivalent to six months of experience.
3.5 This document is not intended as a methodology
for monitoring on-going particulate performance once a
particular process panel has been tested. Also, this
method does not include extended dynamic particle
count testing (particle generation after thousands of
cycles).
3.6 Auto-crossover systems are not within the scope of
this test procedure.
NOTE 1: It should be mentioned that test results from panels
equipped with final outlet filters can differ significantly from
test results from panels without filters.
4 Referenced Standards
4.1 ISO Standards
1
ISO 14644-1 — Cleanrooms and associated controlled
environments Part 1: Classification of air cleanliness.
1 International Organization for Standardization, ISO Central
Secretariat, 1, rue de Varembé, Case postale 56, CH-1211 Geneva 20,
Switzerland. Telephone: 41.22.749.01.11; Fax: 41.22.733.34.30,
Website: www.iso.ch